# Patient Monitor Simulation (PaMSim) — v0.4.0
|
2026-11-12 09:36:13 +01:00
|
|
|
|
|
|
|
|
|
|
A two-window, Rust/GTK4 patient-monitor simulator for teaching on Fedora Linux.
|
|
|
|
|
|
|
|
|
|
|
|
> **Synthetic training signals only — not for clinical use.** ECG, arterial pressure,
|
|
|
|
|
|
> plethysmography and capnography are schematic mathematical approximations. The
|
|
|
|
|
|
> simulation has no clinical validation and is not a medical device.
|
|
|
|
|
|
|
|
|
|
|
|
## Build and run on Fedora
|
|
|
|
|
|
|
|
|
|
|
|
```bash
|
|
|
|
|
|
sudo dnf install rust cargo gcc pkgconf-pkg-config gtk4-devel
|
|
|
|
|
|
cargo test
|
|
|
|
|
|
cargo run --release
|
|
|
|
|
|
```
|
|
|
|
|
|
|
|
|
|
|
|
Run these commands from the extracted `patient-monitor-simulation` folder.
|
|
|
|
|
|
Window titles show the application version. GTK4 requires a functional graphical
|
|
|
|
|
|
session (e.g., Fedora KDE/Wayland).
|
|
|
|
|
|
|
|
|
|
|
|
## Windows
|
|
|
|
|
|
|
2026-11-12 10:43:11 +01:00
|
|
|
|
**Admin:** Controls patient name; ECG rhythm and rate; mechanical capture;
|
|
|
|
|
|
SpO₂, signal quality and sensor connection; invasive arterial pressure,
|
|
|
|
|
|
waveform shape and pulse-pressure variation; NIBP; and capnography. Numeric inputs have bidirectionally linked sliders and text fields.
|
2026-11-12 09:36:13 +01:00
|
|
|
|
Each of the five monitor parameters (ECG, SpO₂/pleth, ART, CO₂, NIBP) has its
|
|
|
|
|
|
own **Show on monitor** checkbox.
|
|
|
|
|
|
|
|
|
|
|
|
**Monitor:** Black background; ECG green, pleth yellow, ART red and capnogram
|
2026-11-12 10:43:11 +01:00
|
|
|
|
light grey. The patient name entered in Admin replaces **PATIENT MONITOR** in
|
|
|
|
|
|
the header immediately; leaving the name blank restores the generic title.
|
|
|
|
|
|
Long names are automatically resized or clipped to fit the header.
|
|
|
|
|
|
Waveforms and vital signs disappear together when their respective
|
2026-11-12 09:36:13 +01:00
|
|
|
|
visibility checkbox is cleared. Remaining waveform rows expand to use the space.
|
|
|
|
|
|
Hiding a display channel does **not** stop signal generation or physiological
|
|
|
|
|
|
coupling: for example, hiding ECG does not stop ART or pleth pulses. NIBP can
|
|
|
|
|
|
be hidden independently; this also removes its footer.
|
|
|
|
|
|
|
|
|
|
|
|
The monitor can be resized and toggled fullscreen with **F11**; **Esc** exits
|
|
|
|
|
|
fullscreen. On Wayland, move the monitor to the presentation screen before
|
|
|
|
|
|
activating fullscreen. Fullscreen does not lock down the computer.
|
|
|
|
|
|
|
2026-11-12 11:30:50 +01:00
|
|
|
|
## Numerical alarm limits (new in v0.4.0)
|
|
|
|
|
|
|
|
|
|
|
|
Each displayed numeric measurement has an **independent optional lower (↓) and
|
|
|
|
|
|
upper (↑) alarm threshold** in the corresponding Admin section. Alarm fields
|
|
|
|
|
|
are text inputs; leave a field **blank** (displayed as `off`) to disable just
|
|
|
|
|
|
that threshold. With both fields blank, the whole numeric alarm is disabled.
|
|
|
|
|
|
The value may equal either limit without triggering an alarm; only readings
|
|
|
|
|
|
strictly below/above the respective threshold are out of range. If both limits
|
|
|
|
|
|
are present, the lower limit must not exceed the upper limit. Invalid input is
|
|
|
|
|
|
highlighted and the last valid configuration stays active.
