# Patient Monitor Simulation (PaMSim) — v0.5.4
|
||
|
||
A two-window, Rust/GTK4 patient-monitor simulator for teaching on Fedora Linux.
|
||
|
||
> **Synthetic training signals only — not for clinical use.** ECG, arterial pressure,
|
||
> plethysmography and capnography are schematic mathematical approximations. The
|
||
> simulation has no clinical validation and is not a medical device.
|
||
|
||
## Chest compressions (v0.5.4)
|
||
|
||
The **Chest compressions** Admin group provides a large **Start Chest Compressions**
|
||
button (green background). While active it changes to **Stop Chest Compressions**
|
||
(red background). **Compression quality** can be changed at any time:
|
||
**Good**, **Adequate** (default), **Poor**, or **Bad**.
|
||
|
||
- Compressions are generated independently of the ECG at **90–110/min**, with
|
||
a new reproducible pseudo-random compression rate and pressure each stroke.
|
||
**The ECG rhythm and electrical HR setting are not changed.**
|
||
- Every compression adds movement artifact to the ECG. In a patient with
|
||
**existing intrinsic arterial perfusion**, the native arterial waveform is
|
||
preserved but distorted by compression-synchronous pressure artifact.
|
||
- With **no intrinsic arterial pulse** (e.g., VF, PVT, asystole or PEA),
|
||
each effective compression generates a broad, blunt arterial pressure wave:
|
||
|
||
| Compression quality | Compression-generated ART systolic peak |
|
||
| --- | --- |
|
||
| Good | 90 mmHg ±10% (81–99 mmHg) |
|
||
| Adequate (default) | 70 mmHg ±10% (63–77 mmHg) |
|
||
| Poor | 50 mmHg, +10% / −50% (25–55 mmHg) |
|
||
| Bad | No compression-generated arterial pressure; 0–5 mmHg baseline drift |
|
||
|
||
During pulseless CPR, **ART SYS** shows the current compression peak,
|
||
while **DIA/MAP** remain `--` (not estimable from this synthetic pressure).
|
||
- SpO₂ numeric saturation and the pleth waveform remain visible **only when
|
||
compression quality is Good**, provided the sensor is connected and not
|
||
occluded by an NIBP cuff. Their effective signal quality is forced to
|
||
**Poor (★)**, irrespective of the normal Admin quality setting. The displayed
|
||
pleth-derived **PR follows the compression rate (90–110/min)**, not the ECG.
|
||
Adequate, Poor and Bad CPR display **NO READING**; detached sensors display
|
||
**NO SENSOR**. Compression-induced Poor quality does not schedule the usual
|
||
intermittent dropout; that resumes when ordinary Poor signal mode returns.
|
||
- Existing numerical alarms and rhythm alarms remain active. ART SYS limits
|
||
apply to compression-generated systolic peaks; NIBP scheduling remains unchanged; its cycle duration still depends on
|
||
effective SpO₂ quality at the start of a measurement.
|
||
|
||
These pressure targets are **didactic assumptions**, not validated CPR
|
||
hemodynamic criteria. The exercise must not be used to judge real compression
|
||
quality or clinical perfusion.
|
||
|
||
## Build and run on Fedora
|
||
|
||
```bash
|
||
sudo dnf install rust cargo gcc pkgconf-pkg-config gtk4-devel pipewire-utils
|
||
cargo test
|
||
cargo run --release
|
||
```
|
||
|
||
Run these commands from the extracted `patient-monitor-simulation` folder.
|
||
Window titles show the application version. GTK4 requires a functional graphical
|
||
session (e.g., Fedora KDE/Wayland).
|
||
|
||
## Windows
|
||
|
||
**Admin sections** have a narrow colored stripe on their left edge to match the
|
||
monitor: ECG **green**, SpO₂ **yellow**, invasive arterial pressure **red**,
|
||
capnography **light grey**, NIBP **red**. The Patient section remains
|
||
neutral. These accents are drawn by GTK and remain visible independently of
|
||
light/dark themes. This is a visual-only change; control behavior is unchanged.
