v0.6.4 — NIBP touch controls restored
The monitor-side NIBP settings menu again contains **Measure now**, **Automatic measurement ON/OFF** and the **measurement interval** (1, 3, 5, 10, 20, 30, 60 minutes). These controls are above the horizontal SYS/MAP/DIA alarm matrix and share its monochrome touchscreen styling. Auto and interval are saved with Apply; Measure now starts a cuff cycle immediately regardless of Auto. Both monitor and Admin display the selected settings. Other monitor menus remain limited to alarm thresholds; NIBP cuff placement, the simulated failed-reading option and non-NIBP signal settings remain Admin-only.
This commit is contained in:
parent
54012ab960
commit
00dc22b1aa
9 changed files with 702 additions and 56 deletions
2
patient-monitor-simulation/Cargo.lock
generated
2
patient-monitor-simulation/Cargo.lock
generated
|
|
@ -438,7 +438,7 @@ dependencies = [
|
|||
|
||||
[[package]]
|
||||
name = "patient-monitor-simulation"
|
||||
version = "0.6.3"
|
||||
version = "0.6.4"
|
||||
dependencies = [
|
||||
"gtk4",
|
||||
]
|
||||
|
|
|
|||
|
|
@ -1,6 +1,6 @@
|
|||
[package]
|
||||
name = "patient-monitor-simulation"
|
||||
version = "0.6.3"
|
||||
version = "0.6.4"
|
||||
edition = "2021"
|
||||
license = "MIT"
|
||||
description = "Synthetic patient-monitor simulation for teaching (not for clinical use)"
|
||||
|
|
|
|||
|
|
@ -26,7 +26,7 @@ vierten Feldes; die gesamte untere Wertezeile bleibt bestehen.
|
|||
Beim Ausblenden anderer Parameter werden die verbleibenden Kurvenzeilen wie
|
||||
bisher dynamisch verteilt.
|
||||
|
||||
## Touch-Bedienung am Monitor: ausschließlich Alarmgrenzen (v0.6.1)
|
||||
## Touch-Bedienung am Monitor: Alarmgrenzen und NIBP-Funktionen (v0.6.4)
|
||||
|
||||
Durch Berühren oder Anklicken der **Kurve oder des numerischen Wertefeldes**
|
||||
von EKG, SpO₂, ART oder CO₂ öffnet sich das entsprechende Alarmmenü. Bei
|
||||
|
|
@ -36,7 +36,7 @@ Die Menüs funktionieren auch im Vollbildmodus und verwenden **schwarzen
|
|||
Hintergrund, sehr hellgraue Schrift sowie 0,8 Pixel breite Begrenzungs-
|
||||
und Trennlinien** wie der Hauptmonitor.
|
||||
|
||||
**Am Monitor können ausschließlich Alarmgrenzen eingestellt werden.**
|
||||
**In allen Monitor-Menüs können die Alarmgrenzen eingestellt werden.**
|
||||
Für jede untere und obere Alarmgrenze stehen große **−/+**-Tasten sowie
|
||||
separate **ON/OFF**-Tasten zur Verfügung. Nicht gesetzte Grenzwerte sind
|
||||
deaktiviert. Ein unterer Grenzwert darf den oberen nicht überschreiten.
|
||||
|
|
@ -55,18 +55,20 @@ Bei manueller Änderung wechselt das Alarm-Preset zu **Custom**. Mit
|
|||
**Cancel** oder **Esc** werden nicht übernommene Änderungen verworfen.
|
||||
|
||||
Die Auswahl der EKG-Ableitung, SpO₂-Signalqualität und Sensorverbindung,
|
||||
der arteriellen Kurvenform, der Kapnogrammform sowie **sämtliche NIBP-
|
||||
Messfunktionen** sind ausschließlich im **Admin-Fenster** verfügbar. Dazu
|
||||
zählen Messintervall, automatische Messung, **Measure now**, simulierter
|
||||
Messausfall und Manschettenlokalisation. Diese Parameter können über die
|
||||
Alarmmenüs am Monitor nicht verändert werden.
