@ -83,6 +83,29 @@ pub enum CapnoShape {
|
|||
Obstructive,
|
||||
}
|
||||
|
||||
/// Instructor-selected efficacy of manual chest compressions.
|
||||
/// This affects synthetic pressure and pleth signals, not ECG rhythm or HR.
|
||||
#[derive(Clone, Copy, Debug, PartialEq, Eq)]
|
||||
pub enum ChestCompressionQuality {
|
||||
Good,
|
||||
Adequate,
|
||||
Poor,
|
||||
Bad,
|
||||
}
|
||||
|
||||
impl ChestCompressionQuality {
|
||||
/// Approximate compression-generated systolic peak pressure (mmHg).
|
||||
/// None means that compressions generate no measurable arterial pressure.
|
||||
pub fn peak_range(self) -> Option<(f64, f64)> {
|
||||
match self {
|
||||
Self::Good => Some((81.0, 99.0)), // 90 +/- 10%
|
||||
Self::Adequate => Some((63.0, 77.0)), // 70 +/- 10%
|
||||
Self::Poor => Some((25.0, 55.0)), // 50 -50% / +10%
|
||||
Self::Bad => None,
|
||||
}
|
||||
}
|
||||
}
|
||||
|
||||
/// Instructor-selected plethysmography signal quality; separate from the
|
||||
/// physiological oxygen-saturation setting and mechanical pulse capture.
|
||||
#[derive(Clone, Copy, Debug, PartialEq, Eq)]
|
||||
|
|
@ -202,6 +225,8 @@ pub struct Parameters {
|
|||
pub patient_name: String, // displayed in the monitor header, blank = generic title
|
||||
pub alarms: AlarmSettings,
|
||||
pub audible_alarms: bool, // mute only the sound, never the visual alarms
|
||||
pub chest_compressions_active: bool,
|
||||
pub chest_compression_quality: ChestCompressionQuality,
|
||||
pub ecg_rhythm: EcgRhythm,
|
||||
pub heart_rate: i32, // bpm, organized electrical rate; not mechanical pulse rate
|
||||
pub pulse_conduction: PulseConduction,
|
||||
|
|
@ -277,6 +302,8 @@ impl Default for Parameters {
|
|||
patient_name: String::new(),
|
||||
alarms: AlarmSettings::default(),
|
||||
audible_alarms: true,
|
||||
chest_compressions_active: false,
|
||||
chest_compression_quality: ChestCompressionQuality::Adequate,
|
||||
ecg_rhythm: EcgRhythm::Sinus,
|
||||
heart_rate: 75,
|
||||
pulse_conduction: PulseConduction::OneToOne,
|
||||
|
|
@ -313,6 +340,17 @@ pub fn mean_arterial_pressure(systolic: i32, diastolic: i32) -> i32 {
|
|||
mod tests {
|
||||
use super::*;
|
||||
|
||||
#[test]
|
||||
fn compression_quality_defaults_to_adequate_and_has_requested_pressure_ranges() {
|
||||
let p = Parameters::default();
|
||||
assert!(!p.chest_compressions_active);
|
||||
assert_eq!(p.chest_compression_quality, ChestCompressionQuality::Adequate);
|
||||
assert_eq!(ChestCompressionQuality::Good.peak_range(), Some((81.0, 99.0)));
|
||||
assert_eq!(ChestCompressionQuality::Adequate.peak_range(), Some((63.0, 77.0)));
|
||||
assert_eq!(ChestCompressionQuality::Poor.peak_range(), Some((25.0, 55.0)));
|
||||
assert_eq!(ChestCompressionQuality::Bad.peak_range(), None);
|
||||
}
|
||||
|
||||
#[test]
|
||||
fn mean_pressure_uses_one_third_pulse_pressure() {
|
||||
assert_eq!(mean_arterial_pressure(120, 75), 90);
|
||||
|
|
|
|||