|
||
|
||
**Admin:** Controls patient name; ECG rhythm and rate; mechanical capture;
|
||
SpO₂, signal quality and sensor connection; invasive arterial pressure,
|
||
waveform shape and pulse-pressure variation; NIBP; and capnography. Numeric inputs have bidirectionally linked sliders and text fields.
|
||
Each of the five monitor parameters (ECG, SpO₂/pleth, ART, CO₂, NIBP) has its
|
||
own **Show on monitor** checkbox.
|
||
|
||
**Monitor:** Black background; ECG green, pleth yellow, ART red and capnogram
|
||
light grey. The patient name entered in Admin replaces **PATIENT MONITOR** in
|
||
the header immediately; leaving the name blank restores the generic title.
|
||
Long names are automatically resized or clipped to fit the header.
|
||
Waveforms and vital signs disappear together when their respective
|
||
visibility checkbox is cleared. Remaining waveform rows expand to use the space.
|
||
Hiding a display channel does **not** stop signal generation or physiological
|
||
coupling: for example, hiding ECG does not stop ART or pleth pulses. NIBP occupies a dedicated right-column row with an **empty waveform cell** on the
|
||
left. All waveform rows and their corresponding numeric boxes share the same
|
||
height and horizontal boundaries. NIBP can be hidden independently; its entire
|
||
row is then removed and the remaining rows expand evenly.
|
||
|
||
The monitor can be resized and toggled fullscreen with **F11**; **Esc** exits
|
||
fullscreen. On Wayland, move the monitor to the presentation screen before
|
||
activating fullscreen. Fullscreen does not lock down the computer.
|
||
|
||
## Audible alarms (v0.5.3)
|
||
|
||
Sound is **enabled by default**. The Admin window has an **Audible alarms**
|
||
checkbox for muting and a **Test tone** button (works even while muted).
|
||
Visual alarms keep blinking normally when sound is off.
|
||
|
||
- **Critical rhythm alarm:** VT, PVT, VF and asystole produce a double beep
|
||
(two short 880 Hz tones), repeated approximately every 1.5 seconds.
|
||
- **Numerical and SpO₂ technical alarms:** out-of-range values, **NO SENSOR**
|
||
and **NO READING** produce a single 700 Hz beep, repeated approximately
|
||
every 2.5 seconds.
|
||
- Critical alarms have priority: only one tone pattern is played even if
|
||
multiple alarms are active. The sound stops when all alarms clear or when
|
||
the checkbox is unchecked. Alarms on hidden monitor channels are excluded.
|
||
- Values not yet available (e.g. NIBP before a completed measurement) do not
|
||
trigger numerical limit alarms. The cuff measurement cycle by itself is not
|
||
an alarm. Audio playback is separate from waveform rendering and simulation.
|
||
|
||
On Fedora, sound is played through **`pw-play`** from `pipewire-utils`, with
|
||
**`paplay`** as a fallback. No additional Rust crates or sound files are needed:
|
||
the application creates its own short WAV tones at runtime. If neither player
|
||
is installed, Admin reports that audio is unavailable; the monitor continues
|
||
to work normally. `Test tone` helps verify the desktop audio output. The
|
||
application uses the system output device and mixer volume.
|
||
|
||
**Important:** These simple teaching tones do not comply with medical-device
|
||
alarm standards (including IEC 60601-1-8) and must never be relied on for
|
||
clinical monitoring.
|
||
|
||
## Numerical alarm limits (v0.4.0; layout and presets updated in v0.4.1)
|
||
|
||
Each displayed numeric measurement has an **independent optional lower (↓) and
|
||
upper (↑) alarm threshold** in the corresponding Admin section. The lower
|
||
and upper entries are placed **in the same row as the corresponding slider**.
