v0.4.3: Debug and optimize alarms and rhythms

This commit is contained in:
GABRIEL Sebastian 2026-11-12 12:30:54 +01:00
commit d4162c49e8
8 changed files with 995 additions and 54 deletions

View file

@ -438,7 +438,7 @@ dependencies = [
[[package]]
name = "patient-monitor-simulation"
version = "0.4.2"
version = "0.4.3"
dependencies = [
"gtk4",
]

View file

@ -1,6 +1,6 @@
[package]
name = "patient-monitor-simulation"
version = "0.4.2"
version = "0.4.3"
edition = "2021"
license = "MIT"
description = "Synthetic patient-monitor simulation for teaching (not for clinical use)"

View file

@ -1,4 +1,4 @@
# Patient Monitor Simulation (PaMSim) — v0.4.2
# Patient Monitor Simulation (PaMSim) — v0.4.3
A two-window, Rust/GTK4 patient-monitor simulator for teaching on Fedora Linux.
@ -131,15 +131,21 @@ Admin → **Pulse oximetry** has two additional controls:
| Setting | Pleth waveform | Monitor display |
| --- | --- | --- |
| **0 — None** | No waveform | Flashing **NO READING**, PR `--`; no stars |
| **1 — Poor** | Small, irregularly amplitude-distorted pulse | 1 star (★) |
| **1 — Poor** | Small, irregularly distorted pulse, with temporary losses of signal | 1 star (★) when available; blinking **NO READING** during loss |
| **2 — Adequate** | Medium, smoothly damped pulse | 2 stars (★★) |
| **3 — Good** | Full nominal, smoothly damped pulse | 3 stars (★★★) |
Default: sensor attached; signal quality **3 — Good**. At levels 1–3, the
instructor-set SpO₂ number is shown only when mechanical pulses are detected.
The signal quality does **not** randomly alter that number: it affects the
pleth morphology and the quality indicator only. The simulated pulse rate
remains derived from mechanically captured QRS complexes.
In **Poor** quality, the signal also has intermittent complete dropouts:
**30–180 seconds of signal** between interruptions, with each interruption
lasting **5–20 seconds**. During an interruption the pleth trace disappears,
SpO₂ and PR are unavailable, and **NO READING** flashes. Following recovery,
the Poor waveform and numerical readings return automatically. These intervals
are pseudo-random and driven by simulation time, independent of frame rate;
they reset when Poor quality is selected again. The dropout does not affect
the actual saturation setting, ECG, derived mechanical pulse generation, or ART.
The simulated pulse rate remains derived from captured QRS complexes.
The degraded waveform uses deterministic, smoothly interpolated pseudo-noise
that scales only existing pleth pulses: it cannot manufacture pulses during
@ -159,8 +165,12 @@ can be taught.
controlled by a pulse-rate slider.
- A captured QRS leads to an ART upstroke after a schematic **120 ms**, and a
peripheral pleth pulse after **300 ms**. Pleth is strongly damped/filtered.
- VT and PVT selections start with a heart rate of **160 bpm**, and update
both the HR slider and numeric entry. The instructor may modify this afterward.
- PVT, VF, asystole and PEA suppress mechanical pulses; the ART trace then
drifts smoothly at 0–5 mmHg rather than representing a pulsatile pressure.
The ART **SYS** number shows the rounded low transducer pressure (0–5 mmHg),
while DIA and MAP show `--` because they cannot be reliably estimated.
Previous pressure samples scroll out of view. NIBP is a static,
separately entered cuff reading.
@ -297,6 +307,26 @@ environment; run `cargo test` and `cargo run --release` on Fedora.
- **VT** remains potentially perfusing. The new **PVT** (pulseless ventricular tachycardia) rhythm retains organized broad QRS complexes but produces no arterial/pleth pulses. VF and asystole are also pulseless regardless of capture selection.
- **ECG**: a flashing **ARRHYTHMIA** message for VT, PVT and VF; a flashing **ASYSTOLE** message for asystole. These event alarms do not depend on numerical HR limits.
- **SpO₂**: when the probe is disconnected, a flashing **NO SENSOR** replaces the saturation reading. Quality 0, loss of mechanical capture or a pulseless rhythm results in flashing **NO READING** (no numerical SpO₂/PR). Sensor disconnection takes precedence. No pleth waveform is displayed in these conditions.
- **ART**: without mechanical pulses (including PVT/VF/asystole/PEA), the simulated invasive pressure line has slow non-pulsatile drift between **0 and 5 mmHg**. The numerical blood-pressure reading remains unavailable (`--/--`), rather than misrepresenting this baseline as true systolic and diastolic pressures.
- **ART**: without mechanical pulses (including PVT/VF/asystole/PEA), the simulated invasive pressure line has slow non-pulsatile drift between **0 and 5 mmHg**. In v0.4.2, the numeric display showed `--/--`; v0.4.3 changes this to a rounded measured SYS of 0–5 mmHg, with `--` for DIA and MAP.
- **ECG rhythm name** has been removed from the monitor; only HR and the associated alarm message, when present, are visible. The instructor continues to select the rhythm in Admin.
- Numerical alarm thresholds and the Adult preset from v0.4.1 are unchanged. Synthetic demonstration behavior only; not suitable for clinical decisions.
## v0.4.3 — Low arterial pressure, VT rate preset and intermittent Poor SpO₂
- Pulseless rhythms, PEA or otherwise absent mechanical capture: invasive SYS
reports the **current measured 0–5 mmHg** pressure, e.g. `3/--` with
`MAP -- mmHg`. **Only SYS** is tested against configured ART SYS limits;
no spurious MAP/DIA threshold alarms are generated from missing readings.
- Selecting **VT** or **PVT** in Admin sets HR to **160 bpm** and updates the
associated slider and text field. Subsequent manual rate changes are allowed.
Perfusing VT still produces QRS-linked pulses subject to capture settings;
PVT remains pulseless.
- **Poor (★)** SpO₂ signal periodically changes to **NO READING** for a
pseudo-random **5–20 seconds**, after a pseudo-random **30–180 seconds** of
usable low-quality signal. The technical alarm blinks, SpO₂/PR readings are
suppressed, and the pleth waveform is absent. On recovery it resumes at Poor
quality. The simulated signal history retains visible gaps for outages.
Choosing another signal quality immediately clears the intermittent outage.
- No changes to Adult alarm presets, alarm limit ordering, or the 0–200 mmHg
arterial scale. All behavior is synthetic and intended for teaching only.