|
|
|
|
|
|
|
|
|
|
|
|
| Monitor section | Independently configurable alarms |
|
|
|
|
|
|
| --- | --- |
|
|
|
|
|
|
| ECG | HR |
|
|
|
|
|
|
| SpO₂ | Oxygen saturation (SpO₂), derived pulse rate (PR) |
|
|
|
|
|
|
| Invasive arterial pressure (ART) | Systolic (SYS), mean (MAP), diastolic (DIA), PPV |
|
|
|
|
|
|
| NIBP | Systolic (SYS), mean (MAP), diastolic (DIA) |
|
|
|
|
|
|
| Capnography | End-tidal CO₂ (EtCO₂), respiratory rate (RR) |
|
|
|
|
|
|
|
|
|
|
|
|
The current limits are displayed in small type at the **upper-right corner** of
|
|
|
|
|
|
each channel, with an arrow for the corresponding direction. A disabled
|
|
|
|
|
|
boundary is displayed as `---`. For instance:
|
|
|
|
|
|
|
|
|
|
|
|
```text
|
|
|
|
|
|
⚠ SYS ↑180 ↓80
|
|
|
|
|
|
MAP ↑--- ↓60
|
|
|
|
|
|
DIA ↑--- ↓---
|
|
|
|
|
|
```
|
|
|
|
|
|
|
|
|
|
|
|
When a number violates a configured limit, **that number flashes by inverting
|
|
|
|
|
|
its foreground and background colors** (500 ms inverted / 500 ms normal).
|
|
|
|
|
|
Each pressure component (SYS, DIA, MAP) and the PPV reading flash independently. The waveform
|
|
|
|
|
|
itself remains unchanged. All alarms are **disabled by default** so old
|
|
|
|
|
|
scenarios do not unexpectedly flash.
|
|
|
|
|
|
|
|
|
|
|
|
Numerical threshold alarms only evaluate actual displayed measurements:
|
|
|
|
|
|
`--`, `NO SENSOR`, and other unavailable readings do not trigger low-value
|
|
|
|
|
|
alarms. ECG-derived PR and signal visibility continue to work as in v0.3.2.
|
|
|
|
|
|
Alarms are presentation-only: there is **no audible alarm, alarm latching,
|
|
|
|
|
|
priority classification or clinical alarm validation**. Instructor settings
|
|
|
|
|
|
are held in memory and reset on restart.
|
|
|
|
|
|
|
2026-11-12 10:43:11 +01:00
|
|
|
|
## Patient name and pulse oximeter signal state (new in v0.3.2)
|
|
|
|
|
|
|
|
|
|
|
|
The new **Patient name** text field at the top of Admin immediately replaces the
|
|
|
|
|
|
monitor headline. It is optional; an empty or whitespace-only name displays
|
|
|
|
|
|
**PATIENT MONITOR**. This is a local teaching label, not a patient record.
|
|
|
|
|
|
|
|
|
|
|
|
Admin → **Pulse oximetry** has two additional controls:
|
|
|
|
|
|
|
|
|
|
|
|
- **Sensor not attached** (checkbox): immediately removes the pleth waveform,
|
|
|
|
|
|
suppresses SpO₂ and pulse-rate readings, and displays **NO SENSOR** instead
|
|
|
|
|
|
of the numeric saturation. This takes precedence over signal quality.
|
|
|
|
|
|
- **Signal quality** (dropdown), which changes the visible pleth waveform and
|
|
|
|
|
|
quality indicator, independently of the instructor-set SpO₂ percentage:
|
|
|
|
|
|
|
|
|
|
|
|
| Setting | Pleth waveform | Monitor display |
|
|
|
|
|
|
| --- | --- | --- |
|
|
|
|
|
|
| **0 — None** | No waveform | SpO₂ `--`, PR `--`, **NO SIGNAL**; no stars |
|
|
|
|
|
|
| **1 — Poor** | Small, irregularly amplitude-distorted pulse | 1 star (★) |
|
|
|
|
|
|
| **2 — Adequate** | Medium, smoothly damped pulse | 2 stars (★★) |
|
|
|
|
|
|
| **3 — Good** | Full nominal, smoothly damped pulse | 3 stars (★★★) |
|
|
|
|
|
|
|
|
|
|
|
|
Default: sensor attached; signal quality **3 — Good**. At levels 1–3, the
|
|
|
|
|
|
instructor-set SpO₂ number is shown only when mechanical pulses are detected.
