• v0.16.12 f0267f5493

    v0.16.12 Pre-release

    gase released this 2026-11-13 19:53:17 +00:00 | 2 commits to master since this release

    SimPaM — Patient Monitor Simulation v0.6.12

    A Rust/GTK4 patient-monitor simulator with two main windows for teaching on Fedora Linux.

    Synthetic training signals only — not for clinical use. ECG, arterial pressure,
    plethysmography and capnography are schematic mathematical approximations. The
    simulation has no clinical validation and is not a medical device.

    Overview

    SimPaM is a touchscreen-capable patient-monitor simulation for teaching
    anesthesia, intensive care and emergency medicine. An instructor changes the
    patient's physiology in the Admin window, while learners see a simulated
    bedside monitor that can be resized or shown fullscreen. The application uses
    synthetic signals and is intended for demonstrations and scenarios—not patient
    monitoring.

    Area What you can do
    Direct Mode Set ECG rhythm and HR, mechanical capture, oxygen saturation and signal quality, invasive blood pressure, capnography, NIBP and visibility.
    Actions & Occurrences Trigger time-based deterioration, defibrillation artifacts, simulated interventions and chest compressions; pause, resume or stop scripted changes.
    Monitor Display ECG, SpO₂/plethysmography, ART, capnography, alarms and numerical measurements. Touch a parameter to adjust its alarm limits.
    NIBP Start a cuff measurement manually or turn on automatic measurements; at startup there is no NIBP reading and Auto is off.
    Sound Hear two-tone alarms and the optional saturation-dependent QRS/pulse beep; selectively pause already-active audible alarms for 120 s.

    Guide contents

    Getting started

    Build and run on Fedora

    sudo dnf install rust cargo gcc pkgconf-pkg-config gtk4-devel pulseaudio-libs-devel
    cargo test
    cargo run --release
    

    Run these commands from the extracted patient-monitor-simulation folder.
    Window titles show the application version. GTK4 requires a functional graphical
    session (e.g., Fedora KDE/Wayland).

    First session

    1. Open the Admin and Monitor windows. In Direct Mode, choose a
      rhythm and set the initial values. The monitor updates continuously.
    2. If needed, enter a patient name and adjust alarm thresholds or choose
      Adult under Alert presets.
    3. For a blood-pressure cuff reading, use Measure now in Admin or the
      NIBP touch menu. Automatic measurement is initially OFF.
    4. Switch to Actions & Occurrences to trigger an event, use the yellow
      Defibrillation shortcut or control chest compressions. To present the
      monitor fullscreen, press F11; use Esc to exit.

    The following sections provide the complete controls, simulation behavior and
    technical details. The Changelog is kept at the end.

    Windows, monitor layout and touch operation

    Windows

    Admin sections have a narrow colored stripe on their left edge to match the
    monitor: ECG green, SpO₂ yellow, invasive arterial pressure red,
    capnography light grey, NIBP red. The Patient section remains
    neutral. These accents are drawn by GTK and remain visible independently of
    light/dark themes. This is a visual-only change; control behavior is unchanged.

    Admin: Controls patient name; ECG rhythm and rate; mechanical capture;
    SpO₂, signal quality and sensor connection; invasive arterial pressure,
    waveform shape and pulse-pressure variation; NIBP; and capnography. Numeric inputs have bidirectionally linked sliders and text fields.
    Each of the five monitor parameters (ECG, SpO₂/pleth, ART, CO₂, NIBP) has its
    own Show on monitor checkbox.

    Monitor: Black background; ECG green, pleth yellow, ART red and capnogram
    light grey. The patient name entered in Admin replaces PATIENT MONITOR in
    the header immediately; leaving the name blank restores the generic title.
    Long names are automatically resized or clipped to fit the header.
    Waveforms and vital signs disappear together when their respective
    visibility checkbox is cleared. Remaining waveform rows expand to use the space.
    Hiding a display channel does not stop signal generation or physiological
    coupling: for example, hiding ECG does not stop ART or pleth pulses.
    The final row is split into four equal-width measurement groups, with NIBP in
    the fourth group
    , aligned beneath the regular numeric column. The first
    three groups are reserved for future measurements. Hiding NIBP only clears
    its own fourth group, without removing the final row.

    The monitor can be resized and toggled fullscreen with F11; Esc exits
    fullscreen. On Wayland, move the monitor to the presentation screen before
    activating fullscreen. Fullscreen does not lock down the computer.

    Monitor layout (v0.5.7)

    The monitor uses a fixed 3:1 column ratio over the drawable area:
    75% waveform column / 25% numeric column, regardless of window size.
    The last row is reserved for four equal-width numerical measurement groups:

    Group 1 (25%) Group 2 (25%) Group 3 (25%) Group 4 (25%)
    CO / CI SVRI · GEDI · ELWI (empty) NIBP

    The first two groups currently display only their headings, with no numerical
    values or simulation logic. The third group is completely empty. The fourth
    retains the existing red NIBP readings, cuff gauge, progress bar and alarms.
    The NIBP visibility toggle hides only the fourth group's content; the
    bottom measurement row and other three groups remain present. Disabling one
    of the four waveform channels still reflows the remaining rows above it.

    Display options (new in v0.3)

    The five Show on monitor checkboxes affect the presentation only:

    Checkbox Hidden monitor content
    ECG ECG waveform, HR and rhythm alarm (when present)
    Pulse oximetry Pleth waveform, SpO₂ and PR
    Invasive arterial blood pressure ART waveform, pressure values, MAP and PPV label
    Capnography CO₂ waveform, EtCO₂ and RR
    NIBP Timed cuff-measurement values and red right-column value box

    Remaining visible waveform rows and the bottom measurements row share an equal
    portion of the available height and remain aligned across both columns; zero
    waveform channels is supported. Hiding NIBP leaves the three future measurement
    placeholders visible.