View file

@ -221,12 +221,29 @@ pub struct Parameters {
}
impl Parameters {
/// Selecting a ventricular tachycardia starts at the teaching default of
/// 160 bpm. The instructor can subsequently adjust the heart rate.
pub fn select_ecg_rhythm(&mut self, rhythm: EcgRhythm) {
self.ecg_rhythm = rhythm;
if matches!(rhythm, EcgRhythm::VentricularTachycardia
| EcgRhythm::PulselessVentricularTachycardia) {
self.heart_rate = 160;
}
}
/// No saturation without an attached sensor, usable signal quality and
/// two observed mechanical pulse events. Thresholds do not control this alarm.
pub fn spo2_status(&self, pulse_rate: Option<i32>) -> Spo2Status {
self.spo2_status_during(pulse_rate, false)
}
/// Poor-quality signal interruptions are generated by the simulator and
/// override the normal reading, but not a physically disconnected sensor.
pub fn spo2_status_during(&self, pulse_rate: Option<i32>, signal_lost: bool) -> Spo2Status {
if !self.spo2_sensor_attached {
Spo2Status::NoSensor
} else if self.spo2_quality == Spo2Quality::None || pulse_rate.is_none()
} else if (signal_lost && self.spo2_quality == Spo2Quality::Poor)
|| self.spo2_quality == Spo2Quality::None || pulse_rate.is_none()
|| !self.ecg_rhythm.can_generate_pulse()
|| self.pulse_conduction == PulseConduction::None
|| self.heart_rate <= 0
@ -368,6 +385,34 @@ mod tests {
assert_eq!(AtrialFibrillation.rhythm_alarm(), None);
}
#[test]
fn tachycardia_selection_sets_160_bpm_but_other_rhythms_do_not() {
let mut p = Parameters::default();
p.select_ecg_rhythm(EcgRhythm::VentricularTachycardia);
assert_eq!(p.heart_rate, 160);
p.heart_rate = 185; // a subsequent instructor override remains possible
p.select_ecg_rhythm(EcgRhythm::Sinus);
assert_eq!(p.heart_rate, 185);
p.select_ecg_rhythm(EcgRhythm::PulselessVentricularTachycardia);
assert_eq!(p.heart_rate, 160);
assert!(!p.ecg_rhythm.can_generate_pulse());
}
#[test]
fn transient_poor_signal_loss_preserves_sensor_precedence() {
let mut p = Parameters::default();
p.spo2_quality = Spo2Quality::Poor;
assert_eq!(p.spo2_status_during(Some(80), false), Spo2Status::Available);
assert_eq!(p.spo2_status_during(Some(80), true), Spo2Status::NoReading);
p.spo2_sensor_attached = false;
assert_eq!(p.spo2_status_during(Some(80), true), Spo2Status::NoSensor);
p.spo2_sensor_attached = true;
p.spo2_quality = Spo2Quality::Good;
assert_eq!(p.spo2_status_during(Some(80), false), Spo2Status::Available);
assert_eq!(p.spo2_status_during(Some(80), true), Spo2Status::Available);
}
#[test]
fn sensor_and_no_reading_alarms_override_spo2_thresholds() {
let mut p = Parameters::default();