|
||
Derived values with no separate slider (PR, ART MAP, NIBP MAP) have their own
|
||
compact alarm rows. The Admin window opens wider to accommodate these controls.
|
||
Alarm fields are text inputs; leave a field **blank** (displayed as `off`) to disable just
|
||
that threshold. With both fields blank, the whole numeric alarm is disabled.
|
||
The value may equal either limit without triggering an alarm; only readings
|
||
strictly below/above the respective threshold are out of range. If both limits
|
||
are present, the lower limit must not exceed the upper limit. Invalid input is
|
||
highlighted and the last valid configuration stays active.
|
||
|
||
| Monitor section | Independently configurable alarms |
|
||
| --- | --- |
|
||
| ECG | HR |
|
||
| SpO₂ | Oxygen saturation (SpO₂), derived pulse rate (PR) |
|
||
| Invasive arterial pressure (ART) | Systolic (SYS), mean (MAP), diastolic (DIA), PPV |
|
||
| NIBP | Systolic (SYS), mean (MAP), diastolic (DIA) |
|
||
| Capnography | End-tidal CO₂ (EtCO₂), respiratory rate (RR) |
|
||
|
||
The current limits are displayed in small type at the **upper-right corner** of
|
||
each channel, with an arrow for the corresponding direction. A disabled
|
||
boundary is displayed as `---`. For instance:
|
||
|
||
```text
|
||
⚠ SYS ↓80 ↑180
|
||
MAP ↓60 ↑---
|
||
DIA ↓--- ↑---
|
||
```
|
||
|
||
### Alert presets
|
||
|
||
The **Alert presets** dropdown at the top of Admin offers **Custom** (manual)
|
||
and **Adult**. Choosing **Adult** immediately sets all alarm limits and updates
|
||
the lower/upper text fields alongside the sliders. It changes *only* alarm limits:
|
||
it does not change heart rate, blood pressure, channel visibility or any other
|
||
simulated physiological setting. Editing any alarm field after applying Adult
|
||
switches the selector to **Custom** without changing other alarm limits. Select
|
||
**Adult** again to reapply the complete preset.
|
||
|
||
| Parameter | Lower | Upper |
|
||
| --- | ---: | ---: |
|
||
| HR (bpm) | 60 | 100 |
|
||
| SpO₂ (%) | 90 | --- |
|
||
| PR (bpm) | --- | --- |
|
||
| ART SYS (mmHg) | 80 | 180 |
|
||
| ART MAP (mmHg) | 60 | 100 |
|
||
| ART DIA (mmHg) | --- | --- |
|
||
| PPV (%) | --- | 10 |
|
||
| NIBP SYS (mmHg) | 80 | 180 |
|
||
| NIBP MAP (mmHg) | 60 | 100 |
|
||
| NIBP DIA (mmHg) | --- | --- |
|
||
| EtCO₂ (mmHg) | 30 | 50 |
|
||
| RR (/min) | 10 | 20 |
|
||
|
||
`---` denotes a disabled threshold. Adult is a **teaching convenience preset**,
|
||
not a clinically validated set of alarm limits. Initially the selector is
|
||
**Custom**, with all limits disabled (as in v0.4.0).
|
||
|
||
When a number violates a configured limit, **that number flashes by inverting
|
||
its foreground and background colors** (500 ms inverted / 500 ms normal).
|
||
Each pressure component (SYS, DIA, MAP) and the PPV reading flash independently. The waveform
|
||
itself remains unchanged. All alarms are **disabled by default** so old
|
||
scenarios do not unexpectedly flash.
|
||
|
||
Numerical threshold alarms only evaluate actual displayed measurements:
|
||
`--`, `NO SENSOR`, and other unavailable readings do not trigger low-value
|
||
alarms. ECG-derived PR and signal visibility continue to work as in v0.3.2.