    Header controls: selective alarm pause and settings (v0.6.8)

    Two touch/mouse controls are drawn at the far right of the monitor header:

    • Yellow bell (rightmost): pauses only the currently active audible
      alarm conditions for 120 seconds. The caption paused for x s counts
      down the remaining seconds. Pressing the bell again replaces the suppressed
      set with the conditions active at that moment and restarts the 120-second
      timer. A newly triggered alarm condition sounds normally, even if another
      alarm of the same priority is paused. When a condition clears, it loses its
      paused status so a later recurrence can sound. Pauses never change visual
      alarms or the separate beat-synchronous QRS sound. Pressing the bell when
      no alarms are active does not start a countdown.
    • Gear (left of bell): opens a black/light-grey monitor-side settings panel
      with QRS beep, Audible alarms, Alert presets (Custom, Adult),
      an editable Patient name field and the attribution
      PaMSim by Critical Care Club Austria, (c) 2026. It does not create
      an on-screen keyboard. Apply synchronizes settings with Admin;
      Cancel/Esc discards changes. The Adult preset affects alarm limits only.

    Other channel touch menus are unchanged. NIBP retains its explicitly requested
    manual measurement, automatic measurement and interval controls.

    Monitor touchscreen: alarm limits and NIBP controls (v0.6.4)

    Tap or click the waveform or numerical readout of ECG, SpO₂, ART or CO₂
    to open that channel's alarm-limit panel. For NIBP, tap the rightmost
    quarter of the bottom measurement row
    . The other three bottom-row groups
    are reserved for future measurements and remain non-interactive. The panels
    are available in fullscreen and retain a black background, very light-grey
    text, and 0.8-pixel borders and dividers
    , in the style of the main monitor.

    Alarm thresholds can be modified from every monitor-side menu.
    No keyboard is required: every lower/upper limit has large − / + controls
    with a parameter-specific increment and an independent ON/OFF control.
    Unset limits remain disabled. The lower limit cannot exceed the upper limit.

    Touch area Available alarm limits Increment
    ECG HR 5 bpm
    SpO₂ SpO₂; pulse rate (PR) 1% SpO₂; 5 bpm PR
    ART SYS; MAP; DIA; PPV 5 mmHg pressure; 1% PPV
    CO₂ EtCO₂; respiratory rate (RR) 1 mmHg EtCO₂; 1 breath/min RR
    NIBP SYS; MAP; DIA 5 mmHg

    Apply commits the threshold changes and synchronizes the Admin window.
    Changing a threshold sets the alarm preset to Custom. Cancel or
    Esc discards the draft changes.

    NIBP is the exception: its monitor-side menu additionally provides Measure now,
    Automatic measurement ON/OFF, and the 1/3/5/10/20/30/60-minute interval.
    Interval and Auto changes take effect with Apply (or are discarded with
    Cancel/Esc); Measure now starts immediately, even with Auto OFF and before
    Apply. The touch controls synchronize with Admin. A measurement already in
    progress is not restarted. The simulated failure option and cuff placement
    remain Admin-only, as do ECG lead, SpO₂ sensor/quality and waveform choices.

    NIBP startup and automatic measurements

    Auto measurement is OFF by default. On startup the monitor shows
    ---/--- (MAP ---) and no measurement is initiated. To obtain a reading,
    select Measure now in the NIBP monitor menu (or in Admin), or enable
    Automatic NIBP measurement via the menu (or Admin). The first automatic cycle begins immediately after enabling Auto;
    subsequent measurement intervals count from completion of the prior cycle.
    Disabling Auto stops further scheduled measurements without interrupting a
    cycle already in progress. Manual measurement remains possible while Auto
    is off. The interval bar is empty and the monitor displays AUTO OFF
    whenever automatic measurements are disabled.

    Horizontal touchscreen alarm matrix (v0.6.3)

    All five monitor touch menus now share a compact horizontal alarm-limit matrix.
    Each parameter occupies its own row (for ART: SYS, MAP, DIA,
    PPV), with ↓ Lower and ↑ Upper displayed in adjacent columns.
    Within each limit cell, the touch controls are ordered + · value · − ·
    OFF/ON
    (the OFF/ON button is immediately right of minus). The original
    large button styling, per-parameter increments, last-value restoration,
    optional bounds and Apply/Cancel semantics are unchanged. The matrix scrolls
    if necessary while Apply and Cancel remain accessible below it. Other
    physiological controls remain Admin-only, except for the three explicitly
    requested NIBP measurement controls.

    Instructor controls and scenarios

    Actions & Occurrences (v0.6.9)

    The Admin window now has two tabs: Direct Mode (all original controls) and
    Actions & Occurrences (time-based instructor scripts). Select an action and
    Trigger action. Each running script appears in the action history with
    Pause / Resume / Stop and elapsed percentage. Stop freezes the
    physiological values at their current level and cancels any pending rhythm
    changes; it does not restore baseline. Restore pre-action baseline is
    an explicit reset that also stops all running scripts. Different actions can
    run concurrently; the latest action takes ownership of overlapping parameters.
    Manually adjusting any parameter in Direct Mode releases that parameter from
    script control, without stopping unrelated parameters. NIBP remains its own
    manual/automatic cuff measurement: its subsequent readings derive from ART.