View file

@ -105,8 +105,9 @@ fn pleth_trace_enabled(p: &Parameters) -> bool {
p.spo2_sensor_attached && p.spo2_quality != Spo2Quality::None
}
fn spo2_vitals(p: &Parameters, pulse_rate: Option<i32>) -> (String, String, String) {
match p.spo2_status(pulse_rate) {
fn spo2_vitals(p: &Parameters, pulse_rate: Option<i32>, signal_lost: bool)
-> (String, String, String) {
match p.spo2_status_during(pulse_rate, signal_lost) {
Spo2Status::NoSensor => (
"SpO₂".to_owned(), "NO SENSOR".to_owned(), "PR -- bpm".to_owned()),
Spo2Status::NoReading => (
@ -143,7 +144,7 @@ fn draw_vitals(cr: &Context, ch: Channel, color: Color, x: f64,
let ecg_hr = if p.ecg_rhythm.has_qrs() && p.heart_rate > 0 {
Some(p.heart_rate)
} else { None };
let spo2_status = p.spo2_status(pulse_rate);
let spo2_status = p.spo2_status_during(pulse_rate, sim.poor_signal_lost());
let oxygen = if spo2_status == Spo2Status::Available { Some(p.spo2) } else { None };
let art_map = mean_arterial_pressure(p.art_systolic, p.art_diastolic);
@ -173,7 +174,7 @@ fn draw_vitals(cr: &Context, ch: Channel, color: Color, x: f64,
}
}
Channel::Pleth => {
let (label, value, info) = spo2_vitals(p, pulse_rate);
let (label, value, info) = spo2_vitals(p, pulse_rate, sim.poor_signal_lost());
text(cr, &label, x, top + row_h * 0.24, label_size, color, false);
let technical_alarm = spo2_status != Spo2Status::Available;
let value_size = if technical_alarm {
@ -212,8 +213,14 @@ fn draw_vitals(cr: &Context, ch: Channel, color: Color, x: f64,
alarmed(p.alarms.art_map, Some(art_map)), flash);
}
} else {
draw_alarm_number(cr, "--/--", x, value_y, number_size * 0.78,
width, color, false, flash);
// Without mechanical activity the transducer still reads the
// slowly drifting 0..5 mmHg pressure. DIA/MAP are indeterminate.
let pulseless = !has_capture;
let sys = if pulseless { sim.measured_art_pressure() } else { None };
let displayed_sys = sys.map_or_else(|| "--".to_owned(), |v| v.to_string());
draw_pressure_components(cr, &displayed_sys, "--", x, value_y,
number_size * 0.78, width, color,
alarmed(p.alarms.art_sys, sys), false, flash);
text(cr, "MAP -- mmHg", x, info_y, sub_size, color, false);
}
}
@ -270,17 +277,24 @@ fn draw_pressure_value(cr: &Context, systolic: i32, diastolic: i32,
x: f64, baseline: f64, nominal_size: f64,
available_width: f64, color: Color,
sys_alarm: bool, dia_alarm: bool, flash: bool) {
let sys = systolic.to_string();
let dia = diastolic.to_string();
draw_pressure_components(cr, &systolic.to_string(), &diastolic.to_string(),
x, baseline, nominal_size, available_width, color, sys_alarm, dia_alarm, flash);
}
#[allow(clippy::too_many_arguments)]
fn draw_pressure_components(cr: &Context, sys: &str, dia: &str,
x: f64, baseline: f64, nominal_size: f64,
available_width: f64, color: Color,
sys_alarm: bool, dia_alarm: bool, flash: bool) {
let full = format!("{}/{}", sys, dia);
let size = fitting_size(cr, &full, nominal_size, available_width, true);
let _ = cr.save();
cr.rectangle(x, baseline - size * 1.28, available_width.max(1.0), size * 1.65);
cr.clip();
let mut cursor = x;
cursor += draw_segment(cr, &sys, cursor, baseline, size, color, sys_alarm && flash, true);
cursor += draw_segment(cr, sys, cursor, baseline, size, color, sys_alarm && flash, true);
cursor += draw_segment(cr, "/", cursor, baseline, size, color, false, true);
draw_segment(cr, &dia, cursor, baseline, size, color, dia_alarm && flash, true);
draw_segment(cr, dia, cursor, baseline, size, color, dia_alarm && flash, true);
let _ = cr.restore();
}
@ -378,7 +392,7 @@ fn draw_trace(cr: &Context, channel: Channel, color: Color,
// Disconnected or absent pleth signal: no curve, including historical pulses.
