v0.6.6 — Optional QRS/pulse beep

Added a separate **QRS tone** checkbox (enabled by default). An 85 ms
single-note beep follows actual ECG or mechanical beat events. With a valid
SpO₂ reading, the delayed pleth pulse determines the timing and the oxygen
saturation determines the pitch in eleven steps from 740 to 87 Hz; otherwise
ECG QRS or ART pulses trigger a fixed 500 Hz tone. The two-tone alarm sounds
remain unchanged. Added regression tests for band boundaries, event selection
and single-tone WAV output.
This commit is contained in:
GABRIEL Sebastian 2026-11-12 22:36:24 +01:00
commit f98fc7412f
11 changed files with 1247 additions and 34 deletions

View file

@ -1,4 +1,4 @@
# Patient Monitor Simulation (PaMSim) — v0.6.5
# Patient Monitor Simulation (PaMSim) — v0.6.6
A Rust/GTK4 patient-monitor simulator with two main windows for teaching on Fedora Linux.
@ -166,7 +166,7 @@ The monitor can be resized and toggled fullscreen with **F11**; **Esc** exits
fullscreen. On Wayland, move the monitor to the presentation screen before
activating fullscreen. Fullscreen does not lock down the computer.
## Audible alarms (updated v0.6.5)
## Audible alarms and QRS tone (updated v0.6.6)
Sound is **enabled by default**. The Admin window has an **Audible alarms**
checkbox for muting and a **Test tone** button (works even while muted).
@ -179,8 +179,8 @@ Visual alarms keep blinking normally when sound is off.
and **NO READING** produce a lower-pitched **two-tone** alert
(700 Hz followed by 950 Hz), repeated approximately every 2.5 seconds.
- Both alarm signals use two **different successive pitches**, separated by
a brief silence. A distinct **single QRS beep is reserved for future work**;
QRS-related audio is not yet implemented.
a brief silence. The single QRS/pulse beep is independent and configurable
using the new **QRS tone** checkbox in Admin (ON by default).
- Critical alarms have priority: only one tone pattern is played even if
multiple alarms are active. The sound stops when all alarms clear or when
the checkbox is unchecked. Alarms on hidden monitor channels are excluded.
@ -188,6 +188,41 @@ Visual alarms keep blinking normally when sound is off.
trigger numerical limit alarms. The cuff measurement cycle by itself is not
an alarm. Audio playback is separate from waveform rendering and simulation.
### Beat-synchronous QRS/pulse tone
The **QRS tone** checkbox is independent of **Audible alarms**; turning off
one does not mute the other. Each brief (85 ms), **single-frequency** beep
follows an actual simulated beat event, not a separate rate timer:
- **SpO₂ reading available:** beep on the delayed peripheral pleth pulse;
the pitch follows the *displayed* saturation (including during effective
good-quality chest compressions). Missed mechanical captures produce no
peripheral beep.
- **SpO₂ reading unavailable:** a fixed **500 Hz** beep follows an organized
ECG QRS if ECG is displayed; otherwise it follows an arterial pulse if
ART is displayed. ECG and ART together do not cause duplicate beeps.
- **VF/asystole without effective CPR:** no beat beep is generated. ART pulses
created by CPR can generate the fallback tone, while good CPR with a valid
SpO₂ reading uses the saturation-dependent tone.
| SpO₂ (%) | Tone (Hz) |
| --- | ---: |
| 97–100 | 740 |
| 94–96 | 675 |
| 90–93 | 609 |
| 80–89 | 544 |
| 70–79 | 479 |
| 60–69 | 413 |
| 50–59 | 348 |
| 40–49 | 283 |
| 30–39 | 218 |
| 20–29 | 152 |
| 0–19 | 87 |
Saturation tone pitches are deliberately simplified teaching tones, not
calibrated clinical pulse-oximeter sounds. The QRS beep and two-tone alarm can
sound independently; no alarm priority or visual alarm logic was changed.
On Fedora, sound is played through **`pw-play`** from `pipewire-utils`, with
**`paplay`** as a fallback. No additional Rust crates or sound files are needed:
the application creates its own short WAV tones at runtime. If neither player
@ -647,3 +682,14 @@ successive pitches. Critical alarms retain their priority and faster repeat
interval; standard alarms retain their slower cadence. The Admin test button
plays the standard two-tone alert. A single QRS beep is reserved for a later
release and is not generated yet. No visual alarm or simulation logic changed.
### v0.6.6 — Optional QRS/pulse beep
Added a separate **QRS tone** checkbox (enabled by default). An 85 ms
single-note beep follows actual ECG or mechanical beat events. With a valid
SpO₂ reading, the delayed pleth pulse determines the timing and the oxygen
saturation determines the pitch in eleven steps from 740 to 87 Hz; otherwise
ECG QRS or ART pulses trigger a fixed 500 Hz tone. The two-tone alarm sounds
remain unchanged. Added regression tests for band boundaries, event selection
and single-tone WAV output.