# Patient Monitor Simulation (PaMSim) — v0.3.0
A two-window, Rust/GTK4 patient-monitor simulator for teaching on Fedora Linux.
> **Synthetic training signals only — not for clinical use.** ECG, arterial pressure,
> plethysmography and capnography are schematic mathematical approximations. The
> simulation has no clinical validation and is not a medical device.
## Build and run on Fedora
```bash
sudo dnf install rust cargo gcc pkgconf-pkg-config gtk4-devel
cargo test
cargo run --release
```
Run these commands from the extracted `patient-monitor-simulation` folder.
Window titles show the application version. GTK4 requires a functional graphical
session (e.g., Fedora KDE/Wayland).
## Windows
**Admin:** Controls ECG rhythm and rate; mechanical capture; SpO₂; invasive
arterial pressure, waveform shape and pulse-pressure variation; NIBP; and
capnography. Numeric inputs have bidirectionally linked sliders and text fields.
Each of the five monitor parameters (ECG, SpO₂/pleth, ART, CO₂, NIBP) has its
own **Show on monitor** checkbox.
**Monitor:** Black background; ECG green, pleth yellow, ART red and capnogram
light grey. Waveforms and vital signs disappear together when their respective
visibility checkbox is cleared. Remaining waveform rows expand to use the space.
Hiding a display channel does **not** stop signal generation or physiological
coupling: for example, hiding ECG does not stop ART or pleth pulses. NIBP can
be hidden independently; this also removes its footer.
The monitor can be resized and toggled fullscreen with **F11**; **Esc** exits
fullscreen. On Wayland, move the monitor to the presentation screen before
activating fullscreen. Fullscreen does not lock down the computer.
## ECG and mechanical coupling
- ECG rhythms: sinus, atrial fibrillation, monomorphic ventricular tachycardia,
ventricular fibrillation and asystole.
- Mechanical capture: **1:1**, **2:1**, **3:1**, **None (PEA)**, or deterministic
pseudo-random capture (~65%). PR is derived from conducted QRS events, never
controlled by a pulse-rate slider.
- A captured QRS leads to an ART upstroke after a schematic **120 ms**, and a
peripheral pleth pulse after **300 ms**. Pleth is strongly damped/filtered.
- VF, asystole and PEA suppress mechanical pulses; old samples scroll out of
view. NIBP is a static, separately entered cuff reading.
## Invasive arterial pressure and PPV (new in v0.3)
The arterial waveform has a fixed **0–200 mmHg** display axis, with tick marks
at **0, 50, 100, 150 and 200 mmHg**. Nominal ART systolic and diastolic inputs
are limited to 0–200 mmHg.
**Pulse pressure variation (PPV)** is adjustable from **0% to 60%**, in 1%
increments, via synchronized slider and text field (Admin → Invasive arterial
blood pressure). Its default is **0%** (no respiratory modulation). The
respiratory cycle comes from the same respiratory-rate parameter as the
capnogram. **RR = 0** disables the respiratory modulation (apnea).
The synthetic PPV definition is:
```text
PPV (%) = 100 × (PPmax − PPmin) / ((PPmax + PPmin) / 2)
```
The model applies a sinusoidal variation to **pulse pressure**, sampled for
each mechanically conducted beat. The diastolic baseline remains nominally
fixed, so this is primarily a **beat-to-beat systolic swing**, not an arbitrary
vertical translation of the entire arterial trace. It does not affect ECG,
mechanical capture or pleth amplitude. This is a visual teaching model, not a
validated predictor of fluid responsiveness.
**Example:** ART 120/75 mmHg has nominal pulse pressure 45 mmHg. At PPV 20%,
the ideal pulse-pressure extrema are 40.5 and 49.5 mmHg, corresponding to
systolic peaks of approximately 115.5 and 124.5 mmHg (assuming the diastolic
baseline stays at 75 mmHg). The rendered waveform shape and finite number of
beats per respiratory cycle mean the visible maxima may not reach the ideal
extrema exactly.
The ART sidebar displays the **nominal** systolic/diastolic pressures and MAP,
plus the **configured** PPV percentage when PPV is greater than 0. It does not
measure PPV from sampled arterial data. At very high nominal systolic pressure,
the highest modulated pulses can exceed the 200 mmHg scale; the displayed
curve is then **clipped** at the top, without rescaling the axis.
## Display options (new in v0.3)
The five **Show on monitor** checkboxes affect the presentation only:
| Checkbox | Hidden monitor content |
| --- | --- |
| ECG | ECG waveform, HR and rhythm label |
| Pulse oximetry | Pleth waveform, SpO₂ and PR |
| Invasive arterial blood pressure | ART waveform, pressure values, MAP and PPV label |
| Capnography | CO₂ waveform, EtCO₂ and RR |
| NIBP | Static cuff pressure values and NIBP footer |
Remaining visible waveforms keep their relative order and receive an equal
share of the available area; zero waveform channels is supported.
## Signal implementation
- 100 Hz synthetic signal samples, 10 s retained history, 8 s scrolling trace.
- Cairo draws into the GTK4 `DrawingArea` and scales with the window.
- A conducted QRS records its own respiratory PPV multiplier at event time.
The arterial pulse thus retains the same amplitude throughout the beat.
- Varying PPV does not alter the peripheral pleth pulse or the number of
mechanical pulses.
- Clinical accuracy, dynamic NIBP measurements, alarms, artifacts, calibration,
precise waveform timing, and patient physiology are outside this prototype.
## Project layout
```text
src/
main.rs GTK application entry point
model.rs Parameters, rhythm/capture settings, display flags
simulation.rs QRS-timed mechanical signals, PPV and respiratory synthesis
render.rs Cairo monitor layout, dynamic visibility, fixed ART axis
ui.rs Admin controls and monitor window
```
## Changes since v0.2.2
1. Respiratory PPV with a 0–60% instructor setting, coupled to the simulated
respiratory cycle and applied **per arterial beat**.
2. Five independent display visibility toggles, dynamic waveform row layout,
and independent NIBP-footer hiding.
3. ART axis reduced from **0–250** to **0–200 mmHg**, with inputs adjusted to
the new range.
4. Added regression tests for PPV arithmetic and its effects on ART, for
hiding ECG while retaining mechanical pulses, for active channel selection,
and for arterial-axis normalization.
**Verification:** The archive is generated from these sources. Run `cargo test`
locally to verify the Rust/GTK build; Rust and GTK4 development components
were not available in the environment where this ZIP was prepared.