# Patient Monitor Simulation (PaMSim) — v0.3.0
|
||||
# Patient Monitor Simulation (PaMSim) — v0.3.2
|
||||
|
||||
A two-window, Rust/GTK4 patient-monitor simulator for teaching on Fedora Linux.
|
||||
|
||||
|
|
@ -20,14 +20,17 @@ session (e.g., Fedora KDE/Wayland).
|
|||
|
||||
## Windows
|
||||
|
||||
**Admin:** Controls ECG rhythm and rate; mechanical capture; SpO₂; invasive
|
||||
arterial pressure, waveform shape and pulse-pressure variation; NIBP; and
|
||||
capnography. Numeric inputs have bidirectionally linked sliders and text fields.
|
||||
**Admin:** Controls patient name; ECG rhythm and rate; mechanical capture;
|
||||
SpO₂, signal quality and sensor connection; invasive arterial pressure,
|
||||
waveform shape and pulse-pressure variation; NIBP; and capnography. Numeric inputs have bidirectionally linked sliders and text fields.
|
||||
Each of the five monitor parameters (ECG, SpO₂/pleth, ART, CO₂, NIBP) has its
|
||||
own **Show on monitor** checkbox.
|
||||
|
||||
**Monitor:** Black background; ECG green, pleth yellow, ART red and capnogram
|
||||
light grey. Waveforms and vital signs disappear together when their respective
|
||||
light grey. The patient name entered in Admin replaces **PATIENT MONITOR** in
|
||||
the header immediately; leaving the name blank restores the generic title.
|
||||
Long names are automatically resized or clipped to fit the header.
|
||||
Waveforms and vital signs disappear together when their respective
|
||||
visibility checkbox is cleared. Remaining waveform rows expand to use the space.
|
||||
Hiding a display channel does **not** stop signal generation or physiological
|
||||
coupling: for example, hiding ECG does not stop ART or pleth pulses. NIBP can
|
||||
|
|
@ -37,6 +40,41 @@ The monitor can be resized and toggled fullscreen with **F11**; **Esc** exits
|
|||
fullscreen. On Wayland, move the monitor to the presentation screen before
|
||||
activating fullscreen. Fullscreen does not lock down the computer.
|
||||
|
||||
## Patient name and pulse oximeter signal state (new in v0.3.2)
|
||||
|
||||
The new **Patient name** text field at the top of Admin immediately replaces the
|
||||
monitor headline. It is optional; an empty or whitespace-only name displays
|
||||
**PATIENT MONITOR**. This is a local teaching label, not a patient record.
|
||||
|
||||
Admin → **Pulse oximetry** has two additional controls:
|
||||
|
||||
- **Sensor not attached** (checkbox): immediately removes the pleth waveform,
|
||||
suppresses SpO₂ and pulse-rate readings, and displays **NO SENSOR** instead
|
||||
of the numeric saturation. This takes precedence over signal quality.
|
||||
- **Signal quality** (dropdown), which changes the visible pleth waveform and
|
||||
quality indicator, independently of the instructor-set SpO₂ percentage:
|
||||
|
||||
| Setting | Pleth waveform | Monitor display |
|
||||
| --- | --- | --- |
|
||||
| **0 — None** | No waveform | SpO₂ `--`, PR `--`, **NO SIGNAL**; no stars |
|
||||
| **1 — Poor** | Small, irregularly amplitude-distorted pulse | 1 star (★) |
|
||||
| **2 — Adequate** | Medium, smoothly damped pulse | 2 stars (★★) |
|
||||
| **3 — Good** | Full nominal, smoothly damped pulse | 3 stars (★★★) |
|
||||
|
||||
Default: sensor attached; signal quality **3 — Good**. At levels 1–3, the
|
||||
instructor-set SpO₂ number is shown only when mechanical pulses are detected.
|
||||
The signal quality does **not** randomly alter that number: it affects the
|
||||
pleth morphology and the quality indicator only. The simulated pulse rate
|
||||
remains derived from mechanically captured QRS complexes.