if matches!(channel, Channel::Pleth)
&& (!pleth_trace_enabled(p)
|| p.spo2_status(sim.pulse_rate()) != Spo2Status::Available)
|| p.spo2_status_during(sim.pulse_rate(), sim.poor_signal_lost()) != Spo2Status::Available)
{
let _ = cr.restore();
return;
@ -393,6 +407,10 @@ fn draw_trace(cr: &Context, channel: Channel, color: Color,
let x = right - (last_t - s.time) / VISIBLE_SECONDS * trace_width;
let amplitude = normalized(channel, s, p);
let y = bottom - amplitude * trace_height;
if matches!(channel, Channel::Pleth) && !s.pleth_valid {
if started { let _ = cr.stroke(); started = false; }
continue;
}
if started { cr.line_to(x, y); }
else { cr.move_to(x, y); started = true; }
}
@ -540,13 +558,13 @@ mod tests {
assert!(pleth_trace_enabled(&p));
p.spo2_sensor_attached = false;
assert!(!pleth_trace_enabled(&p));
assert_eq!(spo2_vitals(&p, Some(75)).1, "NO SENSOR");
assert!(spo2_vitals(&p, Some(75)).2.contains("PR --"));
assert_eq!(spo2_vitals(&p, Some(75), false).1, "NO SENSOR");
assert!(spo2_vitals(&p, Some(75), false).2.contains("PR --"));
p.spo2_sensor_attached = true;
p.spo2_quality = Spo2Quality::None;
assert!(!pleth_trace_enabled(&p));
assert_eq!(spo2_vitals(&p, Some(75)).1, "NO READING");
assert!(spo2_vitals(&p, Some(75)).2.contains("PR --"));
assert_eq!(spo2_vitals(&p, Some(75), false).1, "NO READING");
assert!(spo2_vitals(&p, Some(75), false).2.contains("PR --"));
assert!(p.show_spo2); // no readings, but visible row and status
}
@ -554,29 +572,37 @@ mod tests {
fn poor_and_good_quality_keep_derived_pr_and_quality_indicators() {
let mut p = Parameters::default();
p.spo2_quality = Spo2Quality::Poor;
assert_eq!(spo2_vitals(&p, Some(75)).0, "SpO₂ ★");
assert_eq!(spo2_vitals(&p, Some(75)).1, "98");
assert!(spo2_vitals(&p, Some(75)).2.contains("PR 75 bpm"));
assert_eq!(spo2_vitals(&p, Some(75), false).0, "SpO₂ ★");
assert_eq!(spo2_vitals(&p, Some(75), false).1, "98");
assert!(spo2_vitals(&p, Some(75), false).2.contains("PR 75 bpm"));
p.spo2_quality = Spo2Quality::Good;
assert_eq!(spo2_vitals(&p, Some(75)).0, "SpO₂ ★★★");
assert_eq!(spo2_vitals(&p, None).1, "NO READING");
assert!(spo2_vitals(&p, None).2.contains("PR --"));
assert_eq!(spo2_vitals(&p, Some(75), false).0, "SpO₂ ★★★");
assert_eq!(spo2_vitals(&p, None, false).1, "NO READING");
assert!(spo2_vitals(&p, None, false).2.contains("PR --"));
}
#[test]
fn no_sensor_precedes_no_reading_and_both_are_spo2_alarms() {
let mut p = Parameters::default();
assert_eq!(spo2_vitals(&p, None).1, "NO READING");
assert_eq!(spo2_vitals(&p, None, false).1, "NO READING");
p.ecg_rhythm = crate::model::EcgRhythm::PulselessVentricularTachycardia;
assert_eq!(spo2_vitals(&p, Some(100)).1, "NO READING");
assert_eq!(spo2_vitals(&p, Some(100), false).1, "NO READING");
p.spo2_sensor_attached = false;
assert_eq!(spo2_vitals(&p, Some(100)).1, "NO SENSOR");
assert_eq!(spo2_vitals(&p, Some(100), false).1, "NO SENSOR");
}
#[test]
fn intermittent_poor_spo2_loss_raises_no_reading_not_no_sensor() {
let p = Parameters { spo2_quality: Spo2Quality::Poor, ..Parameters::default() };
assert_eq!(spo2_vitals(&p, Some(75), true).1, "NO READING");
assert!(spo2_vitals(&p, Some(75), true).2.contains("PR --"));
assert_eq!(spo2_vitals(&p, Some(75), false).1, "98");
}
#[test]
fn arterial_axis_is_absolute_and_fixed_to_200_mmhg() {
let p = Parameters::default();
let mut s = Sample { time: 0.0, ecg: 0.0, pleth: 0.0, art: 100.0, co2: 0.0 };
let mut s = Sample { time: 0.0, ecg: 0.0, pleth: 0.0, pleth_valid: true, art: 100.0, co2: 0.0 };
assert_eq!(normalized(Channel::Art, &s, &p), 0.5);
s.art = 200.0;
assert_eq!(normalized(Channel::Art, &s, &p), 1.0);