|
||
Alarms are presentation-only: there is **no audible alarm, alarm latching,
|
||
priority classification or clinical alarm validation**. Instructor settings
|
||
are held in memory and reset on restart.
|
||
|
||
## Patient name and pulse oximeter signal state (new in v0.3.2)
|
||
|
||
The new **Patient name** text field at the top of Admin immediately replaces the
|
||
monitor headline. It is optional; an empty or whitespace-only name displays
|
||
**PATIENT MONITOR**. This is a local teaching label, not a patient record.
|
||
|
||
Admin → **Pulse oximetry** has two additional controls:
|
||
|
||
- **Sensor not attached** (checkbox): immediately removes the pleth waveform,
|
||
suppresses SpO₂ and pulse-rate readings, and displays **NO SENSOR** instead
|
||
of the numeric saturation. This takes precedence over signal quality.
|
||
- **Signal quality** (dropdown), which changes the visible pleth waveform and
|
||
quality indicator, independently of the instructor-set SpO₂ percentage:
|
||
|
||
| Setting | Pleth waveform | Monitor display |
|
||
| --- | --- | --- |
|
||
| **0 — None** | No waveform | Flashing **NO READING**, PR `--`; no stars |
|
||
| **1 — Poor** | Small, irregularly distorted pulse, with temporary losses of signal | 1 star (★) when available; blinking **NO READING** during loss |
|
||
| **2 — Adequate** | Medium, smoothly damped pulse | 2 stars (★★) |
|
||
| **3 — Good** | Full nominal, smoothly damped pulse | 3 stars (★★★) |
|
||
|
||
Default: sensor attached; signal quality **3 — Good**. At levels 1–3, the
|
||
instructor-set SpO₂ number is shown only when mechanical pulses are detected.
|
||
In **Poor** quality, the signal also has intermittent complete dropouts:
|
||
**30–180 seconds of signal** between interruptions, with each interruption
|
||
lasting **5–20 seconds**. During an interruption the pleth trace disappears,
|
||
SpO₂ and PR are unavailable, and **NO READING** flashes. Following recovery,
|
||
the Poor waveform and numerical readings return automatically. These intervals
|
||
are pseudo-random and driven by simulation time, independent of frame rate;
|
||
they reset when Poor quality is selected again. The dropout does not affect
|
||
the actual saturation setting, ECG, derived mechanical pulse generation, or ART.
|
||
The simulated pulse rate remains derived from captured QRS complexes.
|
||
|
||
The degraded waveform uses deterministic, smoothly interpolated pseudo-noise
|
||
that scales only existing pleth pulses: it cannot manufacture pulses during
|
||
PEA, PVT, ventricular fibrillation or asystole. Unplugging a sensor or selecting
|
||
quality 0 hides even previously buffered pleth samples immediately. ECG and
|
||
ART continue unaffected. The **Show on monitor** checkbox still hides the
|
||
entire SpO₂ row; sensor/quality controls leave the row visible so the status
|
||
can be taught.
|
||
|
||
## NIBP measurement cycle (v0.5.0)
|
||
|
||
NIBP is **red**, with a dedicated value box in the **right-hand numeric column**.
|
||
An **empty left-hand waveform cell** occupies the same row, keeping every numeric
|
||
value box horizontally aligned with its waveform region. The previous static
|
||
NIBP footer has been removed. In Admin, choose an interval
|
||
of **1, 3, 5, 10, 20, 30, or 60 minutes** (default 5). The initial cuff cycle
|
||
begins when the simulation starts; subsequent intervals run **from the end of
|
||
the last measurement**, not from its beginning. **Measure now** starts an
|
||
additional cycle if none is in progress. The small **horizontal progress bar**
|
||
shows the fraction of the waiting interval elapsed; a **vertical 0–200 mmHg
|
||
gauge** shows the current cuff pressure during measurement.