|
|
|
|
|
|
The signal quality does **not** randomly alter that number: it affects the
|
|
|
|
|
|
pleth morphology and the quality indicator only. The simulated pulse rate
|
|
|
|
|
|
remains derived from mechanically captured QRS complexes.
|
|
|
|
|
|
|
|
|
|
|
|
The degraded waveform uses deterministic, smoothly interpolated pseudo-noise
|
|
|
|
|
|
that scales only existing pleth pulses: it cannot manufacture pulses during
|
|
|
|
|
|
PEA, ventricular fibrillation or asystole. Unplugging a sensor or selecting
|
|
|
|
|
|
quality 0 hides even previously buffered pleth samples immediately. ECG and
|
|
|
|
|
|
ART continue unaffected. The **Show on monitor** checkbox still hides the
|
|
|
|
|
|
entire SpO₂ row; sensor/quality controls leave the row visible so the status
|
|
|
|
|
|
can be taught.
|
|
|
|
|
|
|
2026-11-12 09:36:13 +01:00
|
|
|
|
## ECG and mechanical coupling
|
|
|
|
|
|
|
|
|
|
|
|
- ECG rhythms: sinus, atrial fibrillation, monomorphic ventricular tachycardia,
|
|
|
|
|
|
ventricular fibrillation and asystole.
|
|
|
|
|
|
- Mechanical capture: **1:1**, **2:1**, **3:1**, **None (PEA)**, or deterministic
|
|
|
|
|
|
pseudo-random capture (~65%). PR is derived from conducted QRS events, never
|
|
|
|
|
|
controlled by a pulse-rate slider.
|
|
|
|
|
|
- A captured QRS leads to an ART upstroke after a schematic **120 ms**, and a
|
|
|
|
|
|
peripheral pleth pulse after **300 ms**. Pleth is strongly damped/filtered.
|
|
|
|
|
|
- VF, asystole and PEA suppress mechanical pulses; old samples scroll out of
|
|
|
|
|
|
view. NIBP is a static, separately entered cuff reading.
|
|
|
|
|
|
|
|
|
|
|
|
## Invasive arterial pressure and PPV (new in v0.3)
|
|
|
|
|
|
|
|
|
|
|
|
The arterial waveform has a fixed **0–200 mmHg** display axis, with tick marks
|
|
|
|
|
|
at **0, 50, 100, 150 and 200 mmHg**. Nominal ART systolic and diastolic inputs
|
|
|
|
|
|
are limited to 0–200 mmHg.
|
|
|
|
|
|
|
|
|
|
|
|
**Pulse pressure variation (PPV)** is adjustable from **0% to 60%**, in 1%
|
|
|
|
|
|
increments, via synchronized slider and text field (Admin → Invasive arterial
|
|
|
|
|
|
blood pressure). Its default is **0%** (no respiratory modulation). The
|
|
|
|
|
|
respiratory cycle comes from the same respiratory-rate parameter as the
|
|
|
|
|
|
capnogram. **RR = 0** disables the respiratory modulation (apnea).
|
|
|
|
|
|
|
|
|
|
|
|
The synthetic PPV definition is:
|
|
|
|
|
|
|
|
|
|
|
|
```text
|
|
|
|
|
|
PPV (%) = 100 × (PPmax − PPmin) / ((PPmax + PPmin) / 2)
|
|
|
|
|
|
```
|
|
|
|
|
|
|
|
|
|
|
|
The model applies a sinusoidal variation to **pulse pressure**, sampled for
|
|
|
|
|
|
each mechanically conducted beat. The diastolic baseline remains nominally
|
|
|
|
|
|
fixed, so this is primarily a **beat-to-beat systolic swing**, not an arbitrary
|
|
|
|
|
|
vertical translation of the entire arterial trace. It does not affect ECG,
|
|
|
|
|
|
mechanical capture or pleth amplitude. This is a visual teaching model, not a
|
|
|
|
|
|
validated predictor of fluid responsiveness.