View file

@ -2,7 +2,7 @@
//! A conducted QRS schedules an arterial pulse and a delayed, smoothed pleth pulse.
use std::collections::VecDeque;
use crate::model::{ArterialShape, CapnoShape, EcgRhythm, Parameters, PulseConduction};
use crate::model::{ArterialShape, CapnoShape, EcgRhythm, Parameters, PulseConduction, Spo2Quality};
const SAMPLE_STEP: f64 = 0.004; // 250 Hz; adequately samples the QRS upstroke
const HISTORY_SECONDS: f64 = 10.0;
@ -32,10 +32,59 @@ pub struct Sample {
pub time: f64,
pub ecg: f64, // arbitrary amplitude
pub pleth: f64, // attenuated, low-pass-filtered amplitude 0..1
pub pleth_valid: bool, // gaps during intermittent signal dropout
pub art: f64, // absolute pressure in mmHg; display clips above 200
pub co2: f64, // mmHg
}
/// Reproducible pseudo-random outages of the pulse-oximeter in Poor quality.
/// Each recovery is followed by 30..180 s of signal before a 5..20 s dropout.
/// The schedule advances in simulation seconds and does not depend on FPS.
#[derive(Debug)]
struct PoorSignalOutages {
rng: u64,
enabled: bool,
active: bool,
next_start: f64,
end: f64,
}
impl PoorSignalOutages {
fn new() -> Self {
Self { rng: 0xC3A5_C85C_97CB_3127, enabled: false,
active: false, next_start: 0.0, end: 0.0 }
}
fn random_range(&mut self, lower: f64, upper: f64) -> f64 {
// 64-bit LCG; deterministic, not cryptographic randomness.
self.rng = self.rng.wrapping_mul(6_364_136_223_846_793_005)
.wrapping_add(1_442_695_040_888_963_407);
let fraction = (self.rng >> 11) as f64 / (1_u64 << 53) as f64;
lower + (upper - lower) * fraction
}
fn update(&mut self, now: f64, poor: bool) {
if !poor {
self.enabled = false;
self.active = false;
return;
}
// Deliberately do not schedule during the negative-time trace warmup.
if now < 0.0 { return; }
if !self.enabled {
self.enabled = true;
self.active = false;
self.next_start = now + self.random_range(30.0, 180.0);
} else if self.active && now >= self.end {
self.active = false;
self.next_start = now + self.random_range(30.0, 180.0);
} else if !self.active && now >= self.next_start {
self.active = true;
self.end = now + self.random_range(5.0, 20.0);
}
}
}
pub struct Simulator {
samples: VecDeque<Sample>,
time: f64,
@ -48,6 +97,7 @@ pub struct Simulator {
previous_rhythm: EcgRhythm,
previous_conduction: PulseConduction,
previous_heart_rate: i32,
poor_signal_outages: PoorSignalOutages,
pending_time: f64,
}
@ -65,6 +115,7 @@ impl Simulator {
previous_rhythm: params.ecg_rhythm,
previous_conduction: params.pulse_conduction,
previous_heart_rate: params.heart_rate,
poor_signal_outages: PoorSignalOutages::new(),
pending_time: 0.0,
};
// Prepopulate the trace and pulse history, including the initial pulse rate.
@ -83,6 +134,15 @@ impl Simulator {
}
}
pub fn poor_signal_lost(&self) -> bool {
self.poor_signal_outages.active
}
/// Last arterial transducer sample, in mmHg (including slow 0..5 drift).
pub fn measured_art_pressure(&self) -> Option<i32> {
self.samples.back().map(|sample| sample.art.round() as i32)
}
pub fn samples(&self) -> &VecDeque<Sample> {
&self.samples
}
@ -106,6 +166,8 @@ impl Simulator {
