v0.6.0 — Additional touchscreen modules and NIBP manual-first startup
Added touch/click settings panels for SpO₂, ART, CO₂ and the bottom-right NIBP cell. Their threshold editors use large, unit-specific +/− and ON/OFF controls; the panels expose relevant waveform/sensor options and stay synchronized with Admin. NIBP additionally provides interval selection, manual **Measure now**, Auto ON/OFF and cuff options. Startup no longer initiates a cuff cycle or shows any NIBP reading; Auto is OFF by default. Existing NIBP cycle physics and signal quality behavior are unchanged.
This commit is contained in:
parent
3f191e4c46
commit
41e1971fae
13 changed files with 1954 additions and 93 deletions
|
|
@ -35,6 +35,7 @@ struct Cycle {
|
|||
pub struct NibpMachine {
|
||||
cycle: Option<Cycle>,
|
||||
ever_started: bool,
|
||||
auto_enabled_on_last_update: bool,
|
||||
ever_successful: bool,
|
||||
last_completed: Option<f64>,
|
||||
last_reading: Option<NibpReading>,
|
||||
|
|
@ -47,6 +48,7 @@ impl NibpMachine {
|
|||
Self {
|
||||
cycle: None,
|
||||
ever_started: false,
|
||||
auto_enabled_on_last_update: false,
|
||||
ever_successful: false,
|
||||
last_completed: None,
|
||||
last_reading: None,
|
||||
|
|
@ -146,17 +148,22 @@ impl NibpMachine {
|
|||
self.ever_started = true;
|
||||
}
|
||||
|
||||
/// The initial cycle starts at simulation time zero. Afterwards, the
|
||||
/// interval is counted from the last completed measurement, not cycle start.
|
||||
/// Automatic cycles only start if the instructor enabled Auto measurement.
|
||||
/// In-flight manual/automatic cycles complete even after Auto is turned off.
|
||||
/// When Auto is switched on, a cycle begins immediately (if idle);
|
||||
/// later cycles start at the interval after the previous completion.
|
||||
pub fn update(&mut self, now: f64, p: &Parameters, quality_at_start: Spo2Quality) {
|
||||
if now < 0.0 { return; }
|
||||
let just_enabled_auto = p.nibp_auto_measurement && !self.auto_enabled_on_last_update;
|
||||
self.auto_enabled_on_last_update = p.nibp_auto_measurement;
|
||||
if let Some(cycle) = self.cycle {
|
||||
if now - cycle.start >= cycle.duration {
|
||||
self.complete(now, p, cycle);
|
||||
}
|
||||
} else if !self.ever_started || self.last_completed.is_some_and(|last| {
|
||||
now - last >= f64::from(p.nibp_interval_minutes.max(1)) * 60.0
|
||||
}) {
|
||||
} else if p.nibp_auto_measurement &&
|
||||
(just_enabled_auto || !self.ever_started || self.last_completed.is_some_and(|last| {
|
||||
now - last >= f64::from(p.nibp_interval_minutes.max(1)) * 60.0
|
||||
})) {
|
||||
self.start(now, p, quality_at_start);
|
||||
}
|
||||
}
|
||||
|
|
@ -206,9 +213,58 @@ mod tests {
|
|||
use super::*;
|
||||
|
||||
#[test]
|
||||
fn intervals_and_first_values() {
|
||||
fn startup_is_blank_and_auto_is_disabled() {
|
||||
let p = Parameters::default();
|
||||
assert!(!p.nibp_auto_measurement);
|
||||
let mut m = NibpMachine::new();
|
||||
for t in [0.0, 20.0, 3600.0] {
|
||||
m.update(t, &p, Spo2Quality::Good);
|
||||
assert!(!m.measuring());
|
||||
assert_eq!(m.reading(), None);
|
||||
}
|
||||
// The first manual cycle completes even though automatic mode is OFF.
|
||||
m.measure_now(3601.0, &p, Spo2Quality::Good);
|
||||
assert!(m.measuring());
|
||||
m.update(3621.0, &p, Spo2Quality::Good);
|
||||
assert_eq!(m.reading(), Some(NibpReading { systolic: 118, diastolic: 74 }));
|
||||
m.update(4400.0, &p, Spo2Quality::Good);
|
||||
assert!(!m.measuring());
|
||||
// Enabling Auto starts the next cycle if the selected interval elapsed.
|
||||
let auto = Parameters { nibp_auto_measurement: true, ..p };
|
||||
m.update(4400.01, &auto, Spo2Quality::Good);
|
||||
assert!(m.measuring());
|
||||
// Disabling Auto again doesn't cancel this running cycle.