    Action Scripted behavior
    Defibrillation Immediate high-amplitude ECG shock artifact; no rhythm change.
    Defibrillation + rhythm change Immediate selected rhythm and ECG artifact; instructor-selected ART SYS/DIA reached in 60 s, EtCO₂ target in 30 s only if effective perfusion exists.
    Deterioration: Hypoxia SpO₂ reaches randomized 40–50% (45 ±5) over 300 s. Below 60% saturation, HR gradually declines toward 50/min (or stays lower if initially under 50).
    Deterioration: Shock Over 600 s, ART MAP trends toward 45 mmHg via proportional SYS/DIA reduction. HR trends to 140 ±5% at 300 s; rhythm becomes AF and HR reaches 160 ±5% at 600 s. SpO₂ quality becomes Poor at MAP ≤59 and None / NO READING at MAP ≤50.
    Chest compressions Start/Stop and quality (Good/Adequate/Poor/Bad), using the existing Direct Mode CPR model.
    Norepinephrine bolus At 30 s, ART SYS 200 mmHg / HR 160; at 60 s, SYS 260 / HR 180; original SYS and HR recovered over the next 300 s.
    Too much propofol Over 60 s, ART reaches 60/45 mmHg and HR ~130 (±5%).
    Lethal propofol Over 60 s, ART reaches 40/20 mmHg and HR ~150 (±5%), followed by VF.

    Scope and assumptions: Scripts modify the invasive arterial-pressure model,
    not the NIBP settings or initial cuff reading. The ART numeric display may
    exceed the fixed 0–200 mmHg waveform axis during norepinephrine simulation;
    the trace clips at the top as before. Hypoxia bradycardia targets 50/min because
    only the onset threshold (60% saturation) was specified. The shock quality
    state “Bad” is mapped to the existing None / NO READING sensor-quality
    state. All actions are illustrative teaching scenarios, not pharmacokinetic
    or validated clinical models. Audible/visual alarms, the selective 120-second
    alarm pause, QRS sounds and monitor-side touchscreen menus are unchanged.

    Instructor shortcuts on the Actions tab (v0.6.10)

    The Actions & Occurrences tab keeps Chest compressions (Start/Stop button
    and quality selector: Good, Adequate, Poor, Bad) permanently visible above
    the scripted action selector. These are the same controls as on Direct Mode,
    backed by the same state; changes in either tab are reflected in the other.
    Chest compressions therefore no longer needs a redundant entry in the action
    dropdown. Its clinical simulation behavior is unchanged.

    A wide yellow Defibrillation shortcut with a red lightning-bolt icon is
    also always available above the scripted actions. One press produces the
    existing transient ECG defibrillation artifact only. It does not change
    rhythm, ART, EtCO₂ or other physiological parameters, and it does not add a
    long-running scripted action. For rhythm conversion and hemodynamic targets,
    continue to select Defibrillation + rhythm change from the dropdown.

    Chest compressions (v0.5.4)

    The Chest compressions Admin group provides a large Start Chest Compressions
    button (green background). While active it changes to Stop Chest Compressions
    (red background). Compression quality can be changed at any time:
    Good, Adequate (default), Poor, or Bad.

    • Compressions are generated independently of the ECG at 90–110/min, with
      a new reproducible pseudo-random compression rate and pressure each stroke.
      The ECG rhythm and electrical HR setting are not changed.

    • Every compression adds movement artifact to the ECG. In a patient with
      existing intrinsic arterial perfusion, the native arterial waveform is
      preserved but distorted by compression-synchronous pressure artifact.

    • With no intrinsic arterial pulse (e.g., VF, PVT, asystole or PEA),
      each effective compression generates a broad, blunt arterial pressure wave:

      Compression quality Compression-generated ART systolic peak
      Good 90 mmHg ±10% (81–99 mmHg)
      Adequate (default) 70 mmHg ±10% (63–77 mmHg)
      Poor 50 mmHg, +10% / −50% (25–55 mmHg)
      Bad No compression-generated arterial pressure; 0–5 mmHg baseline drift

      During pulseless CPR, ART SYS shows the current compression peak,
      while DIA/MAP remain -- (not estimable from this synthetic pressure).

    • SpO₂ numeric saturation and the pleth waveform remain visible only when
      compression quality is Good
      , provided the sensor is connected and not
      occluded by an NIBP cuff. Their effective signal quality is forced to
      Poor (★), irrespective of the normal Admin quality setting. The displayed
      pleth-derived PR follows the compression rate (90–110/min), not the ECG.
      Adequate, Poor and Bad CPR display NO READING; detached sensors display
      NO SENSOR. Compression-induced Poor quality does not schedule the usual
      intermittent dropout; that resumes when ordinary Poor signal mode returns.

    • Existing numerical alarms and rhythm alarms remain active. ART SYS limits
      apply to compression-generated systolic peaks; NIBP scheduling remains unchanged; its cycle duration still depends on
      effective SpO₂ quality at the start of a measurement.

    These pressure targets are didactic assumptions, not validated CPR
    hemodynamic criteria. The exercise must not be used to judge real compression
    quality or clinical perfusion.

    Measurements and signal behavior

    ECG and mechanical coupling

    • ECG rhythms: sinus, atrial fibrillation, monomorphic ventricular tachycardia
      (potentially perfusing), pulseless ventricular tachycardia (PVT),
      ventricular fibrillation and asystole.
    • Mechanical capture: 1:1, 2:1, 3:1, None (PEA), or deterministic
      pseudo-random capture (~65%). PR is derived from conducted QRS events, never
      controlled by a pulse-rate slider.
    • A captured QRS leads to an ART upstroke after a schematic 120 ms, and a
      peripheral pleth pulse after 300 ms. Pleth is strongly damped/filtered.
    • VT and PVT selections start with a heart rate of 160 bpm, and update
      both the HR slider and numeric entry. The instructor may modify this afterward.
    • PVT, VF, asystole and PEA suppress mechanical pulses; the ART trace then
      drifts smoothly at 0–5 mmHg rather than representing a pulsatile pressure.
      The ART SYS number shows the rounded low transducer pressure (0–5 mmHg),
      while DIA and MAP show -- because they cannot be reliably estimated.
      Previous pressure samples scroll out of view. NIBP follows a separate,
      time-based simulated cuff-measurement cycle.

    Smooth HR transitions (v0.6.7)

    Changing the HR slider now preserves the recent history of mechanically
    captured
    beats. PR converges to the new rate using the existing rolling
    three-interval average instead of temporarily disappearing. The SpO₂ value,
    plethysmography and saturation-dependent QRS tone remain available during
    ordinary HR adjustments. Actual loss of mechanical capture, disconnected
    sensors, deliberate signal dropout and low-quality/absent signals still
    produce their intended alarms immediately.

