//! Event-driven synthetic QRS, mechanical pulse and respiratory generation. //! A conducted QRS schedules an arterial pulse and a delayed, smoothed pleth pulse. use std::collections::VecDeque; use crate::model::{ArterialShape, CapnoShape, EcgRhythm, Parameters, PulseConduction}; const SAMPLE_STEP: f64 = 0.01; // 100 Hz const HISTORY_SECONDS: f64 = 10.0; const ART_DELAY: f64 = 0.12; // QRS -> arterial upstroke, seconds (schematic) const PLETH_DELAY: f64 = 0.30; // QRS -> peripheral pleth onset, seconds const PLETH_FILTER_TAU: f64 = 0.16; // seconds: deliberately strong damping const QRS_PHASE: f64 = 0.335; /// One mechanically conducted QRS. Respiratory modulation is sampled on the /// beat, so pressure varies from beat to beat rather than within a single beat. #[derive(Clone, Copy, Debug)] struct ConductedBeat { qrs_time: f64, pulse_pressure_scale: f64, } /// PPV = 100 * (PPmax - PPmin) / ((PPmax + PPmin) / 2). /// For a requested PPV of 20%, pulse pressure oscillates between 90% and 110% /// of its nominal value over one respiratory cycle. fn ppv_multiplier(ppv_percent: i32, respiratory_phase: f64) -> f64 { 1.0 + (ppv_percent.clamp(0, 60) as f64 / 200.0) * (std::f64::consts::TAU * respiratory_phase).sin() } #[derive(Clone, Copy, Debug)] pub struct Sample { pub time: f64, pub ecg: f64, // arbitrary amplitude pub pleth: f64, // attenuated, low-pass-filtered amplitude 0..1 pub art: f64, // absolute pressure in mmHg; display clips above 200 pub co2: f64, // mmHg } pub struct Simulator { samples: VecDeque, time: f64, ecg_phase: f64, respiratory_phase: f64, ecg_beat_number: u64, conducted_qrs: VecDeque, // captured QRS with respiratory beat amplitude rate_qrs: VecDeque, // captured beats for measured PR; reset after rate changes filtered_pleth: f64, previous_rhythm: EcgRhythm, previous_conduction: PulseConduction, previous_heart_rate: i32, pending_time: f64, } impl Simulator { pub fn new(params: &Parameters) -> Self { let mut sim = Self { samples: VecDeque::with_capacity(1_100), time: -8.0, ecg_phase: 0.0, respiratory_phase: 0.0, ecg_beat_number: 0, conducted_qrs: VecDeque::new(), rate_qrs: VecDeque::new(), filtered_pleth: 0.0, previous_rhythm: params.ecg_rhythm, previous_conduction: params.pulse_conduction, previous_heart_rate: params.heart_rate, pending_time: 0.0, }; // Prepopulate the trace and pulse history, including the initial pulse rate. for _ in 0..800 { sim.step(params); } sim } pub fn advance(&mut self, elapsed_seconds: f64, params: &Parameters) { // Cap huge steps after a suspended/minimized window. self.pending_time += elapsed_seconds.clamp(0.0, 0.1); while self.pending_time >= SAMPLE_STEP { self.step(params); self.pending_time -= SAMPLE_STEP; } } pub fn samples(&self) -> &VecDeque { &self.samples } /// Average of recent *mechanically captured* beat intervals, not an independent setting. pub fn pulse_rate(&self) -> Option { if self.rate_qrs.len() < 2 { return None; } let end = self.rate_qrs.len() - 1; // Use the last three intervals, so PR settles rapidly after a rate change. let start = end.saturating_sub(3); let first = self.rate_qrs[start]; let