|
|
|
|
|
|
|
|
|
|
|
|
**Example:** ART 120/75 mmHg has nominal pulse pressure 45 mmHg. At PPV 20%,
|
|
|
|
|
|
the ideal pulse-pressure extrema are 40.5 and 49.5 mmHg, corresponding to
|
|
|
|
|
|
systolic peaks of approximately 115.5 and 124.5 mmHg (assuming the diastolic
|
|
|
|
|
|
baseline stays at 75 mmHg). The rendered waveform shape and finite number of
|
|
|
|
|
|
beats per respiratory cycle mean the visible maxima may not reach the ideal
|
|
|
|
|
|
extrema exactly.
|
|
|
|
|
|
|
|
|
|
|
|
The ART sidebar displays the **nominal** systolic/diastolic pressures and MAP,
|
|
|
|
|
|
plus the **configured** PPV percentage when PPV is greater than 0. It does not
|
|
|
|
|
|
measure PPV from sampled arterial data. At very high nominal systolic pressure,
|
|
|
|
|
|
the highest modulated pulses can exceed the 200 mmHg scale; the displayed
|
|
|
|
|
|
curve is then **clipped** at the top, without rescaling the axis.
|
|
|
|
|
|
|
|
|
|
|
|
## Display options (new in v0.3)
|
|
|
|
|
|
|
|
|
|
|
|
The five **Show on monitor** checkboxes affect the presentation only:
|
|
|
|
|
|
|
|
|
|
|
|
| Checkbox | Hidden monitor content |
|
|
|
|
|
|
| --- | --- |
|
|
|
|
|
|
| ECG | ECG waveform, HR and rhythm label |
|
|
|
|
|
|
| Pulse oximetry | Pleth waveform, SpO₂ and PR |
|
|
|
|
|
|
| Invasive arterial blood pressure | ART waveform, pressure values, MAP and PPV label |
|
|
|
|
|
|
| Capnography | CO₂ waveform, EtCO₂ and RR |
|
|
|
|
|
|
| NIBP | Static cuff pressure values and NIBP footer |
|
|
|
|
|
|
|
|
|
|
|
|
Remaining visible waveforms keep their relative order and receive an equal
|
|
|
|
|
|
share of the available area; zero waveform channels is supported.
|
|
|
|
|
|
|
2026-11-12 10:43:11 +01:00
|
|
|
|
## ECG morphology correction (v0.3.1)
|
|
|
|
|
|
|
|
|
|
|
|
At high heart rates, the previous ECG generator expressed QRS width as a fixed
|
|
|
|
|
|
fraction of the R–R interval and sampled at 100 Hz. Narrow R waves could fall
|
|
|
|
|
|
between samples, making successive QRS complexes appear to vary dramatically
|
|
|
|
|
|
in amplitude despite a regular sinus rhythm.
|
|
|
|
|
|
|
|
|
|
|
|
- Synthetic P/QRS/T components now have widths specified in milliseconds,
|
|
|
|
|
|
not fixed fractions of the cardiac cycle. P/T positioning and widths adapt
|
|
|
|
|
|
to the R–R interval, while the QRS width remains essentially constant.
|
|
|
|
|
|
- Sampling increased to **250 Hz** (4 ms intervals) to capture QRS peaks
|
|
|
|
|
|
more consistently, including at **189–240 bpm**.
|
|
|
|
|
|
- The ECG morphology remains independent of the respiratory arterial PPV
|
|
|
|
|
|
setting. The arterial and pleth delays and mechanical capture options are
|
|
|
|
|
|
unchanged.
|
|
|
|
|
|
- Regression tests compare peak heights at 60, 75, 120, 189 and 240 bpm,
|
|
|
|
|
|
and verify that changing PPV does not affect the ECG signal.