|
||||
m.update(4420.01, &Parameters::default(), Spo2Quality::Good);
|
||||
assert!(!m.measuring());
|
||||
}
|
||||
|
||||
#[test]
|
||||
fn turning_auto_on_after_manual_measurement_starts_new_cycle() {
|
||||
let mut m = NibpMachine::new();
|
||||
let p = Parameters::default();
|
||||
m.measure_now(0.0, &p, Spo2Quality::Good);
|
||||
m.update(20.0, &p, Spo2Quality::Good);
|
||||
assert!(m.reading().is_some());
|
||||
assert!(!m.measuring());
|
||||
// No need to wait five minutes after a manual measurement to enable Auto.
|
||||
let auto = Parameters { nibp_auto_measurement: true, ..p };
|
||||
m.update(21.0, &auto, Spo2Quality::Good);
|
||||
assert!(m.measuring());
|
||||
}
|
||||
|
||||
#[test]
|
||||
fn enabling_auto_starts_first_measurement_without_preexisting_reading() {
|
||||
let mut m = NibpMachine::new();
|
||||
let p = Parameters { nibp_auto_measurement: true, ..Parameters::default() };
|
||||
m.update(10.0, &p, Spo2Quality::Good);
|
||||
assert!(m.measuring());
|
||||
assert_eq!(m.reading(), None);
|
||||
}
|
||||
|
||||
#[test]
|
||||
fn intervals_and_first_values() {
|
||||
let mut m = NibpMachine::new();
|
||||
let p = Parameters { nibp_auto_measurement: true, ..Parameters::default() };
|
||||
m.update(0.0, &p, Spo2Quality::Good);
|
||||
assert!(m.measuring());
|
||||
assert_eq!(m.reading(), None);
|
||||
|
|
@ -233,7 +289,8 @@ mod tests {
|
|||
|
||||
#[test]
|
||||
fn failed_measurement_and_retry_cycle() {
|
||||
let p = Parameters { nibp_no_reading: true, ..Parameters::default() };
|
||||
let p = Parameters { nibp_no_reading: true, nibp_auto_measurement: true,
|
||||
..Parameters::default() };
|
||||
let mut m = NibpMachine::new();
|
||||
m.update(0.0, &p, Spo2Quality::Good);
|
||||
assert!(m.measuring());
|
||||
|
|
@ -246,7 +303,8 @@ mod tests {
|
|||
|
||||
#[test]
|
||||
fn cuff_interference_never_improves_quality() {
|
||||
let p = Parameters { nibp_cuff_on_spo2: true, ..Parameters::default() };
|
||||
let p = Parameters { nibp_cuff_on_spo2: true, nibp_auto_measurement: true,
|
||||
..Parameters::default() };
|
||||
let mut m = NibpMachine::new();
|
||||
m.update(0.0, &p, Spo2Quality::Good);
|
||||
assert_eq!(m.effective_spo2_quality(2.6, &p), Spo2Quality::None);
|
||||
|
|
@ -269,7 +327,7 @@ mod more_tests {
|
|||
(Spo2Quality::None, 120.0),
|
||||
] {
|
||||
let mut machine = NibpMachine::new();
|
||||
let p = Parameters::default();
|
||||
let p = Parameters { nibp_auto_measurement: true, ..Parameters::default() };
|
||||
machine.update(0.0, &p, quality);
|
||||
machine.update(seconds - 0.01, &p, quality);
|
||||
assert!(machine.measuring());
|
||||
|
|
@ -278,14 +336,15 @@ mod more_tests {
|
|||
assert_eq!(machine.reading().is_some(), quality != Spo2Quality::None);
|
||||
}
|
||||
let mut machine = NibpMachine::new();
|
||||
let p = Parameters::default();
|
||||
let p = Parameters { nibp_auto_measurement: true, ..Parameters::default() };
|
||||
machine.update(0.0, &p, Spo2Quality::Poor);
|
||||
assert!((60.0..=120.0).contains(&machine.cycle.unwrap().duration));
|
||||
}
|
||||
|
||||
#[test]
|
||||
fn first_successful_reading_uses_admin_even_after_failed_cycle() {
|
||||
let mut p = Parameters { nibp_no_reading: true, ..Parameters::default() };
|
||||
let mut p = Parameters { nibp_no_reading: true, nibp_auto_measurement: true,
|
||||
..Parameters::default() };
|
||||
let mut machine = NibpMachine::new();
|
||||
machine.update(0.0, &p, Spo2Quality::None);
|
||||
machine.update(120.0, &p, Spo2Quality::None);
|
||||
|
|
@ -298,7 +357,8 @@ mod more_tests {
|
|||
|
||||
#[test]
|
||||
fn partial_pressure_uses_diastic_and_systolic_boundaries() {
|
||||
let p = Parameters { nibp_cuff_on_art: true, ..Parameters::default() };
|
||||
let p = Parameters { nibp_cuff_on_art: true, nibp_auto_measurement: true,
|
||||
..Parameters::default() };
|
||||
let mut machine = NibpMachine::new();
|
||||
machine.update(0.0, &p, Spo2Quality::Good);
|
||||
assert_eq!(machine.cuff_zone(2.6, &p), CuffZone::Occluded);
|
||||
|
|
|
|||
Loading…