fn step(&mut self, params: &Parameters) {
self.time += SAMPLE_STEP;
self.poor_signal_outages.update(self.time,
params.spo2_quality == Spo2Quality::Poor && params.spo2_sensor_attached);
if params.ecg_rhythm != self.previous_rhythm
|| params.pulse_conduction != self.previous_conduction
@ -180,7 +242,10 @@ impl Simulator {
capno_amplitude(self.respiratory_phase, params.capno_shape)
};
self.samples.push_back(Sample { time: self.time, ecg, pleth, art, co2 });
let pleth_valid = params.spo2_sensor_attached
&& params.spo2_quality != Spo2Quality::None
&& !self.poor_signal_outages.active;
self.samples.push_back(Sample { time: self.time, ecg, pleth, pleth_valid, art, co2 });
while self.samples.front().is_some_and(|first| first.time < self.time - HISTORY_SECONDS) {
self.samples.pop_front();
}
@ -476,6 +541,68 @@ mod tests {
assert!(spread > 0.25);
}
#[test]
fn low_arterial_pressure_stays_measurable_without_pulses() {
for rhythm in [EcgRhythm::PulselessVentricularTachycardia,
EcgRhythm::VentricularFibrillation, EcgRhythm::Asystole] {
let p = Parameters { ecg_rhythm: rhythm, ..Parameters::default() };
let mut sim = Simulator::new(&p);
for _ in 0..500 { sim.step(&p); }
assert_eq!(sim.pulse_rate(), None);
let pressure = sim.measured_art_pressure().unwrap();
assert!((0..=5).contains(&pressure), "{rhythm:?}: {pressure}");
}
}
#[test]
fn poor_signal_outages_have_random_intervals_and_durations() {
let mut outage = PoorSignalOutages::new();
outage.update(0.0, true);
let mut start = 0.0;
let mut gaps = Vec::new();
let mut durations = Vec::new();
for _ in 0..16 {
let next = outage.next_start;
gaps.push(next - start);
assert!((30.0..=180.0).contains(&(next - start)));
outage.update(next, true);
assert!(outage.active);
let end = outage.end;
durations.push(end - next);
assert!((5.0..=20.0).contains(&(end - next)));
outage.update(end, true);
assert!(!outage.active);
start = end;
}
assert!(gaps.iter().any(|v| (v - gaps[0]).abs() > 1.0));
assert!(durations.iter().any(|v| (v - durations[0]).abs() > 1.0));
outage.update(start + 1.0, false);
assert!(!outage.active);
outage.update(start + 2.0, true);
assert!(!outage.active); // switching back to Poor reschedules
}
#[test]
fn poor_signal_dropout_does_not_change_art_or_ecg() {
let mut p = Parameters::default();
p.spo2_quality = Spo2Quality::Poor;
let mut sim = Simulator::new(&p);
// Force an outage to test signal gaps without synthesizing minutes.
sim.poor_signal_outages.enabled = true;
sim.poor_signal_outages.active = true;
sim.poor_signal_outages.end = sim.time + 10.0;
sim.step(&p);
assert!(!sim.samples.back().unwrap().pleth_valid);
assert!(sim.poor_signal_lost());
assert!((0.0..=200.0).contains(&sim.samples.back().unwrap().art));
assert_eq!(p.spo2_status_during(sim.pulse_rate(), sim.poor_signal_lost()),
crate::model::Spo2Status::NoReading);
p.spo2_quality = Spo2Quality::Good;
sim.step(&p);
assert!(!sim.poor_signal_lost());
assert!(sim.samples.back().unwrap().pleth_valid);
}
#[test]
fn pulse_rate_follows_two_to_one_capture() {
let p = Parameters { pulse_conduction: PulseConduction::TwoToOne, ..Parameters::default() };