|
|
|
|
|
|
|
|
|
|
|
|
The ECG remains a schematic educational signal; no voltage calibration,
|
|
|
|
|
|
lead placement model, diagnostic criteria, or patient-specific physiology
|
|
|
|
|
|
are simulated.
|
|
|
|
|
|
|
2026-11-12 09:36:13 +01:00
|
|
|
|
## Signal implementation
|
|
|
|
|
|
|
2026-11-12 10:43:11 +01:00
|
|
|
|
- 250 Hz synthetic signal samples, 10 s retained history, 8 s scrolling trace.
|
2026-11-12 09:36:13 +01:00
|
|
|
|
- Cairo draws into the GTK4 `DrawingArea` and scales with the window.
|
|
|
|
|
|
- A conducted QRS records its own respiratory PPV multiplier at event time.
|
|
|
|
|
|
The arterial pulse thus retains the same amplitude throughout the beat.
|
|
|
|
|
|
- Varying PPV does not alter the peripheral pleth pulse or the number of
|
|
|
|
|
|
mechanical pulses.
|
|
|
|
|
|
- Clinical accuracy, dynamic NIBP measurements, alarms, artifacts, calibration,
|
|
|
|
|
|
precise waveform timing, and patient physiology are outside this prototype.
|
|
|
|
|
|
|
|
|
|
|
|
## Project layout
|
|
|
|
|
|
|
|
|
|
|
|
```text
|
|
|
|
|
|
src/
|
|
|
|
|
|
main.rs GTK application entry point
|
|
|
|
|
|
model.rs Parameters, rhythm/capture settings, display flags
|
|
|
|
|
|
simulation.rs QRS-timed mechanical signals, PPV and respiratory synthesis
|
|
|
|
|
|
render.rs Cairo monitor layout, dynamic visibility, fixed ART axis
|
|
|
|
|
|
ui.rs Admin controls and monitor window
|
|
|
|
|
|
```
|
|
|
|
|
|
|
|
|
|
|
|
## Changes since v0.2.2
|
|
|
|
|
|
|
|
|
|
|
|
1. Respiratory PPV with a 0–60% instructor setting, coupled to the simulated
|
|
|
|
|
|
respiratory cycle and applied **per arterial beat**.
|
|
|
|
|
|
2. Five independent display visibility toggles, dynamic waveform row layout,
|
|
|
|
|
|
and independent NIBP-footer hiding.
|
|
|
|
|
|
3. ART axis reduced from **0–250** to **0–200 mmHg**, with inputs adjusted to
|
|
|
|
|
|
the new range.
|
|
|
|
|
|
4. Added regression tests for PPV arithmetic and its effects on ART, for
|
|
|
|
|
|
hiding ECG while retaining mechanical pulses, for active channel selection,
|
|
|
|
|
|
and for arterial-axis normalization.
|
|
|
|
|
|
|
|
|
|
|
|
**Verification:** The archive is generated from these sources. Run `cargo test`
|
|
|
|
|
|
locally to verify the Rust/GTK build; Rust and GTK4 development components
|
|
|
|
|
|
were not available in the environment where this ZIP was prepared.
|
2026-11-12 10:43:11 +01:00
|
|
|
|
|
|
|
|
|
|
## Changes in v0.3.2
|
|
|
|
|
|
|
|
|
|
|
|
1. Added live patient-name header in Admin/Monitor with blank-name fallback.
|
|
|
|
|
|
2. Added four levels of SpO₂ signal quality, quality stars and an irregular,
|
|
|
|
|
|
low-amplitude pleth waveform for Poor quality.
|
|
|
|
|
|
3. Added **Sensor not attached** with `NO SENSOR` status and suppression of
|
|
|
|
|
|
saturation, pulse rate and pleth waveform.
|
|
|
|
|
|
4. Added tests for signal-quality scaling, displayed readings and disconnected
|
|
|
|
|
|
sensor behavior.
|
|
|
|
|
|
|
|
|
|
|
|
**Build verification:** ZIP integrity and patch application were checked.
|
|
|
|
|
|
The GTK4 application and Rust tests were not compiled in the packaging
|
|
|
|
|
|
environment; run `cargo test` and `cargo run --release` on Fedora.
|