# Patient Monitor Simulation (PaMSim) — v0.4.0
# Patient Monitor Simulation (PaMSim) — v0.4.1
A two-window, Rust/GTK4 patient-monitor simulator for teaching on Fedora Linux.
@ -40,11 +40,14 @@ The monitor can be resized and toggled fullscreen with **F11**; **Esc** exits
fullscreen. On Wayland, move the monitor to the presentation screen before
activating fullscreen. Fullscreen does not lock down the computer.
## Numerical alarm limits (new in v0.4.0)
## Numerical alarm limits (v0.4.0; layout and presets updated in v0.4.1)
Each displayed numeric measurement has an **independent optional lower (↓) and
upper (↑) alarm threshold** in the corresponding Admin section. Alarm fields
are text inputs; leave a field **blank** (displayed as `off`) to disable just
upper (↑) alarm threshold** in the corresponding Admin section. The lower
and upper entries are placed **in the same row as the corresponding slider**.
Derived values with no separate slider (PR, ART MAP, NIBP MAP) have their own
compact alarm rows. The Admin window opens wider to accommodate these controls.
Alarm fields are text inputs; leave a field **blank** (displayed as `off`) to disable just
that threshold. With both fields blank, the whole numeric alarm is disabled.
The value may equal either limit without triggering an alarm; only readings
strictly below/above the respective threshold are out of range. If both limits
@ -64,11 +67,40 @@ each channel, with an arrow for the corresponding direction. A disabled
boundary is displayed as `---`. For instance:
```text
⚠ SYS ↑180 ↓80
MAP ↑--- ↓60
DIA ↑--- ↓---
⚠ SYS ↓80 ↑180
MAP ↓60 ↑---
DIA ↓--- ↑---
```
### Alert presets
The **Alert presets** dropdown at the top of Admin offers **Custom** (manual)
and **Adult**. Choosing **Adult** immediately sets all alarm limits and updates
the lower/upper text fields alongside the sliders. It changes *only* alarm limits:
it does not change heart rate, blood pressure, channel visibility or any other
simulated physiological setting. Editing any alarm field after applying Adult
switches the selector to **Custom** without changing other alarm limits. Select
**Adult** again to reapply the complete preset.
| Parameter | Lower | Upper |
| --- | ---: | ---: |
| HR (bpm) | 60 | 100 |
| SpO₂ (%) | 90 | --- |
| PR (bpm) | --- | --- |
| ART SYS (mmHg) | 80 | 180 |
| ART MAP (mmHg) | 60 | 100 |
| ART DIA (mmHg) | --- | --- |
| PPV (%) | --- | 10 |
| NIBP SYS (mmHg) | 80 | 180 |
| NIBP MAP (mmHg) | 60 | 100 |
| NIBP DIA (mmHg) | --- | --- |
| EtCO₂ (mmHg) | 30 | 50 |
| RR (/min) | 10 | 20 |
`---` denotes a disabled threshold. Adult is a **teaching convenience preset**,
not a clinically validated set of alarm limits. Initially the selector is
**Custom**, with all limits disabled (as in v0.4.0).
When a number violates a configured limit, **that number flashes by inverting
its foreground and background colors** (500 ms inverted / 500 ms normal).
Each pressure component (SYS, DIA, MAP) and the PPV reading flash independently. The waveform