Added a separate **QRS tone** checkbox (enabled by default). An 85 ms
single-note beep follows actual ECG or mechanical beat events. With a valid
SpO₂ reading, the delayed pleth pulse determines the timing and the oxygen
saturation determines the pitch in eleven steps from 740 to 87 Hz; otherwise
ECG QRS or ART pulses trigger a fixed 500 Hz tone. The two-tone alarm sounds
remain unchanged. Added regression tests for band boundaries, event selection
and single-tone WAV output.
Both standard/technical and critical alarm sounds now consist of two distinct
successive pitches. Critical alarms retain their priority and faster repeat
interval; standard alarms retain their slower cadence. The Admin test button
plays the standard two-tone alert. A single QRS beep is reserved for a later
release and is not generated yet. No visual alarm or simulation logic changed.
The monitor-side NIBP settings menu again contains **Measure now**,
**Automatic measurement ON/OFF** and the **measurement interval** (1, 3, 5,
10, 20, 30, 60 minutes). These controls are above the horizontal SYS/MAP/DIA
alarm matrix and share its monochrome touchscreen styling. Auto and interval
are saved with Apply; Measure now starts a cuff cycle immediately regardless
of Auto. Both monitor and Admin display the selected settings. Other monitor
menus remain limited to alarm thresholds; NIBP cuff placement, the simulated
failed-reading option and non-NIBP signal settings remain Admin-only.
The ECG, SpO₂, ART, CO₂ and NIBP touch menus now edit only lower/upper
alarm limits. Removed ECG lead, signal/sensor, waveform and all NIBP
measurement controls from monitor-side menus. These remain available in
Admin, including manual NIBP measurement and Auto ON/OFF. No physiological
model, alarm evaluation or NIBP startup behavior was changed.
Added touch/click settings panels for SpO₂, ART, CO₂ and the bottom-right
NIBP cell. Their threshold editors use large, unit-specific +/− and ON/OFF
controls; the panels expose relevant waveform/sensor options and stay
synchronized with Admin. NIBP additionally provides interval selection,
manual **Measure now**, Auto ON/OFF and cuff options. Startup no longer
initiates a cuff cycle or shows any NIBP reading; Auto is OFF by default.
Existing NIBP cycle physics and signal quality behavior are unchanged.
Restyled the ECG settings window with light-grey text/controls, a neutral
0.8-pixel outer border and thin 0.8-pixel Cairo separators matching the
monitor. Replaced the on-monitor HR limit text entries with large 5-bpm
**−/+** buttons and independent **ON/OFF** controls; the ECG lead remains a
touch-selectable dropdown. Pending changes still require Apply, and Cancel
leaves the shared parameters unchanged. No signal generation was altered.
Corrects the underlying cause of the v0.5.5 test failure: in pulseless rhythms, the simulation had reset the pleth low-pass filter at every 250 Hz sample, preventing CPR-induced signals from accumulating. The filter now persists during good-quality chest compressions while QRS-triggered mechanical pulses are still cleared. Stopping CPR in VF/PVT immediately restores no reading. Additional regression coverage includes PVT.