|
||||
der arteriellen Kurvenform und der Kapnogrammform bleiben ausschließlich im
|
||||
**Admin-Fenster**. Das **NIBP-Touchmenü** bietet zusätzlich **Measure now**,
|
||||
**Automatic measurement ON/OFF** und **Measurement interval** (1, 3, 5, 10,
|
||||
20, 30 oder 60 Minuten). Intervall und Automatik werden erst mit **Apply**
|
||||
übernommen; **Measure now** startet dagegen sofort, auch bei deaktivierter
|
||||
Automatik. Eine bereits laufende Messung wird nicht erneut gestartet.
|
||||
Simulierter Messausfall und Manschettenlokalisation bleiben Admin-only.
|
||||
|
||||
### NIBP beim Programmstart
|
||||
|
||||
Die **automatische NIBP-Messung ist beim Start ausgeschaltet**. Anfangs
|
||||
zeigt der Monitor `---/---` (MAP `---`); es findet keine Messung statt.
|
||||
Die erste Messung wird über **Measure now** im Admin-Fenster oder durch
|
||||
Aktivieren von **Automatic NIBP measurement** ausgelöst. Beim Einschalten
|
||||
Die erste Messung wird über **Measure now** im NIBP-Touchmenü oder Admin-Fenster
|
||||
oder durch Aktivieren von **Automatic NIBP measurement** ausgelöst. Beim Einschalten
|
||||
von Auto beginnt sofort ein Messzyklus; das Intervall zählt anschließend
|
||||
jeweils ab Ende der vorherigen Messung. Beim Ausschalten von Auto wird
|
||||
eine bereits laufende Messung nicht abgebrochen. **Measure now** funktioniert
|
||||
|
|
@ -83,8 +85,8 @@ rechts **+ · Wert · − · OFF/ON** angeordnet. Die Schaltfläche zum
|
|||
Deaktivieren liegt unmittelbar rechts von Minus. Größe, Gestaltung,
|
||||
Schrittweite, Wiederherstellung des letzten Werts sowie Apply/Cancel
|
||||
bleiben erhalten. Bei Platzmangel lässt sich die Matrix scrollen, während
|
||||
Apply und Cancel zugänglich bleiben. Andere Einstellungen bleiben dem
|
||||
Admin-Fenster vorbehalten.
|
||||
Apply und Cancel zugänglich bleiben. Andere Einstellungen bleiben dem Admin-Fenster vorbehalten, mit Ausnahme der
|
||||
expliziten NIBP-Messfunktionen im Touchmenü.
|
||||
|
||||
## Thoraxkompressionen (v0.5.4)
|
||||
|
||||
|
|
|
|||
|
|
@ -73,16 +73,18 @@ eine bereits laufende Messung nicht abgebrochen. **Measure now** funktioniert
|
|||
auch bei ausgeschalteter Automatik. Dann bleibt die Intervallanzeige leer
|
||||
und der Monitor zeigt `AUTO OFF`.
|
||||
|
||||
### Anordnung der ART-Alarmgrenzen (v0.6.2)
|
||||
### Horizontale Matrix für Alarmgrenzen (v0.6.3)
|
||||
|
||||
Das Touchmenü für die invasive arterielle Druckmessung ordnet **SYS, MAP und
|
||||
DIA** in drei Spalten an. Die obere Bedienzeile enthält die **oberen (↑)**
|
||||
Alarmgrenzen, die zweite die **unteren (↓)** Alarmgrenzen. **PPV** folgt unter
|
||||
SYS mit eigenen oberen und unteren Grenzwerten. Die bisherigen großen
|
||||
**+ / Wert / − / EIN-AUS**-Schaltflächen und ihre Bedienlogik bleiben erhalten.
|
||||
Die Gestaltung verwendet weiterhin Schwarz, Hellgrau und schmale Rahmenlinien.
|
||||
Alle anderen Parametermenüs bleiben unverändert. Alarmverarbeitung und
|
||||
Voreinstellungen sind von dieser Layoutänderung nicht betroffen.
|
||||
Alle fünf Touchmenüs des Monitors verwenden dieselbe kompakte Matrix.