View file

@ -145,11 +145,11 @@ fn create_admin(app: &Application, params: SharedParams,
let ecg = group(&controls, "ECG");
add_visibility(&ecg, params.borrow().show_ecg, params.clone(), |p, show| p.show_ecg = show);
add_dropdown(&ecg, "ECG rhythm", &[
let rhythm_dropdown = add_dropdown_at(&ecg, "ECG rhythm", &[
"Sinus rhythm", "Atrial fibrillation", "Ventricular tachycardia (VT)",
"Pulseless ventricular tachycardia (PVT)", "Ventricular fibrillation", "Asystole",
], params.clone(), |p, selected| {
p.ecg_rhythm = match selected {
], 0, params.clone(), |p, selected| {
let rhythm = match selected {
1 => EcgRhythm::AtrialFibrillation,
2 => EcgRhythm::VentricularTachycardia,
3 => EcgRhythm::PulselessVentricularTachycardia,
@ -157,16 +157,25 @@ fn create_admin(app: &Application, params: SharedParams,
5 => EcgRhythm::Asystole,
_ => EcgRhythm::Sinus,
};
p.select_ecg_rhythm(rhythm);
});
let (hr_row, hr_slider, hr_entry) = add_numeric(&ecg, "Heart rate", "bpm", 0, 240,
params.borrow().heart_rate, params.clone(), |p, v| p.heart_rate = v);
// Programmatic 160 bpm VT/PVT preset must update both controls,
// including an HR field that currently has keyboard focus.
rhythm_dropdown.connect_selected_notify(move |selector| {
if selector.selected() == 2 || selector.selected() == 3 {
hr_slider.set_value(160.0);
hr_entry.set_text("160");
}
});
let hr_row = add_numeric(&ecg, "Heart rate", "bpm", 0, 240, params.borrow().heart_rate,
params.clone(), |p, v| p.heart_rate = v);
alarm_editors.push((add_alarm_limits(&hr_row, "", "", 240,
params.borrow().alarms.hr, params.clone(), |p, v| p.alarms.hr = v,
&preset_dropdown, &updating_preset), |s| s.hr));
let spo2 = group(&controls, "Pulse oximetry");
add_visibility(&spo2, params.borrow().show_spo2, params.clone(), |p, show| p.show_spo2 = show);
let spo2_row = add_numeric(&spo2, "SpO₂", "%", 0, 100, params.borrow().spo2,
let (spo2_row, _, _) = add_numeric(&spo2, "SpO₂", "%", 0, 100, params.borrow().spo2,
params.clone(), |p, v| p.spo2 = v);
alarm_editors.push((add_alarm_limits(&spo2_row, "", "", 100,
params.borrow().alarms.spo2, params.clone(), |p, v| p.alarms.spo2 = v,
@ -181,7 +190,7 @@ fn create_admin(app: &Application, params: SharedParams,
move |button| params.borrow_mut().spo2_sensor_attached = !button.is_active()
});
spo2.append(&sensor);
add_dropdown_at(&spo2, "Signal quality", &[
let _ = add_dropdown_at(&spo2, "Signal quality", &[
"0 — None (no reading)",
"1 — Poor (*)",
"2 — Adequate (**)",
@ -209,6 +218,18 @@ fn create_admin(app: &Application, params: SharedParams,
_ => PulseConduction::OneToOne,
};
});
let dropout_hint = gtk::Label::new(Some(
"Poor signal: randomly loses all SpO₂/PR readings for 5–20 s,"
));
dropout_hint.set_xalign(0.0);
dropout_hint.set_wrap(true);
spo2.append(&dropout_hint);
let dropout_hint2 = gtk::Label::new(Some(
"with 30–180 s of usable poor-quality signal between interruptions."
));
dropout_hint2.set_xalign(0.0);
dropout_hint2.set_wrap(true);
spo2.append(&dropout_hint2);
let conduction_hint = gtk::Label::new(Some(
"Pulse rate is derived from captured QRS events.
ART and pleth are delayed. VT can be perfusing; PVT, VF and asystole are pulseless."
@ -219,7 +240,7 @@ ART and pleth are delayed. VT can be perfusing; PVT, VF and asystole are pulsele
let art = group(&controls, "Invasive arterial blood pressure");
add_visibility(&art, params.borrow().show_art, params.clone(), |p, show| p.show_art = show);
let art_sys_row = add_numeric(&art, "Systolic", "mmHg", 0, 200, params.borrow().art_systolic,
let (art_sys_row, _, _) = add_numeric(&art, "Systolic", "mmHg", 0, 200, params.borrow().art_systolic,
params.clone(), |p, v| p.art_systolic = v);
alarm_editors.push((add_alarm_limits(&art_sys_row, "", "", 200,
params.borrow().alarms.art_sys, params.clone(), |p, v| p.alarms.art_sys = v,
@ -227,12 +248,12 @@ ART and pleth are delayed. VT can be perfusing; PVT, VF and asystole are pulsele
alarm_editors.push((add_alarm_limits(&art, "MAP (derived)", "mmHg", 200,
params.borrow().alarms.art_map, params.clone(), |p, v| p.alarms.art_map = v,
&preset_dropdown, &updating_preset), |s| s.art_map));
let art_dia_row = add_numeric(&art, "Diastolic", "mmHg", 0, 200, params.borrow().art_diastolic,
let (art_dia_row, _, _) = add_numeric(&art, "Diastolic", "mmHg", 0, 200, params.borrow().art_diastolic,