last = self.rate_qrs[end]; // Suppress stale pulse values after a prolonged period without capture. if self.time - last > 12.0 || last <= first { return None; } Some((((end - start) as f64 * 60.0) / (last - first)).round() as i32) } fn step(&mut self, params: &Parameters) { self.time += SAMPLE_STEP; if params.ecg_rhythm != self.previous_rhythm || params.pulse_conduction != self.previous_conduction || !params.ecg_rhythm.has_qrs() || params.pulse_conduction == PulseConduction::None || params.heart_rate <= 0 { // A rhythm/capture change terminates previously queued mechanical events. self.conducted_qrs.clear(); self.rate_qrs.clear(); self.filtered_pleth = 0.0; self.previous_rhythm = params.ecg_rhythm; self.previous_conduction = params.pulse_conduction; } if params.heart_rate != self.previous_heart_rate { // Old captured events remain in the *waveforms* until they pass. // They must not bias the pulse rate after an HR slider change. self.rate_qrs.clear(); self.previous_heart_rate = params.heart_rate; } // Respiratory phase drives both the capnogram and arterial PPV. let respiratory_rate = params.respiratory_rate.max(0) as f64 / 60.0; self.respiratory_phase = (self.respiratory_phase + SAMPLE_STEP * respiratory_rate).fract(); if params.ecg_rhythm.has_qrs() && params.heart_rate > 0 { let irregularity = if params.ecg_rhythm == EcgRhythm::AtrialFibrillation { // Reproducible irregular R-R durations. let x = (self.ecg_beat_number.wrapping_mul(1_103_515_245) .wrapping_add(12_345) % 997) as f64 / 997.0; 0.65 + x * 0.70 } else { 1.0 }; let increment = SAMPLE_STEP * params.heart_rate as f64 / 60.0 / irregularity; let next_phase = self.ecg_phase + increment; if self.ecg_phase < QRS_PHASE && next_phase >= QRS_PHASE { // Interpolate within the 10ms sample so mechanical onsets stay aligned. let qrs_time = self.time - SAMPLE_STEP * (next_phase - QRS_PHASE) / increment; self.ecg_beat_number += 1; if params.pulse_conduction.conducts(self.ecg_beat_number) { let pulse_pressure_scale = if params.respiratory_rate > 0 { ppv_multiplier(params.pulse_pressure_variation, self.respiratory_phase) } else { 1.0 // no respiratory swing during apnea }; self.conducted_qrs.push_back(ConductedBeat { qrs_time, pulse_pressure_scale }); self.rate_qrs.push_back(qrs_time); } } self.ecg_phase = next_phase.fract(); } else { self.ecg_phase = 0.0; } // Preserve up to 30s of delivered events for stable derived PR calculations. while self.conducted_qrs.front().is_some_and(|beat| beat.qrs_time < self.time - 30.0) { self.conducted_qrs.pop_front(); } while self.rate_qrs.front().is_some_and(|&qrs| qrs < self.time - 30.0) { self.rate_qrs.pop_front(); } let ecg = ecg_amplitude(self.ecg_phase, self.time, params.ecg_rhythm, params.heart_rate); let (art, pleth_target) = self.mechanical_signals(params); let smoothing = 1.0 - (-SAMPLE_STEP / PLETH_FILTER_TAU).exp(); self.filtered_pleth += smoothing * (pleth_target - self.filtered_pleth); let pleth = self.filtered_pleth.clamp(0.0, 1.0); let co2 = params.etco2.max(0) as f64 * if params.respiratory_rate <= 0 { 0.0 } else { capno_amplitude(self.respiratory_phase, params.capno_shape) }; self.samples.push_back(Sample { time: self.time, ecg, pleth, art, co2 }); while self.samples.front().is_some_and(|first| first.time < self.time - HISTORY_SECONDS) { self.samples.pop_front(); } } fn mechanical_signals(&self, params: &Parameters) -> (f64, f64) { let systolic = params.art_systolic.clamp(0, 200) as f64; let diastolic = params.art_diastolic.clamp(0, 200).min(params.art_systolic.clamp(0, 200)) as f64; let pulse_pressure = (systolic - diastolic).max(0.0); let mut latest_art_age: Option = None; let mut art_peak: f64 = 0.0; let mut pleth_peak: f64 = 0.0; for beat in &self.conducted_qrs { let art_age = self.time - beat.qrs_time - ART_DELAY; if art_age >= 0.0 { latest_art_age = Some(art_age); art_peak = art_peak.max( arterial_amplitude(art_age, params.arterial_shape) * beat.pulse_pressure_scale, ); } let pleth_age = self.time - beat.qrs_time - PLETH_DELAY; if pleth_age >= 0.0 { pleth_peak = pleth_peak.max(pleth_amplitude(pleth_age)); } } let art = if let Some(age) = latest_art_age { // Pressure decays after an unperfused interval instead of showing a // non-existent 75 mmHg baseline during prolonged missed captures. let baseline = diastolic * (-(age - 0.60).max(0.0) / 1.40).exp(); (baseline + pulse_pressure * art_peak).clamp(0.0, 250.0) } else { 0.0 }; (art, pleth_peak) } } fn gaussian(x: f64, center: f64, width: f64) -> f64 { let d = (x - center) / width; (-0.5 * d * d).exp() } fn phase_gaussian(phase: f64, center: f64, width: f64) -> f64 { let distance = (phase - center + 0.5).rem_euclid(1.0) - 0.5; gaussian(distance, 0.0, width) } fn ecg_amplitude(phase: f64, time: f64, rhythm: EcgRhythm, heart_rate: i32) -> f64 { if rhythm == EcgRhythm::VentricularFibrillation { // Non-periodic-looking composite oscillation with no organized QRS. return 0.29 * (time * 42.1 + 0.40 * (time * 3.1).sin()).sin() + 0.22 * (time * 57.2 + 0.57 * (time * 5.7).sin()).sin() + 0.14 * (time * 31.4).sin(); } if heart_rate <= 0 || rhythm == EcgRhythm::Asystole { return 0.006 * (time * 9.3).sin(); } match rhythm { EcgRhythm::Sinus => { 0.16 * phase_gaussian(phase, 0.16, 0.030) - 0.19 * phase_gaussian(phase, 0.300, 0.011) + 1.26 * phase_gaussian(phase, QRS_PHASE, 0.008) - 0.32 * phase_gaussian(phase, 0.363, 0.014) + 0.32 * phase_gaussian(phase, 0.61, 0.070) } EcgRhythm::AtrialFibrillation => { 0.023 * (time * 39.0).sin() + 0.014 * (time * 57.0).sin() - 0.19 * phase_gaussian(phase, 0.300, 0.011) + 1.26 * phase_gaussian(phase, QRS_PHASE, 0.008) - 0.32 * phase_gaussian(phase, 0.363, 0.014) + 0.28 * phase_gaussian(phase, 0.61, 0.072) } EcgRhythm::VentricularTachycardia => { 1.06 * phase_gaussian(phase, 0.31, 0.086) - 0.49 * phase_gaussian(phase, 0.46, 0.067) - 0.16 * phase_gaussian(phase, 0.72, 0.10) } EcgRhythm::VentricularFibrillation | EcgRhythm::Asystole => unreachable!