|
||
|
||
The duration and success are determined by the signal quality **at the start
|
||
of the cycle** (including an absent sensor or an ongoing Poor-signal dropout).
|
||
Changing quality while a cycle is running does not retroactively change its
|
||
planned duration:
|
||
|
||
| Starting SpO₂ quality | Cycle duration | Result |
|
||
| --- | --- | --- |
|
||
| Good (★★★) | 20 s | Measured |
|
||
| Adequate (★★) | 40 s | Measured |
|
||
| Poor (★) | Randomly 60–120 s | Simulated 65% chance of a reading |
|
||
| None / no sensor | 120 s | No reading |
|
||
| **No reading** checkbox | 120 s | No reading, regardless of quality |
|
||
|
||
The cuff inflates to **200 mmHg** then deflates. When the starting signal is
|
||
Poor or None, it deflates to about **67 mmHg**, reinflates to about **133 mmHg**,
|
||
and finally deflates to zero. At the end, the completed value becomes visible
|
||
and **blinks five times by appearing/disappearing** (without the inverted-color
|
||
alarm effect). If the cycle fails, the displayed reading is `---/---`, with
|
||
MAP `---`. Previous NIBP values remain visible while the next cuff cycle is
|
||
in progress, until the new reading is available.
|
||
|
||
The **first successful** measurement uses the initial NIBP SYS/DIA values from
|
||
Admin. Subsequent successful values derive from the current invasive arterial
|
||
SYS/DIA settings using independently sampled percentage deviations:
|
||
|
||
| Starting quality | Deviation relative to IBP |
|
||
| --- | --- |
|
||
| Good | −5% to +5% |
|
||
| Adequate | −10% to +5% |
|
||
| Poor | −15% to −5%, if a reading succeeds |
|
||
| None | No reading |
|
||
|
||
Optional **Cuff on SpO₂ extremity** changes the *effective* pulse-oximeter
|
||
quality while the cuff is inflated: above invasive systolic pressure = None;
|
||
between invasive systolic and diastolic pressure = Poor; below invasive
|
||
diastolic pressure = the configured quality. The quality never becomes better
|
||
than the instructor's setting. **NO READING**, missing pleth waveform and
|
||
restoration of the waveform are handled as before.
|
||
|
||
Optional **Cuff on ART extremity** modifies the simulated arterial transducer
|
||
trace during cuff inflation: above invasive SYS it freezes at the last measured
|
||
pressure; between invasive SYS and DIA it pulsates between the cuff pressure
|
||
and the currently available NIBP SYS reference; below invasive DIA it returns
|
||
to the normal arterial waveform. The ART numerical SYS/DIA/MAP values also
|
||
reflect this altered trace. With absent mechanical cardiac capture, the
|
||
previous pulseless 0–5 mmHg simulation is retained and no artificial pulses
|
||
are introduced. Both cuff-interference options are off by default.
|
||
|
||
These cuff and SpO₂ interactions are **deliberately simplified training
|
||
artifacts**; they should not be interpreted as validated cuff oscillometry.
|
||
|
||
## ECG and mechanical coupling
|
||
|
||
- ECG rhythms: sinus, atrial fibrillation, monomorphic ventricular tachycardia
|
||
(potentially perfusing), pulseless ventricular tachycardia (PVT),
|
||
ventricular fibrillation and asystole.
|
||
- Mechanical capture: **1:1**, **2:1**, **3:1**, **None (PEA)**, or deterministic
|
||
pseudo-random capture (~65%). PR is derived from conducted QRS events, never
|
||
controlled by a pulse-rate slider.
|
||
- A captured QRS leads to an ART upstroke after a schematic **120 ms**, and a
|
||
peripheral pleth pulse after **300 ms**. Pleth is strongly damped/filtered.
|
||
- VT and PVT selections start with a heart rate of **160 bpm**, and update
|
||
both the HR slider and numeric entry. The instructor may modify this afterward.