|
||||
Jeder Parameter belegt eine Zeile (bei ART: **SYS**, **MAP**, **DIA**,
|
||||
**PPV**); die **untere (↓)** und **obere (↑)** Alarmgrenze stehen
|
||||
nebeneinander. Die Schaltflächen jedes Grenzwerts sind von links nach
|
||||
rechts **+ · Wert · − · OFF/ON** angeordnet. Die Schaltfläche zum
|
||||
Deaktivieren liegt unmittelbar rechts von Minus. Größe, Gestaltung,
|
||||
Schrittweite, Wiederherstellung des letzten Werts sowie Apply/Cancel
|
||||
bleiben erhalten. Bei Platzmangel lässt sich die Matrix scrollen, während
|
||||
Apply und Cancel zugänglich bleiben. Andere Einstellungen bleiben dem
|
||||
Admin-Fenster vorbehalten.
|
||||
|
||||
## Thoraxkompressionen (v0.5.4)
|
||||
|
||||
|
|
|
|||
|
|
@ -1,8 +1,22 @@
|
|||
--- README.de.md
|
||||
+++ README.de.md
|
||||
@@ -1,4 +1,4 @@
|
||||
-# Patient Monitor Simulation (PaMSim) — v0.6.2
|
||||
+# Patient Monitor Simulation (PaMSim) — v0.6.3
|
||||
-# Patient Monitor Simulation (PaMSim) — v0.6.3
|
||||
+# Patient Monitor Simulation (PaMSim) — v0.6.4
|
||||
|
||||
Ein mit Rust und GTK4 entwickelter Patient*innenmonitor-Simulator mit zwei Fenstern für die Lehre unter Fedora Linux.
|
||||
|
||||
@@ -609,3 +611,13 @@
|
||||
Kurvenformen sowie NIBP-Messfunktionen wurden aus den Monitor-Menüs entfernt
|
||||
und bleiben im Admin-Fenster verfügbar. Die Simulationslogik, Alarmprüfung
|
||||
und das NIBP-Startverhalten wurden nicht verändert.
|
||||
+
|
||||
+### v0.6.4 – NIBP-Touchbedienung wiederhergestellt
|
||||
+
|
||||
+Das NIBP-Menü enthält wieder **Measure now**, **Automatic measurement ON/OFF**
|
||||
+und das **Messintervall** (1, 3, 5, 10, 20, 30 oder 60 Minuten) oberhalb der
|
||||
+SYS/MAP/DIA-Alarmmatrix. Die Einstellungen für Intervall und Automatik werden
|
||||
+mit **Apply** gespeichert; **Measure now** startet unmittelbar einen Messzyklus,
|
||||
+auch wenn die Automatik ausgeschaltet ist. Die Admin-Oberfläche wird
|
||||
+synchronisiert. Die anderen Parametermenüs bleiben auf Alarmgrenzen beschränkt.
|
||||
+Manschettenzuordnung und simulierter Messausfall verbleiben im Admin-Fenster.
|
||||
|
|
|
|||
|
|
@ -1,4 +1,4 @@
|
|||
# Patient Monitor Simulation (PaMSim) — v0.6.3
|
||||
# Patient Monitor Simulation (PaMSim) — v0.6.4
|
||||
|
||||
A Rust/GTK4 patient-monitor simulator with two main windows for teaching on Fedora Linux.
|
||||
|
||||
|
|
@ -23,7 +23,7 @@ The NIBP visibility toggle hides **only the fourth group's content**; the
|
|||
bottom measurement row and other three groups remain present. Disabling one
|
||||
of the four waveform channels still reflows the remaining rows above it.
|
||||
|
||||
## Monitor touchscreen: alarm limits only (v0.6.1)
|
||||
## Monitor touchscreen: alarm limits and NIBP controls (v0.6.4)
|
||||
|
||||
Tap or click the **waveform or numerical readout** of ECG, SpO₂, ART or CO₂
|
||||
to open that channel's alarm-limit panel. For NIBP, tap the **rightmost
|
||||
|
|
@ -32,7 +32,7 @@ are reserved for future measurements and remain non-interactive. The panels
|
|||
are available in fullscreen and retain a **black background, very light-grey
|
||||
text, and 0.8-pixel borders and dividers**, in the style of the main monitor.