    Patient name and pulse oximeter signal state (new in v0.3.2)

    The new Patient name text field at the top of Admin immediately replaces the
    monitor headline. It is optional; an empty or whitespace-only name displays
    PATIENT MONITOR. This is a local teaching label, not a patient record.

    Admin → Pulse oximetry has two additional controls:

    • Sensor not attached (checkbox): immediately removes the pleth waveform,
      suppresses SpO₂ and pulse-rate readings, and displays NO SENSOR instead
      of the numeric saturation. This takes precedence over signal quality.
    • Signal quality (dropdown), which changes the visible pleth waveform and
      quality indicator, independently of the instructor-set SpO₂ percentage:
    Setting Pleth waveform Monitor display
    0 — None No waveform Flashing NO READING, PR --; no stars
    1 — Poor Small, irregularly distorted pulse, with temporary losses of signal 1 star (★) when available; blinking NO READING during loss
    2 — Adequate Medium, smoothly damped pulse 2 stars (★★)
    3 — Good Full nominal, smoothly damped pulse 3 stars (★★★)

    Default: sensor attached; signal quality 3 — Good. At levels 1–3, the
    instructor-set SpO₂ number is shown only when mechanical pulses are detected.
    In Poor quality, the signal also has intermittent complete dropouts:
    30–180 seconds of signal between interruptions, with each interruption
    lasting 5–20 seconds. During an interruption the pleth trace disappears,
    SpO₂ and PR are unavailable, and NO READING flashes. Following recovery,
    the Poor waveform and numerical readings return automatically. These intervals
    are pseudo-random and driven by simulation time, independent of frame rate;
    they reset when Poor quality is selected again. The dropout does not affect
    the actual saturation setting, ECG, derived mechanical pulse generation, or ART.
    The simulated pulse rate remains derived from captured QRS complexes.

    The degraded waveform uses deterministic, smoothly interpolated pseudo-noise
    that scales only existing pleth pulses: it cannot manufacture pulses during
    PEA, PVT, ventricular fibrillation or asystole. Unplugging a sensor or selecting
    quality 0 hides even previously buffered pleth samples immediately. ECG and
    ART continue unaffected. The Show on monitor checkbox still hides the
    entire SpO₂ row; sensor/quality controls leave the row visible so the status
    can be taught.

    Invasive arterial pressure and PPV (new in v0.3)

    The arterial waveform has a fixed 0–200 mmHg display axis, with tick marks
    at 0, 50, 100, 150 and 200 mmHg. Nominal ART systolic and diastolic inputs
    are limited to 0–200 mmHg.

    Pulse pressure variation (PPV) is adjustable from 0% to 60%, in 1%
    increments, via synchronized slider and text field (Admin → Invasive arterial
    blood pressure). Its default is 0% (no respiratory modulation). The
    respiratory cycle comes from the same respiratory-rate parameter as the
    capnogram. RR = 0 disables the respiratory modulation (apnea).

    The synthetic PPV definition is:

    PPV (%) = 100 × (PPmax − PPmin) / ((PPmax + PPmin) / 2)
    

    The model applies a sinusoidal variation to pulse pressure, sampled for
    each mechanically conducted beat. The diastolic baseline remains nominally
    fixed, so this is primarily a beat-to-beat systolic swing, not an arbitrary
    vertical translation of the entire arterial trace. It does not affect ECG,
    mechanical capture or pleth amplitude. This is a visual teaching model, not a
    validated predictor of fluid responsiveness.

    Example: ART 120/75 mmHg has nominal pulse pressure 45 mmHg. At PPV 20%,
    the ideal pulse-pressure extrema are 40.5 and 49.5 mmHg, corresponding to
    systolic peaks of approximately 115.5 and 124.5 mmHg (assuming the diastolic
    baseline stays at 75 mmHg). The rendered waveform shape and finite number of
    beats per respiratory cycle mean the visible maxima may not reach the ideal
    extrema exactly.

    The ART sidebar displays the nominal systolic/diastolic pressures and MAP,
    plus the configured PPV percentage when PPV is greater than 0. It does not
    measure PPV from sampled arterial data. At very high nominal systolic pressure,
    the highest modulated pulses can exceed the 200 mmHg scale; the displayed
    curve is then clipped at the top, without rescaling the axis.

    NIBP measurement cycle (v0.5.0)

    NIBP is red, with a dedicated value box in the right-hand numeric column.
    The bottom measurement row has three placeholder groups to its left, keeping
    NIBP aligned with the standard right-hand numeric column. The previous static
    NIBP footer has been removed. In Admin, choose an interval
    of 1, 3, 5, 10, 20, 30, or 60 minutes (default 5).
    Auto measurement is disabled at startup, so no reading is shown and no
    measurement starts until the instructor activates Auto or selects Measure now.
    When enabled, the first automatic measurement starts immediately; subsequent
    intervals run from the end of the last measurement, not from its beginning.
    Measure now starts a cycle even when Auto is OFF, if none is in progress. The small horizontal progress bar
    shows the fraction of the waiting interval elapsed; a vertical 0–200 mmHg
    gauge
    shows the current cuff pressure during measurement.

    The duration and success are determined by the signal quality at the start
    of the cycle
    (including an absent sensor or an ongoing Poor-signal dropout).
    Changing quality while a cycle is running does not retroactively change its
    planned duration:

    Starting SpO₂ quality Cycle duration Result
    Good (★★★) 20 s Measured
    Adequate (★★) 40 s Measured
    Poor (★) Randomly 60–120 s Simulated 65% chance of a reading
    None / no sensor 120 s No reading
    No reading checkbox 120 s No reading, regardless of quality

    The cuff inflates to 200 mmHg then deflates. When the starting signal is
    Poor or None, it deflates to about 67 mmHg, reinflates to about 133 mmHg,
    and finally deflates to zero. At the end, the completed value becomes visible
    and blinks five times by appearing/disappearing (without the inverted-color
    alarm effect). If the cycle fails, the displayed reading is ---/---, with
    MAP ---. Previous NIBP values remain visible while the next cuff cycle is
    in progress, until the new reading is available.