|
||||
|
||||
The degraded waveform uses deterministic, smoothly interpolated pseudo-noise
|
||||
that scales only existing pleth pulses: it cannot manufacture pulses during
|
||||
PEA, ventricular fibrillation or asystole. Unplugging a sensor or selecting
|
||||
quality 0 hides even previously buffered pleth samples immediately. ECG and
|
||||
ART continue unaffected. The **Show on monitor** checkbox still hides the
|
||||
entire SpO₂ row; sensor/quality controls leave the row visible so the status
|
||||
can be taught.
|
||||
|
||||
## ECG and mechanical coupling
|
||||
|
||||
- ECG rhythms: sinus, atrial fibrillation, monomorphic ventricular tachycardia,
|
||||
|
|
@ -102,9 +140,31 @@ The five **Show on monitor** checkboxes affect the presentation only:
|
|||
Remaining visible waveforms keep their relative order and receive an equal
|
||||
share of the available area; zero waveform channels is supported.
|
||||
|
||||
## ECG morphology correction (v0.3.1)
|
||||
|
||||
At high heart rates, the previous ECG generator expressed QRS width as a fixed
|
||||
fraction of the R–R interval and sampled at 100 Hz. Narrow R waves could fall
|
||||
between samples, making successive QRS complexes appear to vary dramatically
|
||||
in amplitude despite a regular sinus rhythm.
|
||||
|
||||
- Synthetic P/QRS/T components now have widths specified in milliseconds,
|
||||
not fixed fractions of the cardiac cycle. P/T positioning and widths adapt
|
||||
to the R–R interval, while the QRS width remains essentially constant.
|
||||
- Sampling increased to **250 Hz** (4 ms intervals) to capture QRS peaks
|
||||
more consistently, including at **189–240 bpm**.
|
||||
- The ECG morphology remains independent of the respiratory arterial PPV
|
||||
setting. The arterial and pleth delays and mechanical capture options are
|
||||
unchanged.
|
||||
- Regression tests compare peak heights at 60, 75, 120, 189 and 240 bpm,
|
||||
and verify that changing PPV does not affect the ECG signal.
|
||||
|
||||
The ECG remains a schematic educational signal; no voltage calibration,
|
||||
lead placement model, diagnostic criteria, or patient-specific physiology
|
||||
are simulated.
|
||||
|
||||
## Signal implementation
|
||||
|
||||
- 100 Hz synthetic signal samples, 10 s retained history, 8 s scrolling trace.
|
||||
- 250 Hz synthetic signal samples, 10 s retained history, 8 s scrolling trace.
|
||||
- Cairo draws into the GTK4 `DrawingArea` and scales with the window.
|
||||
- A conducted QRS records its own respiratory PPV multiplier at event time.
|
||||
The arterial pulse thus retains the same amplitude throughout the beat.
|
||||
|
|
@ -139,3 +199,17 @@ src/
|
|||
**Verification:** The archive is generated from these sources. Run `cargo test`
|
||||
locally to verify the Rust/GTK build; Rust and GTK4 development components
|
||||
were not available in the environment where this ZIP was prepared.
|
||||
|
||||
## Changes in v0.3.2
|
||||
|
||||
1. Added live patient-name header in Admin/Monitor with blank-name fallback.
|
||||
2. Added four levels of SpO₂ signal quality, quality stars and an irregular,
|
||||
low-amplitude pleth waveform for Poor quality.
|
||||
3. Added **Sensor not attached** with `NO SENSOR` status and suppression of
|
||||
saturation, pulse rate and pleth waveform.
|
||||
4. Added tests for signal-quality scaling, displayed readings and disconnected
|
||||
sensor behavior.
|
||||
|
||||
**Build verification:** ZIP integrity and patch application were checked.
|
||||
The GTK4 application and Rust tests were not compiled in the packaging
|
||||
environment; run `cargo test` and `cargo run --release` on Fedora.
|
||||
|
|
|
|||