|
||||
|
||||
**Only alarm thresholds can be modified from the monitor-side menus.**
|
||||
**Alarm thresholds can be modified from every monitor-side menu.**
|
||||
No keyboard is required: every lower/upper limit has large **− / +** controls
|
||||
with a parameter-specific increment and an independent **ON/OFF** control.
|
||||
Unset limits remain disabled. The lower limit cannot exceed the upper limit.
|
||||
|
|
@ -49,18 +49,20 @@ Unset limits remain disabled. The lower limit cannot exceed the upper limit.
|
|||
Changing a threshold sets the alarm preset to **Custom**. **Cancel** or
|
||||
**Esc** discards the draft changes.
|
||||
|
||||
ECG lead selection, SpO₂ signal quality and sensor status, ART waveform shape,
|
||||
capnogram shape, and all NIBP measurement controls are **Admin-only**. This
|
||||
includes NIBP measurement interval, Auto ON/OFF, **Measure now**, simulated
|
||||
failed reading, and cuff placement. None of these physiological or technical
|
||||
parameters is changed by the monitor-side alarm menus.
|
||||
**NIBP is the exception:** its monitor-side menu additionally provides **Measure now**,
|
||||
**Automatic measurement ON/OFF**, and the **1/3/5/10/20/30/60-minute interval**.
|
||||
Interval and Auto changes take effect with **Apply** (or are discarded with
|
||||
Cancel/Esc); **Measure now starts immediately**, even with Auto OFF and before
|
||||
Apply. The touch controls synchronize with Admin. A measurement already in
|
||||
progress is not restarted. The simulated failure option and cuff placement
|
||||
remain Admin-only, as do ECG lead, SpO₂ sensor/quality and waveform choices.
|
||||
|
||||
### NIBP startup and automatic measurements
|
||||
|
||||
**Auto measurement is OFF by default.** On startup the monitor shows
|
||||
`---/---` (MAP `---`) and no measurement is initiated. To obtain a reading,
|
||||
select **Measure now** in Admin, or enable **Automatic NIBP measurement**
|
||||
there. The first automatic cycle begins immediately after enabling Auto;
|
||||
select **Measure now** in the NIBP monitor menu (or in Admin), or enable
|
||||
**Automatic NIBP measurement** via the menu (or Admin). The first automatic cycle begins immediately after enabling Auto;
|
||||
subsequent measurement intervals count from completion of the prior cycle.
|
||||
Disabling Auto stops further scheduled measurements without interrupting a
|
||||
cycle already in progress. Manual measurement remains possible while Auto
|
||||
|
|
@ -77,7 +79,8 @@ OFF/ON** (the OFF/ON button is immediately right of minus). The original
|
|||
large button styling, per-parameter increments, last-value restoration,
|
||||
optional bounds and Apply/Cancel semantics are unchanged. The matrix scrolls
|
||||
if necessary while Apply and Cancel remain accessible below it. Other
|
||||
physiological and NIBP controls remain Admin-only.
|
||||
physiological controls remain Admin-only, except for the three explicitly
|
||||
requested NIBP measurement controls.
|
||||
|
||||
## Chest compressions (v0.5.4)
|
||||
|
||||
|
|
@ -620,3 +623,14 @@ alarm limits. Removed ECG lead, signal/sensor, waveform and all NIBP
|
|||
measurement controls from monitor-side menus. These remain available in
|
||||
Admin, including manual NIBP measurement and Auto ON/OFF. No physiological
|
||||
model, alarm evaluation or NIBP startup behavior was changed.
|
||||
|
||||
### v0.6.4 — NIBP touch controls restored
|
||||
|
||||
The monitor-side NIBP settings menu again contains **Measure now**,
|
||||
**Automatic measurement ON/OFF** and the **measurement interval** (1, 3, 5,
|
||||
10, 20, 30, 60 minutes). These controls are above the horizontal SYS/MAP/DIA
|
||||
alarm matrix and share its monochrome touchscreen styling. Auto and interval
|
||||
are saved with Apply; Measure now starts a cuff cycle immediately regardless
|
||||
of Auto. Both monitor and Admin display the selected settings. Other monitor
|
||||
menus remain limited to alarm thresholds; NIBP cuff placement, the simulated
|
||||
failed-reading option and non-NIBP signal settings remain Admin-only.