    The first successful measurement uses the initial NIBP SYS/DIA values from
    Admin. Subsequent successful values derive from the current invasive arterial
    SYS/DIA settings using independently sampled percentage deviations:

    Starting quality Deviation relative to IBP
    Good −5% to +5%
    Adequate −10% to +5%
    Poor −15% to −5%, if a reading succeeds
    None No reading

    Optional Cuff on SpO₂ extremity changes the effective pulse-oximeter
    quality while the cuff is inflated: above invasive systolic pressure = None;
    between invasive systolic and diastolic pressure = Poor; below invasive
    diastolic pressure = the configured quality. The quality never becomes better
    than the instructor's setting. NO READING, missing pleth waveform and
    restoration of the waveform are handled as before.

    Optional Cuff on ART extremity modifies the simulated arterial transducer
    trace during cuff inflation: above invasive SYS it freezes at the last measured
    pressure; between invasive SYS and DIA it pulsates between the cuff pressure
    and the currently available NIBP SYS reference; below invasive DIA it returns
    to the normal arterial waveform. The ART numerical SYS/DIA/MAP values also
    reflect this altered trace. With absent mechanical cardiac capture, the
    previous pulseless 0–5 mmHg simulation is retained and no artificial pulses
    are introduced. Both cuff-interference options are off by default.

    These cuff and SpO₂ interactions are deliberately simplified training
    artifacts
    ; they should not be interpreted as validated cuff oscillometry.

    Alarm limits and sounds

    Numerical alarm limits (v0.4.0; layout and presets updated in v0.4.1)

    Each displayed numeric measurement has an independent optional lower (↓) and
    upper (↑) alarm threshold
    in the corresponding Admin section. The lower
    and upper entries are placed in the same row as the corresponding slider.
    Derived values with no separate slider (PR, ART MAP, NIBP MAP) have their own
    compact alarm rows. The Admin window opens wider to accommodate these controls.
    Alarm fields are text inputs; leave a field blank (displayed as off) to disable just
    that threshold. With both fields blank, the whole numeric alarm is disabled.
    The value may equal either limit without triggering an alarm; only readings
    strictly below/above the respective threshold are out of range. If both limits
    are present, the lower limit must not exceed the upper limit. Invalid input is
    highlighted and the last valid configuration stays active.

    Monitor section Independently configurable alarms
    ECG HR
    SpO₂ Oxygen saturation (SpO₂), derived pulse rate (PR)
    Invasive arterial pressure (ART) Systolic (SYS), mean (MAP), diastolic (DIA), PPV
    NIBP Systolic (SYS), mean (MAP), diastolic (DIA)
    Capnography End-tidal CO₂ (EtCO₂), respiratory rate (RR)

    The current limits are displayed in small type at the upper-right corner of
    each channel, with an arrow for the corresponding direction. A disabled
    boundary is displayed as ---. For instance:

    ⚠ SYS ↓80 ↑180
      MAP ↓60 ↑---
      DIA ↓--- ↑---
    

    Alert presets

    The Alert presets dropdown at the top of Admin offers Custom (manual)
    and Adult. Choosing Adult immediately sets all alarm limits and updates
    the lower/upper text fields alongside the sliders. It changes only alarm limits:
    it does not change heart rate, blood pressure, channel visibility or any other
    simulated physiological setting. Editing any alarm field after applying Adult
    switches the selector to Custom without changing other alarm limits. Select
    Adult again to reapply the complete preset.

    Parameter Lower Upper
    HR (bpm) 60 100
    SpO₂ (%) 90 ---
    PR (bpm) --- ---
    ART SYS (mmHg) 80 180
    ART MAP (mmHg) 60 100
    ART DIA (mmHg) --- ---
    PPV (%) --- 10
    NIBP SYS (mmHg) 80 180
    NIBP MAP (mmHg) 60 100
    NIBP DIA (mmHg) --- ---
    EtCO₂ (mmHg) 30 50
    RR (/min) 10 20

    --- denotes a disabled threshold. Adult is a teaching convenience preset,
    not a clinically validated set of alarm limits. Initially the selector is
    Custom, with all limits disabled (as in v0.4.0).

    When a number violates a configured limit, that number flashes by inverting
    its foreground and background colors
    (500 ms inverted / 500 ms normal).
    Each pressure component (SYS, DIA, MAP) and the PPV reading flash independently. The waveform
    itself remains unchanged. All alarms are disabled by default so old
    scenarios do not unexpectedly flash.

    Numerical threshold alarms only evaluate actual displayed measurements:
    --, NO SENSOR, and other unavailable readings do not trigger low-value
    alarms. ECG-derived PR and signal visibility continue to work as in v0.3.2.
    Audible alarms with two priorities are supported, but there is no alarm
    latching or clinical alarm validation
    . Instructor settings are held in
    memory and reset on restart.

    Audible alarms and QRS tone (updated v0.6.7)

    Sound is enabled by default. The Admin window has an Audible alarms
    checkbox for muting and a Test tone button (works even while muted).
    Visual alarms keep blinking normally when sound is off.