|
||
- PVT, VF, asystole and PEA suppress mechanical pulses; the ART trace then
|
||
drifts smoothly at 0–5 mmHg rather than representing a pulsatile pressure.
|
||
The ART **SYS** number shows the rounded low transducer pressure (0–5 mmHg),
|
||
while DIA and MAP show `--` because they cannot be reliably estimated.
|
||
Previous pressure samples scroll out of view. NIBP follows a separate,
|
||
time-based simulated cuff-measurement cycle.
|
||
|
||
## Invasive arterial pressure and PPV (new in v0.3)
|
||
|
||
The arterial waveform has a fixed **0–200 mmHg** display axis, with tick marks
|
||
at **0, 50, 100, 150 and 200 mmHg**. Nominal ART systolic and diastolic inputs
|
||
are limited to 0–200 mmHg.
|
||
|
||
**Pulse pressure variation (PPV)** is adjustable from **0% to 60%**, in 1%
|
||
increments, via synchronized slider and text field (Admin → Invasive arterial
|
||
blood pressure). Its default is **0%** (no respiratory modulation). The
|
||
respiratory cycle comes from the same respiratory-rate parameter as the
|
||
capnogram. **RR = 0** disables the respiratory modulation (apnea).
|
||
|
||
The synthetic PPV definition is:
|
||
|
||
```text
|
||
PPV (%) = 100 × (PPmax − PPmin) / ((PPmax + PPmin) / 2)
|
||
```
|
||
|
||
The model applies a sinusoidal variation to **pulse pressure**, sampled for
|
||
each mechanically conducted beat. The diastolic baseline remains nominally
|
||
fixed, so this is primarily a **beat-to-beat systolic swing**, not an arbitrary
|
||
vertical translation of the entire arterial trace. It does not affect ECG,
|
||
mechanical capture or pleth amplitude. This is a visual teaching model, not a
|
||
validated predictor of fluid responsiveness.
|
||
|
||
**Example:** ART 120/75 mmHg has nominal pulse pressure 45 mmHg. At PPV 20%,
|
||
the ideal pulse-pressure extrema are 40.5 and 49.5 mmHg, corresponding to
|
||
systolic peaks of approximately 115.5 and 124.5 mmHg (assuming the diastolic
|
||
baseline stays at 75 mmHg). The rendered waveform shape and finite number of
|
||
beats per respiratory cycle mean the visible maxima may not reach the ideal
|
||
extrema exactly.
|
||
|
||
The ART sidebar displays the **nominal** systolic/diastolic pressures and MAP,
|
||
plus the **configured** PPV percentage when PPV is greater than 0. It does not
|
||
measure PPV from sampled arterial data. At very high nominal systolic pressure,
|
||
the highest modulated pulses can exceed the 200 mmHg scale; the displayed
|
||
curve is then **clipped** at the top, without rescaling the axis.
|
||
|
||
## Display options (new in v0.3)
|
||
|
||
The five **Show on monitor** checkboxes affect the presentation only:
|
||
|
||
| Checkbox | Hidden monitor content |
|
||
| --- | --- |
|
||
| ECG | ECG waveform, HR and rhythm alarm (when present) |
|
||
| Pulse oximetry | Pleth waveform, SpO₂ and PR |
|
||
| Invasive arterial blood pressure | ART waveform, pressure values, MAP and PPV label |
|
||
| Capnography | CO₂ waveform, EtCO₂ and RR |
|
||
| NIBP | Timed cuff-measurement values and red right-column value box |
|
||
|
||
Remaining visible waveform rows and the NIBP-only row share an equal
|
||
portion of the available height and remain aligned across both columns; zero
|
||
waveform channels is supported.
|
||
|
||
## ECG morphology correction (v0.3.1)
|
||
|
||
At high heart rates, the previous ECG generator expressed QRS width as a fixed
|
||
fraction of the R–R interval and sampled at 100 Hz. Narrow R waves could fall
|
||
between samples, making successive QRS complexes appear to vary dramatically
|
||
in amplitude despite a regular sinus rhythm.