|
||||
|
|
|
|||
|
|
@ -279,6 +279,10 @@ impl Simulator {
|
|||
self.spo2_quality(p))
|
||||
}
|
||||
|
||||
pub fn nibp_measuring(&self) -> bool {
|
||||
self.nibp.measuring()
|
||||
}
|
||||
|
||||
pub fn start_nibp_now(&mut self, p: &Parameters) {
|
||||
let quality = self.nibp_start_quality(p);
|
||||
self.nibp.measure_now(self.time, p, quality);
|
||||
|
|
|
|||
|
|
@ -108,8 +108,8 @@ fn create_monitor(app: &Application, params: SharedParams,
|
|||
});
|
||||
window.add_controller(keys);
|
||||
|
||||
// One gesture opens alarm-only panels for visible waveform/numeric rows
|
||||
// and NIBP's bottom-right cell. Mouse and Wayland touch share geometry.
|
||||
// Mouse and Wayland touch open the visible module's settings panel.
|
||||
// NIBP alone also includes its explicitly requested measurement controls.
|
||||
let monitor_menu: MonitorMenuHandle = Rc::new(RefCell::new(None));
|
||||
let click = gtk::GestureClick::new();
|
||||
click.connect_released({
|
||||
|
|
@ -117,13 +117,14 @@ fn create_monitor(app: &Application, params: SharedParams,
|
|||
let window = window.clone();
|
||||
let canvas = canvas.clone();
|
||||
let params = params.clone();
|
||||
let simulator = simulator.clone();
|
||||
move |_gesture, _press_count, x, y| {
|
||||
let touched = {
|
||||
let p = params.borrow();
|
||||
render::monitor_hit_test(x, y, canvas.width(), canvas.height(), &p)
|
||||
};
|
||||
if let Some(channel) = touched {
|
||||
open_module_menu(&app, &window, params.clone(),
|
||||
open_module_menu(&app, &window, params.clone(), simulator.clone(),
|
||||
admin_changed.clone(), monitor_menu.clone(), channel);
|
||||
}
|
||||
}
|
||||
|
|
@ -133,11 +134,25 @@ fn create_monitor(app: &Application, params: SharedParams,
|
|||
window
|
||||
}
|
||||
|
||||
/// Map an Admin NIBP measurement interval to its dropdown selection.
|
||||
/// The touchscreen alarm menus do not expose scheduling controls.
|
||||
/// Both Admin and the monitor NIBP menu use the same interval options.
|
||||
const NIBP_INTERVALS: [i32; 7] = [1, 3, 5, 10, 20, 30, 60];
|
||||
|
||||
fn nibp_interval_index(minutes: i32) -> u32 {
|
||||
const INTERVALS: [i32; 7] = [1, 3, 5, 10, 20, 30, 60];
|
||||
INTERVALS.iter().position(|&value| value == minutes).unwrap_or(2) as u32
|
||||
NIBP_INTERVALS.iter().position(|&value| value == minutes).unwrap_or(2) as u32
|
||||
}
|
||||
|
||||
fn nibp_interval_minutes(selected: u32) -> i32 {
|
||||
NIBP_INTERVALS.get(selected as usize).copied().unwrap_or(5)
|
||||
}
|
||||
|
||||
/// Apply only the explicitly requested monitor-side NIBP measurement settings.
|
||||
/// Neither cuff settings nor any other module's configuration is changed.