    • Critical rhythm alarm: VT, PVT, VF and asystole produce a higher-pitched
      two-tone alert (880 Hz followed by 1160 Hz), repeated approximately
      every 1.5 seconds.
    • Numerical and SpO₂ technical alarms: out-of-range values, NO SENSOR
      and NO READING produce a lower-pitched two-tone alert
      (700 Hz followed by 950 Hz), repeated approximately every 2.5 seconds.
    • Both alarm signals use two different successive pitches, separated by
      a brief silence. The single QRS/pulse beep is independent and configurable
      using the new QRS tone checkbox in Admin (ON by default).
    • Critical alarms have priority: only one tone pattern is played even if
      multiple alarms are active. The sound stops when all alarms clear or when
      the checkbox is unchecked. Alarms on hidden monitor channels are excluded.
    • Values not yet available (e.g. NIBP before a completed measurement) do not
      trigger numerical limit alarms. The cuff measurement cycle by itself is not
      an alarm. Audio playback is separate from waveform rendering and simulation.

    Beat-synchronous QRS/pulse tone

    The QRS tone checkbox is independent of Audible alarms; turning off
    one does not mute the other. Each brief (85 ms), single-frequency beep
    follows an actual simulated beat event, not a separate rate timer:

    • SpO₂ reading available: beep on the delayed peripheral pleth pulse;
      the pitch follows the displayed saturation (including during effective
      good-quality chest compressions). Missed mechanical captures produce no
      peripheral beep.
    • SpO₂ reading unavailable: a fixed 500 Hz beep follows an organized
      ECG QRS if ECG is displayed; otherwise it follows an arterial pulse if
      ART is displayed. ECG and ART together do not cause duplicate beeps.
    • VF/asystole without effective CPR: no beat beep is generated. ART pulses
      created by CPR can generate the fallback tone, while good CPR with a valid
      SpO₂ reading uses the saturation-dependent tone.
    SpO₂ (%) Tone (Hz)
    97–100 740
    94–96 675
    90–93 609
    80–89 544
    70–79 479
    60–69 413
    50–59 348
    40–49 283
    30–39 218
    20–29 152
    0–19 87

    Saturation tone pitches are deliberately simplified teaching tones, not
    calibrated clinical pulse-oximeter sounds. The QRS beep and two-tone alarm can
    sound independently; no alarm priority or visual alarm logic was changed.

    As of v0.6.11, audio uses the Rust libpulse-simple-binding backend
    directly, rather than launching pw-play or paplay. SimPaM synthesizes
    the same 44.1 kHz mono PCM16 tones in memory and sends them to two
    independent PulseAudio streams (alarm and QRS). Each stream has a
    dedicated playback thread so synchronous audio I/O never blocks GTK.
    PipeWire supports this through its PulseAudio compatibility server.
    The system mixer controls output volume. If connecting or writing to
    the audio server fails, the worker logs an error and can retry on the
    next sound. Test tone verifies playback independently of the mute
    checkbox. No temporary WAV files or external player programs are used.

    Important: These simple teaching tones do not comply with medical-device
    alarm standards (including IEC 60601-1-8) and must never be relied on for
    clinical monitoring.

    Technical reference

    ECG morphology correction (v0.3.1)

    At high heart rates, the previous ECG generator expressed QRS width as a fixed
    fraction of the R–R interval and sampled at 100 Hz. Narrow R waves could fall
    between samples, making successive QRS complexes appear to vary dramatically
    in amplitude despite a regular sinus rhythm.

    • Synthetic P/QRS/T components now have widths specified in milliseconds,
      not fixed fractions of the cardiac cycle. P/T positioning and widths adapt
      to the R–R interval, while the QRS width remains essentially constant.
    • Sampling increased to 250 Hz (4 ms intervals) to capture QRS peaks
      more consistently, including at 189–240 bpm.
    • The ECG morphology remains independent of the respiratory arterial PPV
      setting. The arterial and pleth delays and mechanical capture options are
      unchanged.
    • Regression tests compare peak heights at 60, 75, 120, 189 and 240 bpm,
      and verify that changing PPV does not affect the ECG signal.

    The ECG remains a schematic educational signal; no voltage calibration,
    lead placement model, diagnostic criteria, or patient-specific physiology
    are simulated.

    Signal implementation

    • 250 Hz synthetic signal samples, 10 s retained history, 8 s scrolling trace.
    • Cairo draws into the GTK4 DrawingArea and scales with the window.
    • A conducted QRS records its own respiratory PPV multiplier at event time.
      The arterial pulse thus retains the same amplitude throughout the beat.
    • Varying PPV does not alter the peripheral pleth pulse or the number of
      mechanical pulses.
    • Clinical accuracy, calibration, artifact realism and patient-specific physiology
      are outside this prototype; its NIBP cycle and alarms are synthetic.

    Project layout

    src/
      main.rs        GTK application entry point
      audio_backend.rs  Native PulseAudio playback workers
      alarm.rs       Alarm/QRS tone synthesis, scheduling and pause
      model.rs       Parameters, rhythm/capture settings, display flags
      simulation.rs  QRS-timed mechanical signals, PPV and respiratory synthesis
      render.rs      Cairo monitor layout, dynamic visibility, fixed ART axis
      nibp.rs        Scheduled cuff measurements and measurement results
      ui.rs          Admin controls and monitor window
    

    Building audio in Flatpak

    For Flatpak, keep --socket=pulseaudio in finish-args and verify the
    target SDK provides the libpulse-simple development library:

    flatpak run --command=sh org.gnome.Sdk//51 \
        -c 'pkg-config --modversion libpulse-simple'
    

    The Flatpak runtime must also ship the corresponding shared libraries.
    The host's pipewire-utils package is no longer required. Update Cargo.lock
    with Cargo after changing dependencies, then regenerate cargo-sources.json
    using flatpak-cargo-generator before building offline. The project archive
    contains application sources; an existing Flatpak manifest is maintained
    separately and was not modified by this update.

    Application icon and desktop integration (v0.6.12)

    The CCCA logo is used as SimPaM's application icon. It appears on the monitor
    and Admin windows, the Linux desktop launcher, and Flatpak exports. The source
    image is stored as a square, scalable SVG in data/icons/hicolor/scalable/apps/.
    A GTK resource compiled by build.rs supplies the icon for direct Cargo runs.

