v0.6.4 — NIBP touch controls restored
The monitor-side NIBP settings menu again contains **Measure now**, **Automatic measurement ON/OFF** and the **measurement interval** (1, 3, 5, 10, 20, 30, 60 minutes). These controls are above the horizontal SYS/MAP/DIA alarm matrix and share its monochrome touchscreen styling. Auto and interval are saved with Apply; Measure now starts a cuff cycle immediately regardless of Auto. Both monitor and Admin display the selected settings. Other monitor menus remain limited to alarm thresholds; NIBP cuff placement, the simulated failed-reading option and non-NIBP signal settings remain Admin-only.
This commit is contained in:
parent
54012ab960
commit
00dc22b1aa
9 changed files with 702 additions and 56 deletions
|
|
@ -1,4 +1,4 @@
|
|||
# Patient Monitor Simulation (PaMSim) — v0.6.3
|
||||
# Patient Monitor Simulation (PaMSim) — v0.6.4
|
||||
|
||||
A Rust/GTK4 patient-monitor simulator with two main windows for teaching on Fedora Linux.
|
||||
|
||||
|
|
@ -23,7 +23,7 @@ The NIBP visibility toggle hides **only the fourth group's content**; the
|
|||
bottom measurement row and other three groups remain present. Disabling one
|
||||
of the four waveform channels still reflows the remaining rows above it.
|
||||
|
||||
## Monitor touchscreen: alarm limits only (v0.6.1)
|
||||
## Monitor touchscreen: alarm limits and NIBP controls (v0.6.4)
|
||||
|
||||
Tap or click the **waveform or numerical readout** of ECG, SpO₂, ART or CO₂
|
||||
to open that channel's alarm-limit panel. For NIBP, tap the **rightmost
|
||||
|
|
@ -32,7 +32,7 @@ are reserved for future measurements and remain non-interactive. The panels
|
|||
are available in fullscreen and retain a **black background, very light-grey
|
||||
text, and 0.8-pixel borders and dividers**, in the style of the main monitor.
|
||||
|
||||
**Only alarm thresholds can be modified from the monitor-side menus.**
|
||||
**Alarm thresholds can be modified from every monitor-side menu.**
|
||||
No keyboard is required: every lower/upper limit has large **− / +** controls
|
||||
with a parameter-specific increment and an independent **ON/OFF** control.
|
||||
Unset limits remain disabled. The lower limit cannot exceed the upper limit.
|
||||
|
|
@ -49,18 +49,20 @@ Unset limits remain disabled. The lower limit cannot exceed the upper limit.
|
|||
Changing a threshold sets the alarm preset to **Custom**. **Cancel** or
|
||||
**Esc** discards the draft changes.
|
||||
|
||||
ECG lead selection, SpO₂ signal quality and sensor status, ART waveform shape,
|
||||
capnogram shape, and all NIBP measurement controls are **Admin-only**. This
|
||||
includes NIBP measurement interval, Auto ON/OFF, **Measure now**, simulated
|
||||
failed reading, and cuff placement. None of these physiological or technical
|
||||
parameters is changed by the monitor-side alarm menus.
|
||||
**NIBP is the exception:** its monitor-side menu additionally provides **Measure now**,
|
||||
**Automatic measurement ON/OFF**, and the **1/3/5/10/20/30/60-minute interval**.
|
||||
Interval and Auto changes take effect with **Apply** (or are discarded with
|
||||
Cancel/Esc); **Measure now starts immediately**, even with Auto OFF and before
|
||||
Apply. The touch controls synchronize with Admin. A measurement already in
|
||||
progress is not restarted. The simulated failure option and cuff placement
|
||||
remain Admin-only, as do ECG lead, SpO₂ sensor/quality and waveform choices.
|
||||
|
||||
### NIBP startup and automatic measurements
|
||||
|
||||
**Auto measurement is OFF by default.** On startup the monitor shows
|
||||
`---/---` (MAP `---`) and no measurement is initiated. To obtain a reading,
|
||||
select **Measure now** in Admin, or enable **Automatic NIBP measurement**
|
||||
there. The first automatic cycle begins immediately after enabling Auto;
|
||||
select **Measure now** in the NIBP monitor menu (or in Admin), or enable
|
||||
**Automatic NIBP measurement** via the menu (or Admin). The first automatic cycle begins immediately after enabling Auto;
|
||||
subsequent measurement intervals count from completion of the prior cycle.
|
||||
Disabling Auto stops further scheduled measurements without interrupting a
|
||||
cycle already in progress. Manual measurement remains possible while Auto
|
||||
|
|
@ -77,7 +79,8 @@ OFF/ON** (the OFF/ON button is immediately right of minus). The original
|
|||
large button styling, per-parameter increments, last-value restoration,
|
||||
optional bounds and Apply/Cancel semantics are unchanged. The matrix scrolls
|
||||
if necessary while Apply and Cancel remain accessible below it. Other
|
||||
physiological and NIBP controls remain Admin-only.
|
||||
physiological controls remain Admin-only, except for the three explicitly
|
||||
requested NIBP measurement controls.
|
||||
|
||||
## Chest compressions (v0.5.4)
|
||||
|
||||
|
|
@ -620,3 +623,14 @@ alarm limits. Removed ECG lead, signal/sensor, waveform and all NIBP
|
|||
measurement controls from monitor-side menus. These remain available in
|
||||
Admin, including manual NIBP measurement and Auto ON/OFF. No physiological
|
||||
model, alarm evaluation or NIBP startup behavior was changed.
|
||||
|
||||
### v0.6.4 — NIBP touch controls restored
|
||||
|
||||
The monitor-side NIBP settings menu again contains **Measure now**,
|
||||
**Automatic measurement ON/OFF** and the **measurement interval** (1, 3, 5,
|
||||
10, 20, 30, 60 minutes). These controls are above the horizontal SYS/MAP/DIA
|
||||
alarm matrix and share its monochrome touchscreen styling. Auto and interval
|
||||
are saved with Apply; Measure now starts a cuff cycle immediately regardless
|
||||
of Auto. Both monitor and Admin display the selected settings. Other monitor
|
||||
menus remain limited to alarm thresholds; NIBP cuff placement, the simulated
|
||||
failed-reading option and non-NIBP signal settings remain Admin-only.
|
||||
|
|
|
|||