|
||||
fn apply_nibp_measurement_settings(p: &mut Parameters, selected: u32, auto: bool) -> bool {
|
||||
let minutes = nibp_interval_minutes(selected);
|
||||
let changed = p.nibp_interval_minutes != minutes || p.nibp_auto_measurement != auto;
|
||||
p.nibp_interval_minutes = minutes;
|
||||
p.nibp_auto_measurement = auto;
|
||||
changed
|
||||
}
|
||||
|
||||
fn create_admin(app: &Application, params: SharedParams,
|
||||
|
|
@ -437,10 +452,9 @@ ART and pleth are delayed. VT can be perfusing; PVT, VF and asystole are pulsele
|
|||
|
||||
let nibp_interval_dropdown = add_dropdown_at(&nibp, "Measurement interval",
|
||||
&["1 min", "3 min", "5 min", "10 min", "20 min", "30 min", "60 min"],
|
||||
2, params.clone(), |p, selected| {
|
||||
p.nibp_interval_minutes = match selected {
|
||||
0 => 1, 1 => 3, 2 => 5, 3 => 10, 4 => 20, 5 => 30, _ => 60,
|
||||
};
|
||||
nibp_interval_index(params.borrow().nibp_interval_minutes),
|
||||
params.clone(), |p, selected| {
|
||||
p.nibp_interval_minutes = nibp_interval_minutes(selected);
|
||||
});
|
||||
let nibp_auto = gtk::CheckButton::with_label("Automatic NIBP measurement");
|
||||
nibp_auto.set_active(params.borrow().nibp_auto_measurement);
|
||||
|
|
@ -602,6 +616,10 @@ fn install_ecg_menu_style() {
|
|||
min-height: 48px; min-width: 52px;
|
||||
}
|
||||
window.ecg-settings button:hover { background-color: #303437; }
|
||||
window.ecg-settings dropdown {
|
||||
color: #e5e9eb; background-color: #171717;
|
||||
border: 1px solid #6b757d; min-height: 48px;
|
||||
}
|
||||
window.ecg-settings button.ecg-apply {
|
||||
color: #000000; background-color: #e5e9eb; font-weight: bold;
|
||||
}
|
||||
|
|
@ -866,8 +884,85 @@ fn alarm_limit_matrix(specs: Vec<TouchAlarmSpec>,
|
|||
frame
|
||||
}
|
||||
|
||||
/// A monitor-side menu may change only alarm limits. All patient and
|
||||
/// measurement configuration (including NIBP scheduling) stays Admin-only.
|
||||
/// Add the three explicitly requested NIBP actions to the monitor menu.
|
||||
/// Alarm limits, interval and Auto are staged until Apply. "Measure now" is
|
||||
/// an immediate command, independent of Auto and of the staged settings.
|
||||
fn nibp_touch_measurement_controls(
|
||||
current: &Parameters,
|
||||
params: SharedParams,
|
||||
simulator: Rc<RefCell<Simulator>>,
|
||||
) -> (gtk::Box, gtk::DropDown, gtk::ToggleButton) {
|
||||
let section = gtk::Box::new(Orientation::Vertical, 12);
|
||||
section.add_css_class("alarm-limit-frame");
|
||||
section.set_hexpand(true);
|
||||
|
||||
let heading = gtk::Label::new(Some("NIBP measurement"));
|
||||
heading.add_css_class("alarm-limit-heading");
|
||||
heading.set_xalign(0.0);
|
||||
section.append(&heading);
|
||||
|
||||
let row = gtk::Box::new(Orientation::Horizontal, 12);
|
||||
let label = gtk::Label::new(Some("Measurement interval"));
|
||||
label.set_hexpand(true);
|
||||
label.set_xalign(0.0);
|
||||
let interval = gtk::DropDown::from_strings(&[
|
||||
"1 min", "3 min", "5 min", "10 min", "20 min", "30 min", "60 min",
|
||||
]);
|
||||
interval.set_selected(nibp_interval_index(current.nibp_interval_minutes));
|
||||
interval.set_size_request(160, 52);
|
||||
row.append(&label);
|
||||
row.append(&interval);
|
||||
section.append(&row);
|
||||
|
||||
let action_row = gtk::Box::new(Orientation::Horizontal, 12);
|
||||
action_row.set_homogeneous(true);
|
||||
let auto = gtk::ToggleButton::new();
|
||||
auto.set_active(current.nibp_auto_measurement);
|
||||
auto.set_label(if auto.is_active() {
|
||||
"Automatic measurement: ON"
|
||||
} else {
|
||||
"Automatic measurement: OFF"
|
||||
});
|
||||
auto.connect_toggled(|button| {
|
||||
button.set_label(if button.is_active() {
|
||||
"Automatic measurement: ON"
|
||||
} else {
|
||||
"Automatic measurement: OFF"
|
||||
});
|
||||
});
|
||||
action_row.append(&auto);
|
||||
|
||||
let measure_now = gtk::Button::with_label("Measure now");
|
||||
measure_now.set_tooltip_text(Some("Start an NIBP cuff measurement immediately"));
|
||||
action_row.append(&measure_now);
|
||||
section.append(&action_row);
|
||||
|
||||
let feedback = gtk::Label::new(Some(
|
||||
"Measure now acts immediately. Interval and Auto are saved with Apply."