    See FLATPAK-ICON.md for the one-line addition needed in your
    existing Flatpak manifest and for packaging commands. No new Cargo crates or
    new Flatpak permissions are required.

    Changelog

    Changes since v0.2.2

    1. Respiratory PPV with a 0–60% instructor setting, coupled to the simulated
      respiratory cycle and applied per arterial beat.
    2. Five independent display visibility toggles, dynamic waveform row layout,
      and independent NIBP value-box hiding.
    3. ART axis reduced from 0–250 to 0–200 mmHg, with inputs adjusted to
      the new range.
    4. Added regression tests for PPV arithmetic and its effects on ART, for
      hiding ECG while retaining mechanical pulses, for active channel selection,
      and for arterial-axis normalization.

    Verification: The archive is generated from these sources. Run cargo test
    locally to verify the Rust/GTK build; Rust and GTK4 development components
    were not available in the environment where this ZIP was prepared.

    Changes in v0.3.2

    1. Added live patient-name header in Admin/Monitor with blank-name fallback.
    2. Added four levels of SpO₂ signal quality, quality stars and an irregular,
      low-amplitude pleth waveform for Poor quality.
    3. Added Sensor not attached with NO SENSOR status and suppression of
      saturation, pulse rate and pleth waveform.
    4. Added tests for signal-quality scaling, displayed readings and disconnected
      sensor behavior.

    Build verification: ZIP integrity and patch application were checked.
    The GTK4 application and Rust tests were not compiled in the packaging
    environment; run cargo test and cargo run --release on Fedora.

    v0.4.2 — Rhythm, sensor and missing-reading alarms

    • VT remains potentially perfusing. The new PVT (pulseless ventricular tachycardia) rhythm retains organized broad QRS complexes but produces no arterial/pleth pulses. VF and asystole are also pulseless regardless of capture selection.
    • ECG: a flashing ARRHYTHMIA message for VT, PVT and VF; a flashing ASYSTOLE message for asystole. These event alarms do not depend on numerical HR limits.
    • SpO₂: when the probe is disconnected, a flashing NO SENSOR replaces the saturation reading. Quality 0, loss of mechanical capture or a pulseless rhythm results in flashing NO READING (no numerical SpO₂/PR). Sensor disconnection takes precedence. No pleth waveform is displayed in these conditions.
    • ART: without mechanical pulses (including PVT/VF/asystole/PEA), the simulated invasive pressure line has slow non-pulsatile drift between 0 and 5 mmHg. In v0.4.2, the numeric display showed --/--; v0.4.3 changes this to a rounded measured SYS of 0–5 mmHg, with -- for DIA and MAP.
    • ECG rhythm name has been removed from the monitor; only HR and the associated alarm message, when present, are visible. The instructor continues to select the rhythm in Admin.
    • Numerical alarm thresholds and the Adult preset from v0.4.1 are unchanged. Synthetic demonstration behavior only; not suitable for clinical decisions.

    v0.4.3 — Low arterial pressure, VT rate preset and intermittent Poor SpO₂

    • Pulseless rhythms, PEA or otherwise absent mechanical capture: invasive SYS
      reports the current measured 0–5 mmHg pressure, e.g. 3/-- with
      MAP -- mmHg. Only SYS is tested against configured ART SYS limits;
      no spurious MAP/DIA threshold alarms are generated from missing readings.
    • Selecting VT or PVT in Admin sets HR to 160 bpm and updates the
      associated slider and text field. Subsequent manual rate changes are allowed.
      Perfusing VT still produces QRS-linked pulses subject to capture settings;
      PVT remains pulseless.
    • Poor (★) SpO₂ signal periodically changes to NO READING for a
      pseudo-random 5–20 seconds, after a pseudo-random 30–180 seconds of
      usable low-quality signal. The technical alarm blinks, SpO₂/PR readings are
      suppressed, and the pleth waveform is absent. On recovery it resumes at Poor
      quality. The simulated signal history retains visible gaps for outages.
      Choosing another signal quality immediately clears the intermittent outage.
    • No changes to Adult alarm presets, alarm limit ordering, or the 0–200 mmHg
      arterial scale. All behavior is synthetic and intended for teaching only.

    v0.5.1 — Aligned monitor columns

    The left waveform column and right numeric column now share the same row grid.
    NIBP reserves an empty cell in the left waveform column instead of shortening
    the other numeric rows. Hiding any channel reflows both columns together.
    No NIBP measurement logic or alarm behavior was changed.

    v0.5.3 — Audible alarm output

    • Alarm-tone playback through Fedora PipeWire (pw-play) or PulseAudio
      compatibility (paplay), without any new Rust library dependencies.
    • Admin controls to mute/unmute alarms or play a test tone.
    • Distinct critical-rhythm and standard/technical patterns; critical priority.
    • Audio keeps running independently of the monitor's drawing callback.
    • Visual alarms and existing physiological simulation remain unchanged.

    v0.5.5 — CPR plethysmography correction

    The delayed CPR pleth waveform no longer wraps its trailing portion into the next compression. The regression test now reports the maximum valid waveform amplitude when failing. No other simulation controls were changed.

    v0.5.6 — CPR pleth filter persistence

    Corrects the underlying cause of the v0.5.5 test failure: in pulseless rhythms, the simulation had reset the pleth low-pass filter at every 250 Hz sample, preventing CPR-induced signals from accumulating. The filter now persists during good-quality chest compressions while QRS-triggered mechanical pulses are still cleared. Stopping CPR in VF/PVT immediately restores no reading. Additional regression coverage includes PVT.

    v0.5.7 — Four-group numerical measurement row

    The monitor now uses a 75%/25% waveform/numeric column split. The final
    row has four equally sized measurement groups: CO/CI; SVRI/GEDI/ELWI;
    reserved empty; and the unchanged NIBP readout. Only headings are rendered
    in the first two groups; the third has no content. NIBP can be hidden without
    removing the final row. No physiological simulation behavior was modified.