|
||
|
||
- Synthetic P/QRS/T components now have widths specified in milliseconds,
|
||
not fixed fractions of the cardiac cycle. P/T positioning and widths adapt
|
||
to the R–R interval, while the QRS width remains essentially constant.
|
||
- Sampling increased to **250 Hz** (4 ms intervals) to capture QRS peaks
|
||
more consistently, including at **189–240 bpm**.
|
||
- The ECG morphology remains independent of the respiratory arterial PPV
|
||
setting. The arterial and pleth delays and mechanical capture options are
|
||
unchanged.
|
||
- Regression tests compare peak heights at 60, 75, 120, 189 and 240 bpm,
|
||
and verify that changing PPV does not affect the ECG signal.
|
||
|
||
The ECG remains a schematic educational signal; no voltage calibration,
|
||
lead placement model, diagnostic criteria, or patient-specific physiology
|
||
are simulated.
|
||
|
||
## Signal implementation
|
||
|
||
- 250 Hz synthetic signal samples, 10 s retained history, 8 s scrolling trace.
|
||
- Cairo draws into the GTK4 `DrawingArea` and scales with the window.
|
||
- A conducted QRS records its own respiratory PPV multiplier at event time.
|
||
The arterial pulse thus retains the same amplitude throughout the beat.
|
||
- Varying PPV does not alter the peripheral pleth pulse or the number of
|
||
mechanical pulses.
|
||
- Clinical accuracy, calibration, artifact realism and patient-specific physiology
|
||
are outside this prototype; its NIBP cycle and alarms are synthetic.
|
||
|
||
## Project layout
|
||
|
||
```text
|
||
src/
|
||
main.rs GTK application entry point
|
||
model.rs Parameters, rhythm/capture settings, display flags
|
||
simulation.rs QRS-timed mechanical signals, PPV and respiratory synthesis
|
||
render.rs Cairo monitor layout, dynamic visibility, fixed ART axis
|
||
nibp.rs Scheduled cuff measurements and measurement results
|
||
ui.rs Admin controls and monitor window
|
||
```
|
||
|
||
## Changes since v0.2.2
|
||
|
||
1. Respiratory PPV with a 0–60% instructor setting, coupled to the simulated
|
||
respiratory cycle and applied **per arterial beat**.
|
||
2. Five independent display visibility toggles, dynamic waveform row layout,
|
||
and independent NIBP value-box hiding.
|
||
3. ART axis reduced from **0–250** to **0–200 mmHg**, with inputs adjusted to
|
||
the new range.
|
||
4. Added regression tests for PPV arithmetic and its effects on ART, for
|
||
hiding ECG while retaining mechanical pulses, for active channel selection,
|
||
and for arterial-axis normalization.
|
||
|
||
**Verification:** The archive is generated from these sources. Run `cargo test`
|
||
locally to verify the Rust/GTK build; Rust and GTK4 development components
|
||
were not available in the environment where this ZIP was prepared.
|
||
|
||
## Changes in v0.3.2
|
||
|
||
1. Added live patient-name header in Admin/Monitor with blank-name fallback.
|
||
2. Added four levels of SpO₂ signal quality, quality stars and an irregular,
|
||
low-amplitude pleth waveform for Poor quality.
|
||
3. Added **Sensor not attached** with `NO SENSOR` status and suppression of
|
||
saturation, pulse rate and pleth waveform.
|
||
4. Added tests for signal-quality scaling, displayed readings and disconnected
|
||
sensor behavior.
|
||
|
||
**Build verification:** ZIP integrity and patch application were checked.
|
||
The GTK4 application and Rust tests were not compiled in the packaging
|
||
environment; run `cargo test` and `cargo run --release` on Fedora.