|
||||
));
|
||||
feedback.set_xalign(0.0);
|
||||
feedback.set_wrap(true);
|
||||
measure_now.connect_clicked({
|
||||
let feedback = feedback.clone();
|
||||
move |_| {
|
||||
// Drop the immutable Ref before borrowing the simulator mutably.
|
||||
let measuring = { simulator.borrow().nibp_measuring() };
|
||||
if measuring {
|
||||
feedback.set_text("NIBP measurement already in progress.");
|
||||
} else {
|
||||
simulator.borrow_mut().start_nibp_now(¶ms.borrow());
|
||||
feedback.set_text("NIBP measurement started.");
|
||||
}
|
||||
}
|
||||
});
|
||||
// Keep the explanatory label visible while the button changes its status.
|
||||
// It is owned by the section, not by the button callback.
|
||||
section.append(&feedback);
|
||||
(section, interval, auto)
|
||||
}
|
||||
|
||||
/// All monitor modules can edit alarm limits. Only NIBP additionally offers
|
||||
/// the explicitly requested interval, Auto and immediate measurement controls.
|
||||
fn apply_module_alarm_changes(
|
||||
p: &mut Parameters,
|
||||
limits: &[(fn(&mut AlarmSettings, AlarmLimits), AlarmLimits)],
|
||||
|
|
@ -879,12 +974,13 @@ fn apply_module_alarm_changes(
|
|||
p.alarms != previous
|
||||
}
|
||||
|
||||
/// Open a modal alarm-limit panel with the same black, grey, 0.8 px divider
|
||||
/// and border design as the ECG panel. Other parameters are Admin-only.
|
||||
/// Open a modal monitor settings panel in neutral black/grey styling.
|
||||
/// All modules edit alarm limits; NIBP also retains its explicit controls.
|
||||
#[allow(clippy::too_many_arguments)]
|
||||
fn open_module_menu(app: &Application, monitor: &ApplicationWindow,
|
||||
params: SharedParams, on_change: AdminChanged,
|
||||
open_menu: MonitorMenuHandle, channel: render::TouchChannel) {
|
||||
params: SharedParams, simulator: Rc<RefCell<Simulator>>,
|
||||
on_change: AdminChanged, open_menu: MonitorMenuHandle,
|
||||
channel: render::TouchChannel) {
|
||||
if let Some(existing) = open_menu.borrow().as_ref().and_then(|weak| weak.upgrade()) {
|
||||
existing.present();
|
||||
return;
|
||||
|
|
@ -906,7 +1002,7 @@ fn open_module_menu(app: &Application, monitor: &ApplicationWindow,
|
|||
.default_width(800)
|
||||
.default_height(match channel {
|
||||
render::TouchChannel::Art => 560,
|
||||
render::TouchChannel::Nibp => 490,
|
||||
render::TouchChannel::Nibp => 640,
|
||||
render::TouchChannel::Ecg => 310,
|
||||
_ => 420,
|
||||
})
|
||||
|
|
@ -938,6 +1034,18 @@ fn open_module_menu(app: &Application, monitor: &ApplicationWindow,
|
|||
inner.append(&heading);
|
||||
inner.append(&ecg_menu_divider());
|
||||
|
||||
// NIBP is the explicit exception to the alarm-only monitor menu policy.
|
||||
// Put the manual measurement action above the limit matrix for easy access.