    v0.5.8 — ECG touchscreen controls

    Added a monitor-styled ECG settings window, opened by touching either the ECG
    waveform or numeric HR cell. Edits HR lower/upper alarms and displayed lead
    (I, II, III, aVR, aVL, aVF, V1–V6), with input validation and Admin sync.
    Lead changes are schematic display projections only; arrhythmia and pulse
    simulation remain unchanged.

    v0.5.9 — Touch-optimized ECG settings

    Restyled the ECG settings window with light-grey text/controls, a neutral
    0.8-pixel outer border and thin 0.8-pixel Cairo separators matching the
    monitor. Replaced the on-monitor HR limit text entries with large 5-bpm
    −/+ buttons and independent ON/OFF controls; the ECG lead remains a
    touch-selectable dropdown. Pending changes still require Apply, and Cancel
    leaves the shared parameters unchanged. No signal generation was altered.

    v0.6.0 — Additional touchscreen modules and NIBP manual-first startup

    Added touch/click settings panels for SpO₂, ART, CO₂ and the bottom-right
    NIBP cell. Their threshold editors use large, unit-specific +/− and ON/OFF
    controls; the panels expose relevant waveform/sensor options and stay
    synchronized with Admin. NIBP additionally provides interval selection,
    manual Measure now, Auto ON/OFF and cuff options. Startup no longer
    initiates a cuff cycle or shows any NIBP reading; Auto is OFF by default.
    Existing NIBP cycle physics and signal quality behavior are unchanged.

    v0.6.1 — Alarm-only monitor touch menus

    The ECG, SpO₂, ART, CO₂ and NIBP touch menus now edit only lower/upper
    alarm limits. Removed ECG lead, signal/sensor, waveform and all NIBP
    measurement controls from monitor-side menus. These remain available in
    Admin, including manual NIBP measurement and Auto ON/OFF. No physiological
    model, alarm evaluation or NIBP startup behavior was changed.

    v0.6.4 — NIBP touch controls restored

    The monitor-side NIBP settings menu again contains Measure now,
    Automatic measurement ON/OFF and the measurement interval (1, 3, 5,
    10, 20, 30, 60 minutes). These controls are above the horizontal SYS/MAP/DIA
    alarm matrix and share its monochrome touchscreen styling. Auto and interval
    are saved with Apply; Measure now starts a cuff cycle immediately regardless
    of Auto. Both monitor and Admin display the selected settings. Other monitor
    menus remain limited to alarm thresholds; NIBP cuff placement, the simulated
    failed-reading option and non-NIBP signal settings remain Admin-only.

    v0.6.5 — Two-tone audible alarms

    Both standard/technical and critical alarm sounds now consist of two distinct
    successive pitches. Critical alarms retain their priority and faster repeat
    interval; standard alarms retain their slower cadence. The Admin test button
    plays the standard two-tone alert. A single QRS beep is reserved for a later
    release and is not generated yet. No visual alarm or simulation logic changed.

    v0.6.6 — Optional QRS/pulse beep

    Added a separate QRS tone checkbox (enabled by default). An 85 ms
    single-note beep follows actual ECG or mechanical beat events. With a valid
    SpO₂ reading, the delayed pleth pulse determines the timing and the oxygen
    saturation determines the pitch in eleven steps from 740 to 87 Hz; otherwise
    ECG QRS or ART pulses trigger a fixed 500 Hz tone. The two-tone alarm sounds
    remain unchanged. Added regression tests for band boundaries, event selection
    and single-tone WAV output.

    v0.6.8 — Monitor header controls and selective audible-alarm pause

    Added a yellow bell button with a 120-second pause countdown and a grey settings
    gear to the monitor header. The pause suppresses only alarm conditions active
    when pressed, with automatic un-silencing of new/recurrent conditions; visual
    alarms and QRS tones continue. The global monitor settings window contains
    QRS and audible-alarm switches, alert preset selection, patient name and
    attribution, and synchronizes its changes with Admin.

    v0.6.9 — Actions & Occurrences

    Added a separate Admin tab with eight scripted interventions, configurable
    post-defibrillation rhythm and targets, pause/resume/stop per active timeline,
    manual override of individual parameters, overlapping-action ownership and
    an explicit baseline reset. Synthetic defibrillation ECG artifact and scripted
    invasive pressure/HR/SpO₂/EtCO₂ progressions are independent of NIBP cuff
    settings. Existing Direct Mode and monitor menus are retained.

    v0.6.10 — Persistent CPR controls and defibrillation shortcut

    Chest-compression Start/Stop and quality selection are always visible in the
    Actions & Occurrences tab above the scripted action dropdown and stay
    synchronized with Direct Mode. CPR was removed from the dropdown because the
    same controls are now permanently available. A high-contrast yellow one-tap
    Defibrillation button with a red lightning symbol triggers only the transient
    ECG shock artifact; rhythm conversion remains available as a separate action.

    v0.6.11 — Native PulseAudio / PipeWire audio output

    Replaced external pw-play / paplay subprocesses and temporary WAV files
    with two dedicated Rust playback workers using libpulse-simple-binding.
    The existing two-tone alarm sounds, SpO₂-dependent QRS tones, rhythms, mute
    controls, alarm priorities and selective alarm-pause logic are retained.
    A manual test tone remains available with audible alarms switched off.
    Flatpak requires the PulseAudio socket and libpulse-simple in its SDK/runtime.

    v0.6.12 — Application icon and desktop metadata

    Added the supplied CCCA SVG as a square, scalable application icon. GTK embeds
    it for local Cargo launches, and Flatpak assets include a matching desktop
    entry and AppStream metadata under at.criticalcare.SimPaM. Added a small
    Flatpak asset installer and packaging instructions; the physiological
    simulation and audio backend are unchanged.

    Downloads