params.clone(), |p, v| p.art_diastolic = v);
alarm_editors.push((add_alarm_limits(&art_dia_row, "", "", 200,
params.borrow().alarms.art_dia, params.clone(), |p, v| p.alarms.art_dia = v,
&preset_dropdown, &updating_preset), |s| s.art_dia));
let ppv_row = add_numeric(&art, "Pulse pressure variation", "%", 0, 60,
let (ppv_row, _, _) = add_numeric(&art, "Pulse pressure variation", "%", 0, 60,
params.borrow().pulse_pressure_variation, params.clone(),
|p, v| p.pulse_pressure_variation = v);
alarm_editors.push((add_alarm_limits(&ppv_row, "", "", 60,
@ -256,7 +277,7 @@ ART and pleth are delayed. VT can be perfusing; PVT, VF and asystole are pulsele
let nibp = group(&controls, "Non-invasive blood pressure (NIBP)");
add_visibility(&nibp, params.borrow().show_nibp, params.clone(), |p, show| p.show_nibp = show);
let nibp_sys_row = add_numeric(&nibp, "Systolic", "mmHg", 0, 260, params.borrow().nibp_systolic,
let (nibp_sys_row, _, _) = add_numeric(&nibp, "Systolic", "mmHg", 0, 260, params.borrow().nibp_systolic,
params.clone(), |p, v| p.nibp_systolic = v);
alarm_editors.push((add_alarm_limits(&nibp_sys_row, "", "", 260,
params.borrow().alarms.nibp_sys, params.clone(), |p, v| p.alarms.nibp_sys = v,
@ -264,7 +285,7 @@ ART and pleth are delayed. VT can be perfusing; PVT, VF and asystole are pulsele
alarm_editors.push((add_alarm_limits(&nibp, "MAP (derived)", "mmHg", 260,
params.borrow().alarms.nibp_map, params.clone(), |p, v| p.alarms.nibp_map = v,
&preset_dropdown, &updating_preset), |s| s.nibp_map));
let nibp_dia_row = add_numeric(&nibp, "Diastolic", "mmHg", 0, 200, params.borrow().nibp_diastolic,
let (nibp_dia_row, _, _) = add_numeric(&nibp, "Diastolic", "mmHg", 0, 200, params.borrow().nibp_diastolic,
params.clone(), |p, v| p.nibp_diastolic = v);
alarm_editors.push((add_alarm_limits(&nibp_dia_row, "", "", 200,
params.borrow().alarms.nibp_dia, params.clone(), |p, v| p.alarms.nibp_dia = v,
@ -277,12 +298,12 @@ ART and pleth are delayed. VT can be perfusing; PVT, VF and asystole are pulsele
p.capno_shape = if selected == 1 { CapnoShape::Obstructive }
else { CapnoShape::Normal };
});
let etco2_row = add_numeric(&capno, "EtCO₂", "mmHg", 0, 100, params.borrow().etco2,
let (etco2_row, _, _) = add_numeric(&capno, "EtCO₂", "mmHg", 0, 100, params.borrow().etco2,
params.clone(), |p, v| p.etco2 = v);
alarm_editors.push((add_alarm_limits(&etco2_row, "", "", 100,
params.borrow().alarms.etco2, params.clone(), |p, v| p.alarms.etco2 = v,
&preset_dropdown, &updating_preset), |s| s.etco2));
let rr_row = add_numeric(&capno, "Respiratory rate", "/min", 0, 60, params.borrow().respiratory_rate,
let (rr_row, _, _) = add_numeric(&capno, "Respiratory rate", "/min", 0, 60, params.borrow().respiratory_rate,
params.clone(), |p, v| p.respiratory_rate = v);
alarm_editors.push((add_alarm_limits(&rr_row, "", "", 60,
params.borrow().alarms.rr, params.clone(), |p, v| p.alarms.rr = v,
@ -355,11 +376,11 @@ where F: Fn(&mut Parameters, bool) + 'static {
fn add_dropdown<F>(parent: &gtk::Box, label: &str, items: &[&str],
params: SharedParams, setter: F)
where F: Fn(&mut Parameters, u32) + 'static {
add_dropdown_at(parent, label, items, 0, params, setter);
let _ = add_dropdown_at(parent, label, items, 0, params, setter);
}
fn add_dropdown_at<F>(parent: &gtk::Box, label: &str, items: &[&str],
initial: u32, params: SharedParams, setter: F)
initial: u32, params: SharedParams, setter: F) -> gtk::DropDown
where F: Fn(&mut Parameters, u32) + 'static {
let row = gtk::Box::new(Orientation::Horizontal, 8);
let caption = gtk::Label::new(Some(label));
@ -374,12 +395,13 @@ where F: Fn(&mut Parameters, u32) + 'static {
row.append(&caption);
row.append(&combo);
parent.append(&row);
combo
}
#[allow(clippy::too_many_arguments)]
fn add_numeric<F>(parent: &gtk::Box, label: &str, unit: &str,
min: i32, max: i32, initial: i32,
params: SharedParams, setter: F) -> gtk::Box
params: SharedParams, setter: F) -> (gtk::Box, gtk::Scale, gtk::Entry)
where F: Fn(&mut Parameters, i32) + 'static {
let row = gtk::Box::new(Orientation::Horizontal, 8);
let caption = gtk::Label::new(Some(label));
@ -434,7 +456,7 @@ where F: Fn(&mut Parameters, i32) + 'static {
row.append(&entry);
row.append(&suffix);
parent.append(&row);
row
(row, slider, entry)
}
fn normalize_entry(entry: &gtk::Entry, slider: &gtk::Scale, min: i32, max: i32) {