(), } } fn arterial_amplitude(age: f64, shape: ArterialShape) -> f64 { if age < 0.0 { return 0.0; } let value = match shape { ArterialShape::Normal => { let envelope = if age < 0.13 { (age / 0.13).powf(1.8) } else { (-(age - 0.13) / 0.25).exp() }; envelope - 0.12 * gaussian(age, 0.35, 0.023) + 0.055 * gaussian(age, 0.43, 0.045) } ArterialShape::Damped => { let envelope = if age < 0.22 { (age / 0.22).powf(1.3) } else { (-(age - 0.22) / 0.39).exp() }; envelope * 0.82 } ArterialShape::Hyperdynamic => { let envelope = if age < 0.085 { (age / 0.085).powf(1.9) } else { (-(age - 0.085) / 0.18).exp() }; envelope - 0.16 * gaussian(age, 0.27, 0.022) + 0.095 * gaussian(age, 0.36, 0.034) } }; value.clamp(0.0, 1.0) } fn pleth_amplitude(age: f64) -> f64 { if age < 0.0 { return 0.0; } // Heavily attenuated, broad pulse, without a prominent dicrotic notch. 0.75 * (1.0 - (-age / 0.13).exp()) * (-age / 0.48).exp() } fn capno_amplitude(phase: f64, shape: CapnoShape) -> f64 { match shape { CapnoShape::Normal => { if phase < 0.30 { 0.0 } else if phase < 0.38 { (phase - 0.30) / 0.08 * 0.92 } else if phase < 0.76 { 0.92 + 0.08 * ((phase - 0.38) / 0.38) } else if phase < 0.82 { 1.0 - (phase - 0.76) / 0.06 } else { 0.0 } } CapnoShape::Obstructive => { if phase < 0.28 { 0.0 } else if phase < 0.76 { ((phase - 0.28) / 0.48).powf(0.68) } else if phase < 0.83 { 1.0 - (phase - 0.76) / 0.07 } else { 0.0 } } }.clamp(0.0, 1.0) } #[cfg(test)] mod tests { use super::*; fn blank(p: &Parameters) -> Simulator { let mut sim = Simulator::new(p); sim.samples.clear(); sim.conducted_qrs.clear(); sim.rate_qrs.clear(); sim.time = 0.0; sim.ecg_phase = 0.0; sim.ecg_beat_number = 0; sim.filtered_pleth = 0.0; sim } #[test] fn signals_remain_in_expected_ranges() { let p = Parameters::default(); let mut sim = Simulator::new(&p); sim.advance(0.1, &p); assert!(sim.samples().len() >= 800); for s in sim.samples() { assert!((0.0..=1.0).contains(&s.pleth)); assert!((0.0..=250.0).contains(&s.art)); assert!((0.0..=f64::from(p.etco2)).contains(&s.co2)); assert!(s.ecg.is_finite()); } } #[test] fn qrs_triggers_delayed_art_and_later_pleth() { let p = Parameters::default(); let mut sim = blank(&p); while sim.conducted_qrs.is_empty() { sim.step(&p); } let qrs = sim.conducted_qrs.back().unwrap().qrs_time; let art_before = sim.samples.back().unwrap().art; assert_eq!(art_before, 0.0); while sim.time < qrs + ART_DELAY + 0.13 { sim.step(&p); } assert!(sim.samples.back().unwrap().art > p.art_diastolic as f64); assert_eq!(sim.samples.back().unwrap().pleth, 0.0); while sim.time < qrs + PLETH_DELAY + 0.20 { sim.step(&p); } assert!(sim.samples.back().unwrap().pleth > 0.03); } #[test] fn asystole_and_vf_and_pea_produce_no_mechanical_pulses() { for (rhythm, capture) in [ (EcgRhythm::Asystole, PulseConduction::OneToOne), (EcgRhythm::VentricularFibrillation, PulseConduction::OneToOne), (EcgRhythm::Sinus, PulseConduction::None), ] { let p = Parameters { ecg_rhythm: rhythm, pulse_conduction: capture, ..Parameters::default() }; let mut sim = Simulator::new(&p); for _ in 0..250 { sim.step(&p); } assert!(sim.pulse_rate().is_none()); assert!(sim.samples.iter().all(|s| s.art == 0.0 && s.pleth == 0.0)); } } #[test] fn switching_to_pea_cancels_existing_pulses_immediately() { let mut p = Parameters::default(); let mut sim = Simulator::new(&p); assert!(sim.pulse_rate().is_some()); p.pulse_conduction = PulseConduction::None; sim.step(&p); assert!