|
||||
let nibp_controls = if channel == render::TouchChannel::Nibp {
|
||||
let (section, interval, auto) = nibp_touch_measurement_controls(
|
||||
¤t, params.clone(), simulator);
|
||||
inner.append(§ion);
|
||||
inner.append(&ecg_menu_divider());
|
||||
Some((interval, auto))
|
||||
} else {
|
||||
None
|
||||
};
|
||||
|
||||
// A shared horizontal matrix is used in every monitor-side touch menu.
|
||||
let specs = module_alarm_specs(channel, ¤t);
|
||||
let mut alarm_drafts = Vec::new();
|
||||
|
|
@ -983,7 +1091,12 @@ fn open_module_menu(app: &Application, monitor: &ApplicationWindow,
|
|||
}
|
||||
let changed_alarms = {
|
||||
let mut p = params.borrow_mut();
|
||||
apply_module_alarm_changes(&mut p, &selected_limits)
|
||||
let changed = apply_module_alarm_changes(&mut p, &selected_limits);
|
||||
if let Some((interval, auto)) = &nibp_controls {
|
||||
apply_nibp_measurement_settings(&mut p,
|
||||
interval.selected(), auto.is_active());
|
||||
}
|
||||
changed
|
||||
};
|
||||
if let Some(callback) = on_change.borrow().as_ref() {
|
||||
callback(changed_alarms);
|
||||
|
|
@ -1064,7 +1177,7 @@ mod module_touch_menu_tests {
|
|||
}
|
||||
|
||||
#[test]
|
||||
fn touchscreen_applies_only_alarm_limits_and_preserves_other_controls() {
|
||||
fn monitor_alarm_edits_preserve_other_controls() {
|
||||
let mut p = Parameters::default();
|
||||
p.ecg_lead = EcgLead::V1;
|
||||
p.spo2_quality = Spo2Quality::Poor;
|
||||
|
|
@ -1093,6 +1206,40 @@ mod module_touch_menu_tests {
|
|||
assert!(p.nibp_cuff_on_art);
|
||||
}
|
||||
|
||||
#[test]
|
||||
fn nibp_touch_interval_roundtrip_and_startup_off() {
|
||||
let p = Parameters::default();
|
||||
assert!(!p.nibp_auto_measurement);
|
||||
assert_eq!(p.nibp_interval_minutes, 5);
|
||||
for (index, minutes) in NIBP_INTERVALS.into_iter().enumerate() {
|
||||
assert_eq!(nibp_interval_index(minutes), index as u32);
|
||||
assert_eq!(nibp_interval_minutes(index as u32), minutes);
|
||||
}
|
||||
assert_eq!(nibp_interval_minutes(99), 5);
|
||||
}
|
||||
|
||||
#[test]
|
||||
fn nibp_touch_apply_changes_only_interval_and_auto() {
|
||||
let mut p = Parameters::default();
|
||||
p.nibp_no_reading = true;
|
||||
p.nibp_cuff_on_spo2 = true;
|
||||
p.nibp_cuff_on_art = true;
|
||||
p.show_nibp = false;
|
||||
p.alarms.nibp_sys.lower = Some(80);
|
||||
assert!(apply_nibp_measurement_settings(&mut p, 5, true));
|
||||
assert_eq!(p.nibp_interval_minutes, 30);
|
||||
assert!(p.nibp_auto_measurement);
|
||||
assert!(!apply_nibp_measurement_settings(&mut p, 5, true));
|
||||
assert!(apply_nibp_measurement_settings(&mut p, 0, false));
|
||||
assert_eq!(p.nibp_interval_minutes, 1);
|
||||
assert!(!p.nibp_auto_measurement);
|
||||
assert!(p.nibp_no_reading);
|
||||
assert!(p.nibp_cuff_on_spo2);
|
||||
assert!(p.nibp_cuff_on_art);
|
||||
assert!(!p.show_nibp);
|
||||
assert_eq!(p.alarms.nibp_sys.lower, Some(80));
|
||||
}
|
||||
|
||||
}
|
||||
|
||||
/// Frame a parameter section with an optional monitor-colored left sidebar.
|
||||
|
|
|
|||
Loading…
Reference in a new issue