|
||
|
||
|
||
## v0.4.2 — Rhythm, sensor and missing-reading alarms
|
||
|
||
- **VT** remains potentially perfusing. The new **PVT** (pulseless ventricular tachycardia) rhythm retains organized broad QRS complexes but produces no arterial/pleth pulses. VF and asystole are also pulseless regardless of capture selection.
|
||
- **ECG**: a flashing **ARRHYTHMIA** message for VT, PVT and VF; a flashing **ASYSTOLE** message for asystole. These event alarms do not depend on numerical HR limits.
|
||
- **SpO₂**: when the probe is disconnected, a flashing **NO SENSOR** replaces the saturation reading. Quality 0, loss of mechanical capture or a pulseless rhythm results in flashing **NO READING** (no numerical SpO₂/PR). Sensor disconnection takes precedence. No pleth waveform is displayed in these conditions.
|
||
- **ART**: without mechanical pulses (including PVT/VF/asystole/PEA), the simulated invasive pressure line has slow non-pulsatile drift between **0 and 5 mmHg**. In v0.4.2, the numeric display showed `--/--`; v0.4.3 changes this to a rounded measured SYS of 0–5 mmHg, with `--` for DIA and MAP.
|
||
- **ECG rhythm name** has been removed from the monitor; only HR and the associated alarm message, when present, are visible. The instructor continues to select the rhythm in Admin.
|
||
- Numerical alarm thresholds and the Adult preset from v0.4.1 are unchanged. Synthetic demonstration behavior only; not suitable for clinical decisions.
|
||
|
||
|
||
## v0.4.3 — Low arterial pressure, VT rate preset and intermittent Poor SpO₂
|
||
|
||
- Pulseless rhythms, PEA or otherwise absent mechanical capture: invasive SYS
|
||
reports the **current measured 0–5 mmHg** pressure, e.g. `3/--` with
|
||
`MAP -- mmHg`. **Only SYS** is tested against configured ART SYS limits;
|
||
no spurious MAP/DIA threshold alarms are generated from missing readings.
|
||
- Selecting **VT** or **PVT** in Admin sets HR to **160 bpm** and updates the
|
||
associated slider and text field. Subsequent manual rate changes are allowed.
|
||
Perfusing VT still produces QRS-linked pulses subject to capture settings;
|
||
PVT remains pulseless.
|
||
- **Poor (★)** SpO₂ signal periodically changes to **NO READING** for a
|
||
pseudo-random **5–20 seconds**, after a pseudo-random **30–180 seconds** of
|
||
usable low-quality signal. The technical alarm blinks, SpO₂/PR readings are
|
||
suppressed, and the pleth waveform is absent. On recovery it resumes at Poor
|
||
quality. The simulated signal history retains visible gaps for outages.
|
||
Choosing another signal quality immediately clears the intermittent outage.
|
||
- No changes to Adult alarm presets, alarm limit ordering, or the 0–200 mmHg
|
||
arterial scale. All behavior is synthetic and intended for teaching only.
|
||
|
||
## v0.5.1 — Aligned monitor columns
|
||
|
||
The left waveform column and right numeric column now share the same row grid.
|
||
NIBP reserves an empty cell in the left waveform column instead of shortening
|
||
the other numeric rows. Hiding any channel reflows both columns together.
|
||
No NIBP measurement logic or alarm behavior was changed.
|
||
|
||
## v0.5.3 — Audible alarm output
|
||
|
||
- Alarm-tone playback through Fedora PipeWire (`pw-play`) or PulseAudio
|
||
compatibility (`paplay`), without any new Rust library dependencies.
|
||
- Admin controls to mute/unmute alarms or play a test tone.
|
||
- Distinct critical-rhythm and standard/technical patterns; critical priority.
|
||
- Audio keeps running independently of the monitor's drawing callback.
|
||
- Visual alarms and existing physiological simulation remain unchanged.
|