(sim.pulse_rate().is_none()); assert_eq!(sim.samples.back().unwrap().art, 0.0); assert_eq!(sim.samples.back().unwrap().pleth, 0.0); } #[test] fn pulse_rate_follows_two_to_one_capture() { let p = Parameters { pulse_conduction: PulseConduction::TwoToOne, ..Parameters::default() }; let mut sim = blank(&p); for _ in 0..1_400 { sim.step(&p); } // 75 QRS/min, with every second QRS producing a pulse = 37.5/min. assert_eq!(sim.pulse_rate(), Some(38)); } #[test] fn pulse_rate_follows_three_to_one_capture() { let p = Parameters { pulse_conduction: PulseConduction::ThreeToOne, ..Parameters::default() }; let mut sim = blank(&p); for _ in 0..1_400 { sim.step(&p); } assert_eq!(sim.pulse_rate(), Some(25)); } #[test] fn pulse_rate_updates_when_heart_rate_slider_changes() { let mut p = Parameters::default(); let mut sim = Simulator::new(&p); assert_eq!(sim.pulse_rate(), Some(75)); p.heart_rate = 189; sim.step(&p); // No old 75 bpm PR is displayed while the new beat interval is measured. assert_eq!(sim.pulse_rate(), None); for _ in 0..200 { sim.step(&p); } assert_eq!(sim.pulse_rate(), Some(189)); } #[test] fn new_rate_respects_two_to_one_mechanical_capture() { let mut p = Parameters::default(); let mut sim = Simulator::new(&p); p.heart_rate = 180; p.pulse_conduction = PulseConduction::TwoToOne; for _ in 0..400 { sim.step(&p); } assert_eq!(sim.pulse_rate(), Some(90)); } #[test] fn asystole_is_flat_and_vf_has_waveform_without_qrs() { assert!(ecg_amplitude(0.335, 2.0, EcgRhythm::Asystole, 75).abs() < 0.01); assert!(ecg_amplitude(0.335, 2.0, EcgRhythm::VentricularFibrillation, 75).abs() > 0.01); } #[test] fn ppv_produces_requested_peak_to_trough_pulse_pressure_change() { let max_scale = ppv_multiplier(20, 0.25); let min_scale = ppv_multiplier(20, 0.75); assert!((max_scale - 1.10).abs() < 1e-12); assert!((min_scale - 0.90).abs() < 1e-12); let calculated_ppv = 100.0 * (max_scale - min_scale) / ((max_scale + min_scale) / 2.0); assert!((calculated_ppv - 20.0).abs() < 1e-10); assert_eq!(ppv_multiplier(0, 0.25), 1.0); } #[test] fn respiratory_phase_modulates_arterial_pulses_not_pleth() { let p = Parameters::default(); let mut sim = blank(&p); sim.time = ART_DELAY + 0.13; sim.conducted_qrs.push_back(ConductedBeat { qrs_time: 0.0, pulse_pressure_scale: 1.2, }); let (art_high, pleth_high) = sim.mechanical_signals(&p); sim.conducted_qrs.front_mut().unwrap().pulse_pressure_scale = 0.8; let (art_low, pleth_low) = sim.mechanical_signals(&p); assert!(art_high > art_low); assert_eq!(pleth_high, pleth_low); assert!(art_low > f64::from(p.art_diastolic)); } #[test] fn apnea_suppresses_ppv_and_hidden_ecg_does_not_suppress_pulses() { let p = Parameters { pulse_pressure_variation: 40, respiratory_rate: 0, show_ecg: false, ..Parameters::default() }; let sim = Simulator::new(&p); assert!(sim.conducted_qrs.iter().all(|b| b.pulse_pressure_scale == 1.0)); assert!(sim.pulse_rate().is_some()); assert!(sim.samples().iter().any(|s| s.art > 0.0)); } #[test] fn capnogram_reaches_etco2_plateau() { assert!((capno_amplitude(0.76, CapnoShape::Normal) - 1.0).abs() < 1e-8); } }